Accuracy of Dentaport ZX in length determination of resorbed roots and non-resorbed roots in primary teeth
Authors
Abstract
Background: Since radiography has some problems, Electronic Apex Locator is applied in endodontic treatment for young children. But the accuracy of this equipment in roots with resorption is questionable; therefore, the present study was conducted to Compare the accuracy of Dentaport ZX in length determination of resorbed roots and non resorbed roots in primary teeth.
Materials and Methods: Clinical evaluation of the accuracy of Dentaport ZX was carried out in 52 primary teeth with vital pulp tissue planned for extraction. The length of the canals was estimated with Dentaport ZX after removing pulp tissues. Then, the teeth were extracted and the real lengths of canals were measured by K-files # 15 with a precision of 0.1mm. Data were analyzed using Chi-square and intraclass correlation coefficient tests at the 5% level of significance.
Results: Accuracy of the Dentaport ZX in roots without and with resorption in the range of – 0.5 to +0.5 from apical foramen was 96.1% and 89.1% respectively and in the range of -0.2 to + 0.2 from apical foramen was 84.3% and 53.1% respectively.
Discussion: Results showed that the Dentaport ZX’s accuracy in determining the length of non-resorbed roots is higher than that of resorbed roots
Highlights
References
- 1.The References of this article are available on the PDF file.
Copyright
© 2008, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.
Similar Articles
An in vitro comparison of Foramatron and Raypex5 apex locator accuracy
Khalilkhani H, Azmoudeh F, Salamat F. An in vitro comparison of Foramatron and Raypex5 apex locator accuracy. J Inflamm Dis. 2024;17(1):e155741. doi:
Performance of the Working Length Determination using Cone Beam Computed Tomography, Radiography and Electronic Apex Locator, in Comparisons to Actual Length
Yildirim C, Murat Aktan A, Karataslioglu E, Aksoy F, Isman O, et al. Performance of the Working Length Determination using Cone Beam Computed Tomography, Radiography and Electronic Apex Locator, in Comparisons to Actual Length. I J Radiol. 2017;14(1):e13468. doi: https://doi.org/10.5812/iranjradiol.31007
Ex Vivo Evaluation of the Root Form and Root Canal Morphology of the Mandibular First Molar Us ing CBCT Technology
Ali Saberi E, Farhad-Mollashahi N, Niknami M, Mousavi E, Rasuli H. Ex Vivo Evaluation of the Root Form and Root Canal Morphology of the Mandibular First Molar Us ing CBCT Technology. Zahedan J Res Med Sci. 2014;16(7):. doi:
Root and canal morphology of the mesiobuccal root of maxillary first permanent molars in a part of population of Qazvin
Adel M, Hamzehnejad S. Root and canal morphology of the mesiobuccal root of maxillary first permanent molars in a part of population of Qazvin. J Inflamm Dis. 2024;13(3):e155488. doi:
Prevalence of C-shaped mandibular second molars in Zahedan
Noormandi Poor M, nasiri M. Prevalence of C-shaped mandibular second molars in Zahedan. Zahedan J Res Med Sci. 2007;9(4):e94750. doi:
- Scopus by DOI: 0
Last Update: 20 hours ago
- Scopus by Title: 0
Last Update: 20 hours ago
- Scopus by Title (Ref): 2
Last Update: 20 hours ago
- CrossRef: 0
Last Update: 3 days ago