The forced oscillation technique (FOT) is a non-invasive method of measuring respiratory impedance, which is defined as the spectral relationship between pressure and airflow (
1). Impedance is determined by two components: respiratory resistance (Rrs) and respiratory reactance (Xrs); the latter reflects the elastic and inertial properties of the lung. FOT measurement requires no special breathing maneuvers or interference with normal breathing (
2). The clinical use of FOT has progressed as more FOT devices have become commercially available, such as the MostGraph-01® impulse oscillation system (Chest MI, Tokyo, Japan) (
3). The evidence base for the clinical utility of FOT has expanded, especially for the evaluation and management of obstructive pulmonary diseases. Nonetheless, the changes in respiratory impedance that occur as a result of airway inflammation and pulmonary function are not fully understood. The Rrs measured at 5 Hz (R5) and 20 Hz (R20) are representative of low and high frequency resistances, respectively. The difference between R5 and R20 (R5 - R20) is referred to as the frequency dependence of Rrs. The reactance at 5 Hz (X5) reflects the combined effects of tissue elastance and inertia, although at this lower frequency, the effect of tissue elastance would be predominant. Therefore, X5 reflects the elastic recoil of the peripheral airways. The resonant frequency (Fres) is the frequency at which respiratory reactance equals 0 cmH
2O/L/s. The Fres marks the transition from capacitative dominance at low frequencies to inertial dominance at high frequencies. The area of low reactance (ALX) is the area created by three lines (those of frequency 5 Hz, Xrs = 0 cmH
2O/L/s, and the Xrs curve). The ALX is a useful index associated with respiratory compliance and therefore, reflects the patency of the small airways. The ALX is a single quantity that reflects changes in the degree of peripheral airway obstruction and closely correlates with R5 - R20 (
4). We examined the changes in respiratory impedance after endotracheal intubation or the use of a supraglottic airway device (SGD) in mechanically ventilated patients. We hypothesized that endotracheal intubation would bring about more substantial preoperative increases in respiratory resistance and reactance than using SGD.