Cover image of the article: Evaluation of time-stability response of gelatin-Trimesic acid dosimeter

Image Credit:Koomesh

Evaluation of time-stability response of gelatin-Trimesic acid dosimeter

Authors

Zohreh Hoseinpour Yektaei, Ramzan Ali Ramzanpour, Maryam Rahbar, Hadi Hasanzadeh,*
*Corresponding Author: Email: hasanzadeh.h@semums.ac.ir

Koomesh:Vol. 18, issue 1; 159-164
Published online:Sep 28, 2016
Article type:Research Article
Received:Feb 07, 2016
Accepted:Aug 02, 2016
How to Cite:Hoseinpour Yektaei Z, Ramzanpour RA, Rahbar M, Hasanzadeh H. Evaluation of time-stability response of gelatin-Trimesic acid dosimeter. koomesh. 2016;18(1):e151157. doi:

Abstract

Introduction: One of the problems in chemical dosimeters such as Trimesic acid is the liquid environment which makes it difficult to use. Chemical dosimeter merged into a bed of gel can help to partly overcome this problem.According to the importance of stability of dosimeter response after irradiation, in this study, the parameters of gelatin- Trimesic acid dosimeter in the range of doses of a one fraction of radiotherapy was assessed in days one, two, three and four after irradiation. Materials and Methods: After the gel construction at a concentration of 1 mM and a pH of 2.2, it was poured into Perspex vials which opening was tight with parafilm and wrapped in an aluminum foil to eliminate light and was divided into 5 groups to be irradiated with 6 MV photons of linac 24h after the gel synthesis. To do so, vials were place at the isocenter of the linac in a SSD of 100cm and field size of 10×10 cm2. The delivered doses were from 0-200 cGy and vials were kept at the temperature of 4˚C after the irradiation. The procedure of reading the vials were performed using a spectroflourimeter (Jasco 6200) from one to four days after irradiation in excitation wavelengths of 370, 380, 390, 410, 420 and 450 nm. Results: After obtaining the emission spectrum from the spectroflourimeter, the emission spectrum was analyzed and it was found that there is a peak in the wavelengths of 377, 387, 396, 416, 427 and 457 nm seen in one day after irradiation and the most stable response belongs to the first day after irradiation. Conclusion: The results of this study showed that the combined gelatin- Trimesic acid dosimeter in the first day after irradiation has the highest and most stable response and according to its characteristics, has the potential to be assessed as a clinical dosimeter

Highlights

References

  • 1.
    Attix FH. Introduction to radiological physics and radiation dosimetry: John Wiley & Sons; 2008.
  • 2.
    Butson MJ, Yu PK, Cheung T, Metcalfe P. Radiochromic film for medical radiation dosimetry. Mater Sci Eng Rep 2003; 41: 61-120.
  • 3.
    Knoll GF. Radiation detection and measurement: John Wiley & Sons; 2010.
  • 4.
    Gelyan HN, Makhdumi Y, Nikoofar A, Hasanzadeh H. Measurement of surface dose in external radiotherapy of brain frontal lobe: A study on patient and phantom. Koomesh 2016; 17: 323-328. (Persian).
  • 5.
    Nikoofar A, Hoseinpour Z, Mahdavi SR, Hasanzadeh H, Tavirani MR. High-dose-rate 192Ir brachytherapy dose verification: A phantom study. Iran J Cancer Prev 2015; 8: e2330.
  • 6.
    Rafati M, Eftekhari-Moghadam A, Mehri-Kakavand G, Hasanzadeh H, Maftoon A. Dosimetry of organ at risks in orthopantomography. Koomesh 2015; 17: 262-266. (Persian).
  • 7.
    Nazemi-Gelyan H, Hasanzadeh H, Makhdumi Y, Abdollahi S, Akbari F, Varshoee-Tabrizi F, et al. Evaluation of organs at risk's dose in external radiotherapy of brain tumors. Iran J Cancer Prev 2015; 8: 47-52.
  • 8.
    Hasanzadeh H, Sharafi A, Allah Verdi M, Nikoofar A. Assessment of absorbed dose to thyroid, parotid and ovaries in patients undergoing Gamma Knife radiosurgery. Phys Med Biol 2006; 51: 4375-4383.
  • 9.
    Matthews RW, Barker NT, Sangster DF. A comparison of some aqueous chemical dosimeters for absorbed doses of less than 1000 rads. Int J Appl Radiat Isot 1978; 29: 1-8.
  • 10.
    Khan H, Bhatti I. Aqueous phenylacetic acid as a low-dose fluorescence dosimeter. J Nucl Chem 1995; 199: 385-393.
  • 11.
    Khan HM, McLaughlin WL. Gamma-ray dosimetry by spectrofluorimetry of phenylacetic acid solution. Int J Radiat Appl Instrum C 1992; 39: 243-249.
  • 12.
    Matthews RW. Aqueous chemical dosimetry. Int J Appl Radiat Isot 1982; 33: 1159-1170.
  • 13.
    Armstrong W, Facey R, Grant D, Humphreys W. A tissue-equivalent chemical dosimeter sensitive to 1 rad. Can J Chem 1963; 41: 1575-1577.
  • 14.
    Semwal MK, Bansal AK, Thakur PK, Vidyasagar PB. FBX aqueous chemical dosimeter for measurement of virtual wedge profiles. J Appl Clin Med Phys 2008; 9: 2809.
  • 15.
    Serwer P. Agarose gels: Properties and use for electrophoresis. Electrophoresis 1983; 4: 375-382.
  • 16.
    Wikipedia the free encyclopedia. Agarose gel Properties. 2013.
  • 17.
    Audet C, Schreiner L. Multiple-site fast exchange model for spinlattice relaxation in the Fricke-gelatin dosimeter. Med Phys 1997; 24: 201-209.
  • 18.
    Seybold C. Polyacrylamide review: Soil conditioning and environmental fate. Commun Soil Sci Plant Anal 1994; 25: 2171-2185.
  • 19.
    Smith EA, Oehme FW. Acrylamide and polyacrylamide: a review of production, use, environmental fate and neurotoxicity. Rev Environ Health 1991; 9: 215-228.
  • 20.
    Armisen R, Galatas F. Production, properties and uses of agar. FAO Fisheries Technical Paper (FAO) 1987.
  • 21.
    Wikipedia the free encyclopedia. Acrylamide. From Wikipedia the free encyclopedia 2012.
  • 22.
    Gelatin Manufacturers Institute of America. America: GMIA; 2012.
  • 23.
    by Generic A. Agarose gel electrophoresis.
  • 24.
    Azadbakht B, Adinehvand K. Investigation of the post time dependence of PAGAT gel dosimeter by electron beams using MRI technique. Res J Appl Sci 2012; 4: 232-235.
  • 25.
    Matthews RW. Aqueous chemical dosimetry. Int J Appl Rad Isotop 1982; 33: 1159-1170.
  • 26.
    Matthews RW, Wilson JG. Chemical dosimetry at less than 1000 rad: Aqueous trimesic acid solutions. Int J Appl Rad Isotop 1981; 32: 295-301.
  • 27.
    Gupta B, Gomathy K. Consistency of ferrous sulphate-benzoic acid-xylenol orange dosimeter. Int J Appl Rad Isotop 1974; 25: 509-513.
  • 28.
    Baldock C, Burford R, Billingham N, Wagner G, Patval S, Badawi R, Keevil S. Experimental procedure for the manufacture and calibration of polyacrylamide gel (PAG) for magnetic resonance imaging (MRI) radiation dosimetry. Phys Med Biol 1998; 43: 695.
  • 29.
    Smith E, Prues S, Oehme F. Environmental degradation of polyacrylamides. 2. Effects of environmental (outdoor) exposure. Ecotoxicol Environ Saf 1997; 37: 76-91.##.

Copyright

© 2016, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.

Similar Articles

20
Nov
2010
Determination of Residual Methylmethacrylate Monomerin Denture Base Resins by Gas Chromatography

Determination of Residual Methylmethacrylate Monomerin Denture Base Resins by Gas Chromatography

Khosrou Abdi,
Ali Mandegary,
Mohsen Amini,
Mozhgan Bagheri,
Farideh Gerami-Panah

Abdi K, Mandegary A, Amini M, Bagheri M, Gerami-Panah F. Determination of Residual Methylmethacrylate Monomerin Denture Base Resins by Gas Chromatography. Iran J Pharm Res. 2022;4(4):e128249. doi: https://doi.org/10.22037/ijpr.2010.641

16
Feb
2016
Quince Seed Lyophilized Biopolymer as a Medicinal Remedy: Study of Mechanical Properties of the Powder and Tablets

Quince Seed Lyophilized Biopolymer as a Medicinal Remedy: Study of Mechanical Properties of the Powder and Tablets

Abdolhossein Moghbel,
Maryam Tayebi

Moghbel A, Tayebi M. Quince Seed Lyophilized Biopolymer as a Medicinal Remedy: Study of Mechanical Properties of the Powder and Tablets. Jundishapur J Nat Pharm Prod. 2016;11(1):e59678. doi:

25
May
2014
Chemical and Biological Properties of Trachyspermum ammiEncapsulated in Gelatin Nanofilms

Chemical and Biological Properties of Trachyspermum ammiEncapsulated in Gelatin Nanofilms

Somayeh Jahani,
Gholamreza Kavoosi,
Amin Shakiba

Jahani S, Kavoosi G, Shakiba A. Chemical and Biological Properties of Trachyspermum ammiEncapsulated in Gelatin Nanofilms. Int J Infect. 2014;1(1):e18420. doi: https://doi.org/10.17795/iji-18420

20
Nov
2010
The Influence of Various Silicones on the Rheological Parameters of AZG Containing Silicone-Based Gels

The Influence of Various Silicones on the Rheological Parameters of AZG Containing Silicone-Based Gels

Seyed Alireza Mortazavi,
Hosseinali Tabandeh

Mortazavi SA, Tabandeh H. The Influence of Various Silicones on the Rheological Parameters of AZG Containing Silicone-Based Gels. Iran J Pharm Res. 2022;4(4):e128246. doi: https://doi.org/10.22037/ijpr.2010.638

15
Oct
2013
Design of Intelligent Polymer of Gelatin- Poly N-isopropyla- crylamide under Gamma radiation for Cellular Applications

Design of Intelligent Polymer of Gelatin- Poly N-isopropyla- crylamide under Gamma radiation for Cellular Applications

Saeed Heidari-Keshel,
Bita Soleimani,
Maryam Ebrahimi,
Asghar Ashrafi Hafez,
Razieh Fallahakbarpour,
Mohammad Khaledian
,et al.

Heidari-Keshel S, Soleimani B, Ebrahimi M, Ashrafi Hafez A, Fallahakbarpour R, et al. Design of Intelligent Polymer of Gelatin- Poly N-isopropyla- crylamide under Gamma radiation for Cellular Applications. J Rep Pharm Sci. 2013;2(2):e147751. doi:

More by these authors

Zohreh Hoseinpour YektaeiPubMedScholar
Ramzan Ali RamzanpourPubMedScholar
Maryam RahbarPubMedScholar
Hadi HasanzadehPubMedScholar
Share
Cited by
Metrics