Evaluation of 4D-NCAT Phantom to simulate Dynamic Behavior of Radiopharmaceutical in Myocardium

Author(s):
Zohreh ShahpouriZohreh Shahpouri, Ahmad bitarafanAhmad bitarafan,*, Alireza Kamali AslAlireza Kamali Asl, Hadi HasanzadehHadi Hasanzadeh
*Corresponding Author: Email: [email protected]

Koomesh:Vol. 17, issue 1; 250-254
Published online:Dec 28, 2015
Article type:Research Article
Received:Jan 03, 2015
Accepted:Aug 02, 2015
How to Cite:Shahpouri Z, bitarafan A, Kamali Asl A, Hasanzadeh H. Evaluation of 4D-NCAT Phantom to simulate Dynamic Behavior of Radiopharmaceutical in Myocardium. koomesh. 2015;17(1):e150798. doi:

Abstract

 Introduction: Studying the heart tissue kinetics can lead to expedient diagnostic technique to study the spatial and temporal cardiac activity simultaneously. Simulations of this process requires phantom, which applies dynamic activity changes gradually to the regions of interest. The objective of this study was to validate the efficiency of 4D-NCAT phantom data in the dynamic activity changes in course of the time.  Materials and Methods: The accuracy of constant, ascending and descending steep Time activity Curves (TACs) in blood and myocardial tissue were measured after i99m-Tc teboroxime injection. Also, the accuracy of phantom output data was measured in different resolution.  Results: Our study showed that the response of the dynamic 4D NCAT phantom for constant curves had just some slight errors. The average relative differences between input and output data for steeper TACs were noteworthy in both ascending and descending mode. The results showed that by increasing the temporal data sampling, the average difference between input and outputs were decreased. Overall, only 0.6% of the assessed samples depicted relative errors larger than 0.1%. Conclusion: The primary studies with available 4D NCAT software, version 2012, revealed some challenges in applying intended activity with TACs, which referred to and were subsequently resolved by Dr. Paul Segars revisions. Overall, the modified 4D NCAT phantom software is able to accurately simulate the organ dynamics. 

Copyright

© 2015, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.

Similar Articles

10
Aug
2016

Comparison of Reconstruction Methods in a Small Animal Cardiac Positron Emission Tomography Study Using a 18F-Labeled Myocardial Agent, [18F] FPTP

Hyeon-Sik Kim,
Byeong-il Lee,
Ju Han Kim,
Hee-Seung Bom,
Dong-Yeon Kim,
Jung-Joon Min

Kim H, Lee B, Han Kim J, Bom H, Kim D, et al. Comparison of Reconstruction Methods in a Small Animal Cardiac Positron Emission Tomography Study Using a 18F-Labeled Myocardial Agent, [18F] FPTP. I J Radiol. 2017;14(1):e13463. doi: https://doi.org/10.5812/iranjradiol.26806

29
Jan
2014

Automatic Bolus Tracking Versus Fixed Time-Delay Technique in Biphasic Multidetector Computed Tomography of the Abdomen

Atoosa Adibi,
Ali Shahbazi

Adibi A, Shahbazi A. Automatic Bolus Tracking Versus Fixed Time-Delay Technique in Biphasic Multidetector Computed Tomography of the Abdomen. I J Radiol. 2014;11(1):4617. doi: https://doi.org/10.5812/iranjradiol.4617

20
Aug
2016

Radiation Dose and Image Quality Assessment of the Bolus Timing Method for CT Angiography

Myeong Seong Kim,
Jong-Woong Lee,
Sun Geun Kim,
Dae Cheol Kweon

Seong Kim M, Lee J, Geun Kim S, Cheol Kweon D. Radiation Dose and Image Quality Assessment of the Bolus Timing Method for CT Angiography. I J Radiol. 2017;14(2):e31918. doi: https://doi.org/10.5812/iranjradiol.31918

10
Feb
2023
Iran J Radiol

Role of Lesion-to-Background Activity Ratio and Background Activity in Optimization of Reconstruction Protocols for FDG PET/CT Images of Overweight Patients: A Phantom Study

Samira Rezvani,
Pardis Ghafarian,
Mehrdad Bakhshayesh Karam,
Mohammadreza Ay

Rezvani S, Ghafarian P, Bakhshayesh Karam M, Ay M. Role of Lesion-to-Background Activity Ratio and Background Activity in Optimization of Reconstruction Protocols for FDG PET/CT Images of Overweight Patients: A Phantom Study. I J Radiol. 2023;20(1):e129306. doi: https://doi.org/10.5812/iranjradiol-129306

16
Jun
2022
Iran J Radiol

Investigation of the Clinical Efficacy of 99mTc-Sestamibi Washout in Patients with Acute Myocardial Infarction and Comparison with Stress Myocardial Imaging with 99mTc-Sestamibi Using a Two-Day Protocol

Mieko Ota,
Fuminori Hyodo,
Shinro Matsuo,
Takashi Kato,
Nobuyuki Kawai,
Fumihiko Nakamura
,et al.

Ota M, Hyodo F, Matsuo S, Kato T, Kawai N, et al. Investigation of the Clinical Efficacy of 99mTc-Sestamibi Washout in Patients with Acute Myocardial Infarction and Comparison with Stress Myocardial Imaging with 99mTc-Sestamibi Using a Two-Day Protocol. I J Radiol. 2022;19(2):e122423. doi: https://doi.org/10.5812/iranjradiol-122423

Download PDF1.75 MB
Share on
Cited by
Metrics

Ordering Reprints

Articles are published under the Creative Commons license stated on each article. No permission or royalty fee is required for uses permitted by that license. CCC handles optional bulk and customized reprint orders. Any quotation covers production and delivery services only, not copyright permission. > Request Reprints from CCC 

Search Relations

Author(s):

Related Articles