Ultra Rapid and Highly Sensitive Disperser-less Liquid-liquid Microextraction of Organophosphate Pesticides Prior to Gas Chromatography with Mass Spectrometry Detection

authors:

avatar Nadereh Rahbar 1 , * , avatar Mohammad Mehdi Hesami 2 , avatar Zahra Ramezani 1 , avatar Aye Savarizadeh 1

Department of Medicinal Chemistry, School of Pharmacy, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
Student Research Committee, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran

how to cite: Rahbar N, Hesami M M , Ramezani Z , Savarizadeh A. Ultra Rapid and Highly Sensitive Disperser-less Liquid-liquid Microextraction of Organophosphate Pesticides Prior to Gas Chromatography with Mass Spectrometry Detection. Jundishapur J Health Sci. 2020;12(4):e109594. https://doi.org/10.5812/jjhs.109594.

Abstract

Background:

Organophosphates (OPPs) are toxic chemicals that can cause serious health problems through poisoning water and food.

Objectives:

A very simple and fast disperser-less liquid microextraction strategy before chromatographic detection was designed for the analysis of organophosphates in various water solutions.

Methods:

A 60 µL aliquot of chloroform, as extraction solvent (without using disperser), was introduced into the sample solution by rapid injection, and the sedimented organic phase was analyzed to assay some organophosphates.

Results:

Analytical characteristics, including limits of detection (0.0003 - 0.001 µg.L-1), linear dynamic ranges (0.001 - 100 µg.L-1), relative standard deviations (2.5 - 10), enrichment factors (up to 238), and extraction recoveries (84% - 108%), indicated the high efficacy of the developed method for analyzing the target analytes.

Conclusions:

The proposed procedure was effectively used for the analysis of the OPPs in real tap water, river water, and fruit juice samples. In the present study, the examined analytes were in the range of 0.07 - 1.56 µg.L-1.

1. Background

Organophosphates (OPPs) are toxic chemicals that can cause serious health problems through poisoning water and food. These compounds can extensively control pests; however, they also undergo fast degradation in the environment. Therefore, various types of OPPs are extensively used in agriculture (1, 2). The potential threats of increased concentration of OPPS to the environment have attracted significant attention, which has resulted in the development of sensitive and selective detection methods. OPPs are detected in real samples using various methods like gas chromatography with mass detection (GC/MS) (3, 4), high performance liquid chromatography (HPLC) (5), electrochemical techniques (6, 7), micellar electrokinetic chromatography (2), spectrophotometry (8, 9), and spectrofluorimetry (10, 11). OPPs are commonly found in real samples at low concentrations and complicated matrices. Therefore, before performing analyses, sample treatment is essential for both purification and preconcentration. In general, sample treatment and isolation of analytes from the matrix is an important step in the analytical procedure, which affects the whole analytical technique. Reduced time of sample preparation steps, the sample amount, and the consumption of hazardous reagents and solvents can increase the ease of operation and safety for the environment. Besides, it provides a green sample preparation procedure. In the past few decades, several miniaturized extraction techniques have been developed as alternatives to classical methods of sample preparation, namely liquid-liquid extraction (LLE) and solid phase extraction (SPE). However, these techniques contain disadvantages such as large consumption of time, toxic organic solvents, and sample amounts (12). The extraction strategies like solid-phase microextraction (SPME) (13) and micro solid-phase extraction (µ-SPE) (3) have been developed for sample treatment to improve the SPE method.

Besides, a number of new liquid microextraction methods are introduced as alternatives to LLE, such as dispersive liquid-liquid microextraction (DLLME) (14), hollow-fiber liquid-phase microextraction (HF-LPME) (15), and single drop microextraction (SDME) (16).

Among the different types of liquid microextraction methods, the DLLME strategy presents some appealing features such as simplicity, rapidity, and high preconcentration factor (17). In this technique, a water-immiscible organic solvent (extractant) is mixed with a dispersive solvent, which is water-miscible. The mixture is entered into an aqueous sample solution by rapid injection, forming a cloudy phase with fine droplets of the extractant (18, 19). DLLME technique has been operated via different strategies for extraction of OPPs from real matrices (4, 18, 20-26). In all of these procedures, a disperser solvent (0.5 - 1.5 mL) has been used to disperse an organic extraction solvent in the whole sample solution, which accelerates the extraction of analytes. However, the polarity of the sample solution relatively decreases in the presence of a disperser solvent, which in turn increases the solubility of lipophilic analytes in a water medium that causes decreased extraction efficiency (27). In addition, the exclusion of a disperser solvent, which is consumed in fairly larger volumes compared to an extraction solvent, will make the extraction procedure greener. Ultra-sonication (28), vortex (29), and air agitating (27) of the sample solution are proposed alternatives of the disperser solvent. Although these techniques have been successful, being time-consuming is their main drawback, particularly regarding the case of the air-agitated method.

2. Objectives

In the present study, the elimination of the disperser solvent in the DLLME procedure using a semi-soluble extraction solvent has been accomplished for the first time. Then, the proposed disperser-less DLLME (DL-LLME) technique coupled with GC/MS was exploited to separate and determine three OPPs residues, namely malathion (MT), methyl parathion (MP), and fenitrothion (FT), in various water and fruit juice samples.

3. Methods

3.1. Materials and Apparatus

All selected OPPs were purchased from Fluka, and their stock solutions in methanol (500 mg.L-1) were kept at -20°C. Deionized water was used to prepare the working solutions by step-by-step dilution of the stock solutions. All solvents, including methanol (MeOH), acetonitrile (ACN), acetone (ACT), chloroform (CLF), carbon tetrachloride (CTC), and dichloromethane (DCM), were obtained from Merck (Germany). Heidolph vortex model REAX top (Germany) and KUBOTA Centrifuge Centrifuge Model KN-70 (Japan) were used for mixing and centrifugation of the sample solutions. A Mettler TOLEDO model Seven Compact pH meter (Germany) was used to adjust the pH values.

3.2. Chromatographic Conditions

GC analysis was conducted using the Agilent 7890A instrument (Palo Alto, CA, USA) with an Agilent 5975C mass detector (electron impact (EI) mode with ionization energy of 70 eV). The temperatures of the injection port (in the splitless mode), Ion source, quadruple, and GC/MS interface were set at 250°C, 230°C, 150°C, and 280°C, respectively. An HP-5MS column (30 m × 0.32 mm) with a film thickness of 0.25 µm was used for OPPs separation. The flow rate of carrier gas helium 99.999% was 1 mL.min-1. With a solvent delay of 7 min, the total separation time was 22 min. The selected ion monitoring (SIM) mode with the m/z values of 125, 263, and 277 for MT, MP, and FT, respectively, was used for the analysis.

3.3. DL-LLME Experiments

A graphical presentation of the DL-LLME strategy is displayed in Figure 1. For the extraction procedure, a 5.00 mL solution of the OPPs mixture in concentrations ranging from 0.001 to 100 µg.L-1 at pH = 2 was transferred into the conic screw-capped glass tubes. Then, 60 µL aliquots of extraction solvent (CLF) were rapidly injected into these solutions and vortexed for 5 sec to homogenize the sample solution. The obtained cloudy phase solutions were then centrifuged for 3.0 min at 5000 rpm to separate the organic phase (CLF) from the aqueous phase. The volume of the organic phase was obtained using a microsyringe. Afterward, 1 µL of the organic phase was injected into the GC/MS.

Schematic illustration of DL-DLLME procedure: A and B, injection of the extraction solvent into the aqueous sample and formation of the fine droplets of extraction solvent in the aqueous phase; C, vortexing the mixture of extraction solvent and aqueous sample for 5 sec; D, centrifugation of the mixture for 3 min; E and F, removing the aqueous phase using a syringe and obtaining the sedimented organic phase; and G, retracting the sedimented phase and injection into the GC-MS.
Schematic illustration of DL-DLLME procedure: A and B, injection of the extraction solvent into the aqueous sample and formation of the fine droplets of extraction solvent in the aqueous phase; C, vortexing the mixture of extraction solvent and aqueous sample for 5 sec; D, centrifugation of the mixture for 3 min; E and F, removing the aqueous phase using a syringe and obtaining the sedimented organic phase; and G, retracting the sedimented phase and injection into the GC-MS.

3.4. Real Water and Juice Samples

The tap water and the river water samples were collected from the drinking water of Ahvaz City and Karun River (Ahvaz, Khuzestan Province, Iran). Grape and apple juices were provided from local supermarkets in Ahvaz. All real samples were kept at 4°C after filtering by a 0.45 µm membrane filter. Two point five milliliter of real sample was spiked with the appropriate concentration of the standard OPPs and was diluted up to 5 mL with the deionized water. Then, the OPPs were extracted and analyzed according to the optimized procedure. Besides, non-spiked samples were subjected to the proposed method in parallel. Each experiment was carried out in triplicate.

3.5. Extraction Recovery (ER) and Enrichment Factor (EF)

The ratio of OPP concentration in the organic phase (Corg) after liquid microextraction to the initial concentration of the analyte (Caq) in sample the solution was considered as enrichment factor (EF):

Equation 1. EF=CorgC aq

It worth noting that Corg can be calculated via a standard curve, which is constructed by direct analysis of the OPPs standard solutions in CLF (0.1 - 250 µg.L-1) using GC/MS.

The percentage of the total moles of the extracted OPPs into the organic phase (norg) is called ER. This term is related to the EF via the following equation:

Equation 2.ER = Corg ×VorgCaq × Vaq ×100 = EF ×VorgVaq ×100 (2)

4. Results and Discussion

4.1. The Optimization of DL-LLME Operation Conditions

To obtain more EFs and ERs, some extraction parameters that can influence the performance of the microextraction, including pH, ionic strength, extraction solvent (type and volume), presence of disperser solvent, and centrifugation time, were investigated in details.

4.1.1. Effect of pH

The influence of the solution pH on the microextraction performance was examined in the range of 2 - 10 (5 mL of OPPs solution, 10 µg.L-1; the volume of CLF, 50 µL; 5 min centrifugation time, and without the use of disperser solvent). As shown in Figure 2A, the best value for extraction efficiency is achieved at pH = 2. The reason might be due to the existence of high concentrations of ions (H+ and Cl- ions), which increases the ionic strength of the solution leading to the extraction of target analytes into the organic extractant solvent.

Effects of: A, pH; B, type of extraction solvent; and C, volume of extraction solvent
Effects of: A, pH; B, type of extraction solvent; and C, volume of extraction solvent

4.1.2. Extraction Solvent: Type and Volume

An appropriate extraction solvent should have some particular properties including high density, low solubility in aqueous solution, and extraction capability for the analyte. Three solvents (namely DCM, CLF (relatively low polar solvents), and CTC (nonpolar solvent)) with high densities were chosen and compared with each other using DL-LLME (5 mL of OPPs solution, 10 µg.L-1; the volume of extraction solvent, 50 µL; pH = 2; 5 min centrifugation time, and without the use of disperser solvent). As seen in Figure 2B, the best performance was obtained with CLF. It possesses polarity index comparable with DCM (polarity index sequence: CLF (4.1) > DCM (3.5) > CTC (1.56)). However, the partial solubility of CLF (8.1 g.L-1 in 20°C) is significantly lower than DCM with 17.5 g.L-1 solubility in water, and thus, it can be dispersed in the solution, and good performance is achieved in the DL-LLME process. Cases with both more soluble solvent (DCM) and lower soluble solvent (CTC with 0.8 g.L-1 solubility) are accompanied by decreased extraction efficiency. From this observation, it can be concluded that the solubility of the extraction solvent is a more critical parameter in the DL-LLME procedure. The reason can be attributed to the lower tendency of OPPs for extraction in the presence of more soluble solvent (DCM) because of increased miscibility of the solvent with the sample solution. On the other hand, the presence of nearly insoluble solvents like CTC is associated with the immiscibility of the sample solution as well as the decreased volume of the solvent. To optimize CLF volume, further experiments were performed with various volumes (µL) within 50 - 100 range. As shown in Figure 2C, the results revealed that the best microextraction result was achieved with 60 µL CLF and this optimum volume of extraction solvent was used in subsequent experiments. In volumes lower than 50 µL, the obtained organic phase was too little. After applying larger volumes of the extractant, analytical signals were decreased, mainly due to the decreased concentrations of OPPs in the sedimented phase.

4.1.3. Presence of the Disperser on Microextraction Efficiency

The addition of 30 µL disperser solvent and its effects on the extraction performance were studied using MET, can, and ACT (as disperser) according to the optimized conditions. As displayed in Figure 3A, the best extraction efficiency was obtained in the absence of the disperser. The observations indicated that the extraction performance was considerably decreased in the presence of disperser solvent even at low volumes. This observation might be due to the decreased aqueous solution polarity. As a result, administering larger volumes of disperser (i.e., > 30 µL) was associated with considerable declines in the volume of the sedimented organic phase, which can be attributed to the increased solubility of CLF in this solution, which possesses a polarity lower than that of the solution without disperser. Consequently, when the extraction solvent contains suitable partial solubility in the aqueous sample solution, the presence of another reagent as disperser solvent causes an adverse effect on extraction efficiency, and therefore, it should be removed from the DLLME procedure.

Effects of: A, Amount of salt; B, centrifugation time; and C, type of disperser
Effects of: A, Amount of salt; B, centrifugation time; and C, type of disperser

4.1.4. Salt Addition

In some DLLME procedures, adding salt to the sample solution may enhance the efficiency of the extraction. To test the salt effect on the microextraction performance, various concentrations of NaCl in the range of 0% - 10% (w/v) were used in the DL-LLME procedure. As seen in Figure 3B, the salt addition decreased the analytical signals of OPPs. These observations can be possibly due to the increased viscosity of the sample solution and the subsequent reduction in the diffusion rate of OPPs from the aqueous solution into the organic phase (27, 30). According to the results, the subsequent experiments were carried out without adding salt (31).

4.1.5. The Centrifugation Time Effect

After injecting the extraction solvent, the mixture should be centrifuged to sediment the organic phase. Therefore, we investigated the effect of various centrifugation times (3 to 10 min) at 5000 rpm. As shown in Figure 3C, centrifugation times longer than 3 min were associated with decreased extraction efficiency as well as lower analytical signals. Moreover, in centrifugation times longer than 3 min, the sedimented organic phase volume was slightly reduced, possibly due to the higher dissolution of the organic phase in the aqueous phase in longer durations.

4.2. Method Validation

To validate the DL-LLME method, some important characteristics, including limits of quantification (LOQ) and detection (LOD), precision, accuracy, ER, EF, and linear dynamic range (LDR), were considered. The calibration curves were constructed by analyzing standard OPPs solutions at selected levels ranging from 0.001 to 100 µg.L-1. The regression coefficients (R2) were used to evaluate the linearity goodness of the standard curves. The LODs (0.0003-0.001 µg.L-1) and LOQs (0.001 - 0.003 µg.L-1) were obtained using signals of known concentrations of OPPs solutions based on signal to noise ratios of 3 and 10, respectively. Two LDRs were achieved for all selected OPPs in this method, and good linearity was observed with correlation coefficients greater than 0.9970. The features of the analytical strategy are provided in Table 1. To obtain the precision and accuracy of values, the quality control (QC) samples (0.05, 5, and 50 µg.L-1) were analyzed in five runs (on the same day) and on three consecutive days. The results (Table 2) indicated the good reproducibility of the method with the relative standard deviations (RSD %) in the ranges of 2.5% - 9.3% (intra-day precision) and 3.5% - 10.0% (inter-day precision).

Table 1.

Analytical Features for Determination of OPPs by the Current Methoda, b

PesticideLODLOQLDR1AR12BLDR2CR22DEFEERF
Fenitrothion0.00030.00100.001 - 0.10.99920.1 - 1000.9977173 ± 1769 ± 5
Methyl parathion0.00100.00300.003 - 0.10.99860.1 - 1000.9970238 ± 1095 ± 4
Malathion0.00100.00300.003 - 10.99971 - 1000.9978153 ± 2461 ± 6
Table 2.

Relative Recoveries and RSDs% for QC Samples on Three Spiked Standard Levels of OPPs in Deionized Watera

sAddedAFoundBRecoveryCRSD%
Intra-day (N = 5)Inter-day (N = 3)
Fenitrothion0.0500.04590.07.49.1
5.0004.73094.63.25.4
50.00048.83497.72.53.5
Methyl parathion0.0500.04692.09.310.0
5.0005.231104.64.86.1
50.00051.720103.43.44.1
Malathion0.0500.04488.08.49.5
5.0004.67993.64.16.7
50.00049.10298.23.14.9

4.3. Real samples Analysis

To assess the capability of the proposed method to extract and determine the OPPs in real matrices, the water and juice samples were analyzed by the optimized analytical procedure. Matrix effect can be used to compare the response of the detector signal for the analytes in real and solvent sample solutions (26). At first, the matrix-matched calibration curves were obtained in order to attain accurate results and determining the extent of matrix effects. These curves were constructed using the solutions containing the real samples fortified with standard OPPs, and their analytical features were compared to external calibration curves. According to the findings, their slopes were significantly different (> ± 20%). Therefore, the concentrations of OPPs in the spiked real sample solutions were obtained from the matrix-matched calibration curves. Afterward, the water and juice samples were spiked with OPPs at the levels of 0.5, 5, and 50 µg.L-1 to calculate the recovery data and to test the reliability of the method. The results of the analysis of the real sample (after subjection to the optimized procedure) are provided in Table 3. As shown, the recoveries (84% - 108 %) are in the acceptable range, and it can be concluded that the developed method can be successfully exploited for the real samples analysis with suitable accuracy (-20% to +20%). The typical GC/MS chromatograms obtained from water and juice samples (spiked with 5 µg.L-1 of each OPP) are illustrated in Figure 4.

Table 3.

Investigation of Matrix Effect in the Spiked Real Samples with the Proposed Method (All Real Samples Were Diluted with Deionized Water at a Ratio of 1:2)a

SampleAddedAFenitrothionMethyl ParathionMalathion
FoundBRecovery, %CFoundRecovery, %FoundRecovery, %
River water01.26-1.21-0.38-
0.500.4692.00.4794.00.54108.0
5.005.35106.44.8997.84.7294.4
50.0051.49103.048.4796.949.7699.5
Tap water00.12-0.12-0.07-
0.500.4284.00.4386.00.4794.0
5.005.15103.05.11102.24.9198.2
50.0050.54101.148.3396.6651.34102.7
Grape juice00.14-0.08-0.24-
0.500.4794.30.54107.40.4589.8
5.005.25104.94.8697.24.7695.2
50.0049.6199.249.3598.748.7197.4
Apple juice00.61-1.56-0.99-
0.500.54107.30.53105.60.4691.7
5.004.8897.64.7294.35.22104.3
50.0049.7899.649.7799.548.3096.5
The typical chromatograms of: A River water; B, apple juice; C, grape juice, and tap water samples after being spiked with OPPs at 5 µg.L-1.
The typical chromatograms of: A River water; B, apple juice; C, grape juice, and tap water samples after being spiked with OPPs at 5 µg.L-1.

4.4. Comparison of DL-LLME with the Other Techniques

The analytical performances of the developed DL-LLME strategy in comparison with some published DLLME techniques for the extraction and analysis of OPPs are presented in Table 4. As mentioned before, in the present study, the LOQs, LODs, and RSDs % were better or comparable to the currently available alternatives.

Table 4.

Comparison of the Current Method with the Some Reported Studies in OPPs Detection

MethodMatrixLDR, µg.L-1LOD, ng.L-1Extraction time, minRecovery, %References
DLLME-SFO/HPLC-UV aSummer crops5 - 800b1000 - 4000b1088 - 110(20)
NHS-DLLME/GC-MScWater-3.3 - 8.01050 - 91(25)
ET-DLLME/GC-FIDdWater-juice2 - 10000.82 - 2.72> 975 - 99(4)
UA-DLLME/S-MEKCeMedicinal plants25 - 2500b2000 - 8000b> 683 - 122(2)
DLLME/UPLC-MS/MSfWine and beer0.1 - 1000.48 - 18.8476 - 122(1)
DLLME-GC/MSFruit, vegetables and herbal medicines0.1 - 10000.02 - 0.23> 970 - 119(21)
DL-DLLME-GC/MSWater and juice0.001 - 1000. 3 - 1~384 - 108Present work

5. Conclusions

The current study presented a simple, rapid, and environment-friendly strategy with satisfactory EFs and ERs, namely DL-LLME. A highly sensitive and selective methodology was obtained by the coupling of the DL-LLME procedure by the GC/MS method. The analytical features of the presented microextraction procedure, such as LOQs, LODs, and RSDs % were found to be comparable or better than the other strategies for the analysis of OPPs. As seen, for this work, the exclusion of disperser solvent and too short extraction time (< 4 min) are two main advantages of this method over the other compared protocols, which could significantly improve EFs, rapidity, and greenness of the process.

References

  • 1.

    Pang L, Yang H, Yang P, Zhang H, Zhao J. Trace determination of organophosphate esters in white wine, red wine, and beer samples using dispersive liquid-liquid microextraction combined with ultra-high-performance liquid chromatography-tandem mass spectrometry. Food Chem. 2017;229:445-51. [PubMed ID: 28372199]. https://doi.org/10.1016/j.foodchem.2017.02.103.

  • 2.

    Wei JC, Hu J, Cao JL, Wan JB, He CW, Hu YJ, et al. Sensitive Detection of Organophosphorus Pesticides in Medicinal Plants Using Ultrasound-Assisted Dispersive Liquid-Liquid Microextraction Combined with Sweeping Micellar Electrokinetic Chromatography. J Agric Food Chem. 2016;64(4):932-40. [PubMed ID: 26758524]. https://doi.org/10.1021/acs.jafc.5b05369.

  • 3.

    Zare M, Ramezani Z, Rahbar N. Development of zirconia nanoparticles-decorated calcium alginate hydrogel fibers for extraction of organophosphorous pesticides from water and juice samples: Facile synthesis and application with elimination of matrix effects. J Chromatogr A. 2016;1473:28-37. [PubMed ID: 27810103]. https://doi.org/10.1016/j.chroma.2016.10.071.

  • 4.

    Farajzadeh MA, Afshar Mogaddam MR, Rezaee Aghdam S, Nouri N, Bamorrowat M. Application of elevated temperature-dispersive liquid-liquid microextraction for determination of organophosphorus pesticides residues in aqueous samples followed by gas chromatography-flame ionization detection. Food Chem. 2016;212:198-204. [PubMed ID: 27374524]. https://doi.org/10.1016/j.foodchem.2016.05.157.

  • 5.

    Farajzadeh MA, Bahram M, Vardast MR, Bamorowat M. Dispersive liquid-liquid microextraction for the analysis of three organophosphorus pesticides in real samples by high performance liquid chromatography-ultraviolet detection and its optimization by experimental design. Microchimica Acta. 2010;172(3-4):465-70. https://doi.org/10.1007/s00604-010-0451-9.

  • 6.

    Song D, Wang Y, Lu X, Gao Y, Li Y, Gao F. Ag nanoparticles-decorated nitrogen-fluorine co-doped monolayer MoS2 nanosheet for highly sensitive electrochemical sensing of organophosphorus pesticides. Sensors Actuators B. 2018;267:5-13. https://doi.org/10.1016/j.snb.2018.04.016.

  • 7.

    Zhang P, Sun T, Rong S, Zeng D, Yu H, Zhang Z, et al. A sensitive amperometric AChE-biosensor for organophosphate pesticides detection based on conjugated polymer and Ag-rGO-NH2 nanocomposite. Bioelectrochemistry. 2019;127:163-70. [PubMed ID: 30831354]. https://doi.org/10.1016/j.bioelechem.2019.02.003.

  • 8.

    Liang B, Wang G, Yan L, Ren H, Feng R, Xiong Z, et al. Functional cell surface displaying of acetylcholinesterase for spectrophotometric sensing organophosphate pesticide. Sensors Actuators B. 2019;279:483-9. https://doi.org/10.1016/j.snb.2018.09.119.

  • 9.

    Satnami ML, Korram J, Nagwanshi R, Vaishanav SK, Karbhal I, Dewangan HK, et al. Gold nanoprobe for inhibition and reactivation of acetylcholinesterase: An application to detection of organophosphorus pesticides. Sensors Actuators B. 2018;267:155-64. https://doi.org/10.1016/j.snb.2018.03.181.

  • 10.

    Wu X, Song Y, Yan X, Zhu C, Ma Y, Du D, et al. Carbon quantum dots as fluorescence resonance energy transfer sensors for organophosphate pesticides determination. Biosens Bioelectron. 2017;94:292-7. [PubMed ID: 28315592]. https://doi.org/10.1016/j.bios.2017.03.010.

  • 11.

    Sahub C, Tuntulani T, Nhujak T, Tomapatanaget B. Effective biosensor based on graphene quantum dots via enzymatic reaction for directly photoluminescence detection of organophosphate pesticide. Sensors Actuators B. 2018;258:88-97. https://doi.org/10.1016/j.snb.2017.11.072.

  • 12.

    Filippou O, Bitas D, Samanidou V. Green approaches in sample preparation of bioanalytical samples prior to chromatographic analysis. J Chromatogr B Analyt Technol Biomed Life Sci. 2017;1043:44-62. [PubMed ID: 27623110]. https://doi.org/10.1016/j.jchromb.2016.08.040.

  • 13.

    Wang YL, Gao YL, Wang PP, Shang H, Pan SY, Li XJ. Sol-gel molecularly imprinted polymer for selective solid phase microextraction of organophosphorous pesticides. Talanta. 2013;115:920-7. [PubMed ID: 24054683]. https://doi.org/10.1016/j.talanta.2013.06.056.

  • 14.

    March JG, Cerda V. A novel procedure for phase separation in dispersive liquid-liquid microextraction based on solidification of the aqueous phase. Talanta. 2016;156-157:204-8. [PubMed ID: 27260454]. https://doi.org/10.1016/j.talanta.2016.05.027.

  • 15.

    Wang P, Xiao Y, Liu W, Wang J, Yang Y. Vortex-assisted hollow fibre liquid-phase microextraction technique combined with high performance liquid chromatography-diode array detection for the determination of oestrogens in milk samples. Food Chem. 2015;172:385-90. [PubMed ID: 25442568]. https://doi.org/10.1016/j.foodchem.2014.09.092.

  • 16.

    An J, Rahn KL, Anderson JL. Headspace single drop microextraction versus dispersive liquid-liquid microextraction using magnetic ionic liquid extraction solvents. Talanta. 2017;167:268-78. [PubMed ID: 28340720]. https://doi.org/10.1016/j.talanta.2017.01.079.

  • 17.

    Mohamadjafari S, Rastegarzadeh S. A sensing colorimetric method based on in situ formation of gold nanoparticles after dispersive liquid-liquid microextraction for determination of zineb. Microchem J. 2017;132:154-60. https://doi.org/10.1016/j.microc.2017.01.021.

  • 18.

    Berijani S, Assadi Y, Anbia M, Milani Hosseini MR, Aghaee E. Dispersive liquid-liquid microextraction combined with gas chromatography-flame photometric detection. Very simple, rapid and sensitive method for the determination of organophosphorus pesticides in water. J Chromatogr A. 2006;1123(1):1-9. [PubMed ID: 16716329]. https://doi.org/10.1016/j.chroma.2006.05.010.

  • 19.

    Rezaee M, Assadi Y, Milani Hosseini MR, Aghaee E, Ahmadi F, Berijani S. Determination of organic compounds in water using dispersive liquid-liquid microextraction. J Chromatogr A. 2006;1116(1-2):1-9. [PubMed ID: 16574135]. https://doi.org/10.1016/j.chroma.2006.03.007.

  • 20.

    Pirsaheb M, Fattahi N, Shamsipur M. Determination of organophosphorous pesticides in summer crops using ultrasound-assisted solvent extraction followed by dispersive liquid–liquid microextraction based on the solidification of floating organic drop. Food Control. 2013;34(2):378-85. https://doi.org/10.1016/j.foodcont.2013.05.013.

  • 21.

    Ho YM, Tsoi YK, Leung KS. Highly sensitive and selective organophosphate screening in twelve commodities of fruits, vegetables and herbal medicines by dispersive liquid-liquid microextraction. Anal Chim Acta. 2013;775:58-66. [PubMed ID: 23601975]. https://doi.org/10.1016/j.aca.2013.02.043.

  • 22.

    Alves AC, Goncalves MM, Bernardo MM, Mendes BS. Validated dispersive liquid-liquid microextraction for analysis of organophosphorous pesticides in water. J Sep Sci. 2011;34(11):1326-32. [PubMed ID: 21538877]. https://doi.org/10.1002/jssc.201100111.

  • 23.

    Yang Z, Liu Y, Liu D, Zhou Z. Determination of organophosphorus pesticides in soil by dispersive liquid-liquid microextraction and gas chromatography. J Chromatogr Sci. 2012;50(1):15-20. [PubMed ID: 22291051]. [PubMed Central ID: PMC3252121]. https://doi.org/10.1093/chromsci/bmr011.

  • 24.

    Mohd MS, Siti UM, Mazidatul AM, Wan AWI. Determination of organophosphorus pesticides by dispersive liquid-liquid micro extraction coupled with gas chromatography-electron capture detection. Malays J Anal Sci. 2011;15(2):232-9.

  • 25.

    Alves AC, Goncalves MM, Bernardo MM, Mendes BS. Dispersive liquid-liquid microextraction of organophosphorous pesticides using nonhalogenated solvents. J Sep Sci. 2012;35(19):2653-8. [PubMed ID: 22815100]. https://doi.org/10.1002/jssc.201200241.

  • 26.

    Szarka A, Turkova D, Hrouzkova S. Dispersive liquid-liquid microextraction followed by gas chromatography-mass spectrometry for the determination of pesticide residues in nutraceutical drops. J Chromatogr A. 2018;1570:126-34. [PubMed ID: 30119974]. https://doi.org/10.1016/j.chroma.2018.07.072.

  • 27.

    Farajzadeh MA, Mogaddam MR, Aghdam AA. Comparison of air-agitated liquid-liquid microextraction technique and conventional dispersive liquid-liquid micro-extraction for determination of triazole pesticides in aqueous samples by gas chromatography with flame ionization detection. J Chromatogr A. 2013;1300:70-8. [PubMed ID: 23473511]. https://doi.org/10.1016/j.chroma.2013.02.033.

  • 28.

    Wang K, Xie X, Zhang Y, Huang Y, Zhou S, Zhang W, et al. Combination of microwave-assisted extraction and ultrasonic-assisted dispersive liquid-liquid microextraction for separation and enrichment of pyrethroids residues in Litchi fruit prior to HPLC determination. Food Chem. 2018;240:1233-42. [PubMed ID: 28946248]. https://doi.org/10.1016/j.foodchem.2017.08.061.

  • 29.

    Zaruba S, Bozóová V, Vishnikin AB, Bazeľ YR, Šandrejová J, Gavazov K, et al. Vortex-assisted liquid-liquid microextraction procedure for iodine speciation in water samples. Microchem J. 2017;132:59-68. https://doi.org/10.1016/j.microc.2017.01.004.

  • 30.

    Farajzadeh MA, Khorram P, Nabil AA. Solid-based disperser liquid-liquid microextraction for the preconcentration of phthalate esters and di-(2-ethylhexyl) adipate followed by gas chromatography with flame ionization detection or mass spectrometry. J Sep Sci. 2014;37(9-10):1177-84. [PubMed ID: 24616409]. https://doi.org/10.1002/jssc.201301306.

  • 31.

    Rahbar N, Behrouz E, Ramezani Z. One-Step Synthesis of Zirconia and Magnetite Nanocomposite Immobilized Chitosan for Micro-Solid-Phase Extraction of Organophosphorous Pesticides from Juice and Water Samples Prior to Gas Chromatography/Mass Spectroscopy. Food Anal Methods. 2017;10(7):2229-40. https://doi.org/10.1007/s12161-016-0769-y.