Evaluation of the Efficiency of Indirect Blood Pressure Measurement Methods in Comparison to Intra-Arterial Reading among Iranian Individuals
Authors
Abstract
Background:
Detecting the suitable non-invasive Blood Pressure (BP) measurement method is a quandary in clinical settings for accurate diagnosis of Hypertension (HTN) status.
Objectives:
This study aimed to evaluate the efficiency of indirect BP measurement methods in comparison to the gold standard among Iranian patients admitted for Coronary Angiography (CAG).
Methods:
This observational study was conducted on 150 CAG candidates randomly selected using the computerized random numbers from March 2019 to September 2019. The participants' BPs were measured via three different non-invasive methods, including brachial and wrist oscillometric cuffs plus brachial sphygmomanometer, and the results were compared to simultaneous intra-arterial reading as the gold standard. The associations between different non-invasive BP measurements and direct arterial BP reading were assessed using different statistical analyses, including correlation coefficient, chi-square, independent and paired t-test, and Analysis of Variance (ANOVA), as appropriated.
Results:
The mean age of the participants was 60.56 ± 11.16 years. Both Systolic BP (SBP) and Diastolic BP (DBP) were positively correlated to the gold standard reading in all measurement methods (P < 0.001). After adjustment for potential confounders, the findings revealed no significant difference between the pre-defined BP measurement methods, including brachial sphygmomanometer, brachial oscillometric, and wrist oscillometric, and intra-arterial BP reading regarding the means of SBP and DBP (P > 0.05).
Conclusions:
The present study findings showed that the selected solution was appropriate in terms of clinical aspects for the patients undergoing CABG surgery with long surgical duration or low Ejection Fraction (EF) and reduced the costs to half. Considering the significant difference in the CK-MB level and the lower troponin level in the combined group (not statistically significant), further studies are required to confirm the clinical priority of the combined solution.
Highlights
References
- 1.The references are available in pdf.
Copyright
© 2020, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.
Similar Articles
Investigating the Association of Relative Central Blood Pressure Indices and Severity of Coronary Artery Disease
Forouzanmehr PA, Haidary Sardaby E, Hesam S. Investigating the Association of Relative Central Blood Pressure Indices and Severity of Coronary Artery Disease. Jundishapur J Chronic Dis Care. 2019;8(4):e97112. doi: https://doi.org/10.5812/jjcdc.97112
Accuracy Assessment of Measured Blood Pressures in Three Departments of one Tertiary Hospital
Nasrolahi A, Ghanei E, Moalem M. Accuracy Assessment of Measured Blood Pressures in Three Departments of one Tertiary Hospital. Shiraz E-Med J. 2013;14(1):35-43. doi:
Accuracy of Sphygmomanometers Measurements in two Iranian Tertiary Educational Hospitals
Nasrolahi A, Ghanei E, Mehrparvar G. Accuracy of Sphygmomanometers Measurements in two Iranian Tertiary Educational Hospitals. Shiraz E-Med J. 2013;14(4):e16655. doi: https://doi.org/10.17795/semj16655
Comparison Between Continuous Non-invasive Arterial Pressure (CNAP) and Invasive Arterial Pressure Monitoring During Weaning from CPB in Patients Undergoing CABG Surgery: A Comparative Case Study
Jahangirifard A, Zarimeydani N, Daneshmand M, Rafiei K, Mirtajani SB, et al. Comparison Between Continuous Non-invasive Arterial Pressure (CNAP) and Invasive Arterial Pressure Monitoring During Weaning from CPB in Patients Undergoing CABG Surgery: A Comparative Case Study. J Cell Mol Anesth. 2024;9(4):e148317. doi: https://doi.org/10.5812/jcma-148317
The Frequency of Errors of Blood Pressure Measurement among Nurses in the Hospitals Affiliated to Shiraz University of Medical Sciences, 2014
Dokoohaki R, Raeiskarimian F, Rahgosha A, Sharifi M. The Frequency of Errors of Blood Pressure Measurement among Nurses in the Hospitals Affiliated to Shiraz University of Medical Sciences, 2014. Int Cardiovasc Res J. 2017;9(1):e11664. doi:
- Scopus by DOI: 0
Last Update: 1 month ago
- Scopus by Title: 8
Last Update: 1 month ago
- Scopus by Title (Ref): 8
Last Update: 1 month ago
- CrossRef: 0
Last Update: 2 hours ago
- OpenAlex: 0
Last Update: 6 hours ago
- OpenCitations: 0
Last Update: 4 days ago