Evaluating the blood glucose measured via glucometry in diagnosis of hypoglycemia in neonates

Authors

Shamsollah NoriPor, Farhad Azizzadeh, Navid Daniae, mohammad sadeghyazdiha,*, Raheb GhorbaniRaheb Ghorbani ORCID
*Corresponding Author: Email: dr_msyazdiha@yahoo.com

Koomesh:Vol. 13, issue 4; 440-444
Published online:May 25, 2012
Article type:Research Article
Received:Jul 01, 2011
Accepted:Mar 06, 2012
How to Cite:NoriPor S, Azizzadeh F, Daniae N, sadeghyazdiha M, Ghorbani R. Evaluating the blood glucose measured via glucometry in diagnosis of hypoglycemia in neonates. koomesh. 2012;13(4):e152533. doi:

Abstract

  Introduction: Hypoglycemia is one of the most prevalent problems seen in neonates and can lead to irreversible brain damage if not to be diagnosed in time . Today, the blood glucose is mostly checked by laboratory methods that have a significant delay and can result in many adverse effects such as vessel rupture. The goal of this study was to evaluate the value of the glucometry device in diagnosis of hypoglycemia in neonates.   Materials and Methods: This study has been conducted over 156 hospitalized neonates in Amir al momenin Hospital, Semnan, Iran. The blood glucose in all cases has been simultaneously measured by both laboratory methods (glucose oxidase as the golden standard) and glucometer. These measurements along with the relevant data for neonatal risk factors were recorded in the checklist. The blood glucose measurements of less than or equal 35 mg/dL in first two hours after birth, less than or equal 40mg/dL in 3 to 24 hours after birth, less than or equal 45 mg/dL in more than 24 hours after birth are considered as hypoglycemic.   Results: 58 neonates (37.2%) were hypoglycemic.The measurement of blood glucose levels via glucometer for hypoglycemia detection had a high precision (Area under curve = 0.941, Standard error =0.018, P= 0.0001). The optimum cut-off point yielded to equivalent of 51mg/dL. So that for fasting blood glucose of less than or equal 51mg/dL, sensitivity, specificity, positive, and negative predictive values were, respectively, 94.7%, 81.6%, 75.3%, and 96.4%.   Conclusion: Findings show that, blood glucose levels measured by glucometry have good accuracy for diagnosis of hypoglycemia in newborns. Thus, it is recommended that in case of screening and frequent need of monitoring neonatal blood glucose, this device be used as a suitable replacement of laboratory methods .

Highlights

Copyright

© 2012, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.

Similar Articles

25
Jun
2003
The assessment of the difference of capillary blood glucose level and venous glucose level measured by glucometer devices

The assessment of the difference of capillary blood glucose level and venous glucose level measured by glucometer devices

MH Bastanhagh,
b Larijani,
Alireza Khalili-Fard,
A Hossein Nezhad,
AR Shafahi

Bastanhagh M, Larijani B, Khalili-Fard A, Hossein Nezhad A, Shafahi A. The assessment of the difference of capillary blood glucose level and venous glucose level measured by glucometer devices. Zahedan J Res Med Sci. 2003;5(2):e95149. doi:

31
Jan
2026
Neonatal Glycemic Status in Infants of Diabetic Mothers in Sulaimaniyah Governorate, Iraq

Neonatal Glycemic Status in Infants of Diabetic Mothers in Sulaimaniyah Governorate, Iraq

Basta Jalil Abdalla,
Adnan Muhammad Hasan,
Jamal Mohammed Hussein Lateef

Jalil Abdalla B, Muhammad Hasan A, Mohammed Hussein Lateef J. Neonatal Glycemic Status in Infants of Diabetic Mothers in Sulaimaniyah Governorate, Iraq. Int J Endocrinol Metab. 2026;24(1):e164575. doi: https://doi.org/10.5812/ijem-164575

31
Mar
2012
Correlation between Hypoglycemia and Positive Rate of Inborn Error of Metabolism in Neonatal Intensive Care Unit

Correlation between Hypoglycemia and Positive Rate of Inborn Error of Metabolism in Neonatal Intensive Care Unit

Ying Li,
Jian He,
Wen-jun Tu

Li Y, He J, Tu W. Correlation between Hypoglycemia and Positive Rate of Inborn Error of Metabolism in Neonatal Intensive Care Unit. Inn J Pediatr. 2016;22(1):. doi:

30
Sep
2015
Comparison of the Blood Level OF Leptin in Umbilical Cord of Newborns of Mothers With Gestational Diabetes and Normal Mothers and Its Relationship With Growth Indices of Newborns

Comparison of the Blood Level OF Leptin in Umbilical Cord of Newborns of Mothers With Gestational Diabetes and Normal Mothers and Its Relationship With Growth Indices of Newborns

Mohammad Reza Aramesh,
Mosoud Dehdashtian,
Arash Malekian,
Reza Babapour

Aramesh MR, Dehdashtian M, Malekian A, Babapour R. Comparison of the Blood Level OF Leptin in Umbilical Cord of Newborns of Mothers With Gestational Diabetes and Normal Mothers and Its Relationship With Growth Indices of Newborns. Jentashapir J Cell Mol Biol. 2015;6(5):e25797. doi: https://doi.org/10.17795/jjhr-25797

31
Mar
2013
Comparing the frequency of neonatal hypoglycemia and hypocalcemia in macrosomic and non-macrosomic infants of diabetic mothers

Comparing the frequency of neonatal hypoglycemia and hypocalcemia in macrosomic and non-macrosomic infants of diabetic mothers

M Hematyar,
Z Mirmajidi Hashjin

Hematyar M, Mirmajidi Hashjin Z. Comparing the frequency of neonatal hypoglycemia and hypocalcemia in macrosomic and non-macrosomic infants of diabetic mothers. J Inflamm Dis. 2024;16(4):e155735. doi:

More by these authors

Shamsollah NoriPorPubMedScholar
Farhad AzizzadehPubMedScholar
Navid DaniaePubMedScholar
mohammad sadeghyazdihaPubMedScholar
Raheb GhorbaniPubMedScholar
Share
Cited by
Metrics