Comparison between the Precision of Measurements in Two Types of the Micropipettes according to CLSI EP5 and ISO 8655

Authors

Hamid Alavi MajdHamid Alavi Majd ORCID,*, Jamal Hosieni, Hosein Tamadon, Alireza AkbarZadeh
*Corresponding Author: Email: alavimajd@gmail.com

Koomesh:Vol. 10, issue 3; 185-190
Published online:Aug 24, 2009
Article type:Research Article
Received:May 02, 2006
Accepted:Jun 10, 2009
How to Cite:Alavi Majd H, Hosieni J, Tamadon H, AkbarZadeh A. Comparison between the Precision of Measurements in Two Types of the Micropipettes according to CLSI EP5 and ISO 8655. koomesh. 2009;10(3):e152238. doi:

Abstract

Background: In the diagnosis medical labs, it is very important to evaluate the precision of micropipettes in transferring small amounts of liquids. The medical tests will not result accurately, if the liquid volume doesn’t transfer exactly by micropipettes. Thus the doctor faces sort of problems in the disease diagnosis and its control. In the standard CLSI EP5, there is a method to specify and assess the precision of micropipettes, by using CV (coefficients of variation). Also there are other methods to estimate and test the mention CV theory, in the formal statistic texts which could be used to assess the micropipettes precision. It is the main goal of this research to study the precision of lab micropipettes which are shown by A and B and compared by reference standard value from ISO 8655. Material and Methods: In this research we evaluate the precision of the lab micropipettes. Two brands A and B are assigned to measure the distilled water mass by using the accurate scale which is accurate up to 10-6 to measure 50 gram weight. The experimental environment is a metrology lab which is confirmed by Iran Standard and Industrial Researches. A unique technician sampled in the beginning of the work time and after 2 hours repeated the sampling. Totally, each micropipette is used to measure 40 times with 10 times repeat for single measurement in 28 work days. It is used common statistical methods to estimate and test the coefficients of variation theory. Results: Point estimation of CV for micropipettes A and B were 0.50% and 0.64% respectively. Also the 95% confidence upper bounds for these two micropipettes by using likelihood ratio method were 0.53% and 0.64% respectively. The micropipette A confirmed the ISO 8655, but the micropipette B did not. Measurement error in micropipette B was 30% less than micropipette A in average. Conclusion: By using the approach of CLS EP5 and confidence interval for CV, precision of two micropipettes were compared. Only one of them confirmed the ISO 8655, but the other one was failed

Highlights

Copyright

© 2009, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.

Similar Articles

18
Apr
2015
Estimating the blood pressure measurement frequency for optimum reliabilit

Estimating the blood pressure measurement frequency for optimum reliabilit

M Karami,
A Biderafsh

Karami M, Biderafsh A. Estimating the blood pressure measurement frequency for optimum reliabilit. J Inflamm Dis. 2024;19(1):e155893. doi:

1
Oct
2013
Accuracy of Sphygmomanometers Measurements in two Iranian Tertiary Educational Hospitals

Accuracy of Sphygmomanometers Measurements in two Iranian Tertiary Educational Hospitals

Alireza Nasrolahi,
Esmat Ghanei,
Golfam Mehrparvar

Nasrolahi A, Ghanei E, Mehrparvar G. Accuracy of Sphygmomanometers Measurements in two Iranian Tertiary Educational Hospitals. Shiraz E-Med J. 2013;14(4):e16655. doi: https://doi.org/10.17795/semj16655

25
Mar
2013
Accuracy Assessment of Measured Blood Pressures in Three Departments of one Tertiary Hospital

Accuracy Assessment of Measured Blood Pressures in Three Departments of one Tertiary Hospital

Alireza Nasrolahi,
Esmat Ghanei,
Masoumeh Moalem

Nasrolahi A, Ghanei E, Moalem M. Accuracy Assessment of Measured Blood Pressures in Three Departments of one Tertiary Hospital. Shiraz E-Med J. 2013;14(1):35-43. doi:

28
Jan
2019
The Principles of Biomedical Scientific Writing: Materials and Methods

The Principles of Biomedical Scientific Writing: Materials and Methods

Asghar Ghasemi,
Zahra Bahadoran,
Azita Zadeh-Vakili,
Seyed Ali Montazeri,
Farhad Hosseinpanah

Ghasemi A, Bahadoran Z, Zadeh-Vakili A, Montazeri SA, Hosseinpanah F. The Principles of Biomedical Scientific Writing: Materials and Methods. Int J Endocrinol Metab. 2019;17(1):e88155. doi: https://doi.org/10.5812/ijem.88155

25
Jun
2003
The assessment of the difference of capillary blood glucose level and venous glucose level measured by glucometer devices

The assessment of the difference of capillary blood glucose level and venous glucose level measured by glucometer devices

MH Bastanhagh,
b Larijani,
Alireza Khalili-Fard,
A Hossein Nezhad,
AR Shafahi

Bastanhagh M, Larijani B, Khalili-Fard A, Hossein Nezhad A, Shafahi A. The assessment of the difference of capillary blood glucose level and venous glucose level measured by glucometer devices. Zahedan J Res Med Sci. 2003;5(2):e95149. doi:

More by these authors

Hamid Alavi MajdPubMedScholar
Jamal HosieniPubMedScholar
Hosein TamadonPubMedScholar
Alireza AkbarZadehPubMedScholar
Share
Cited by
Metrics