Strategy to manage T-Wave Oversensing in a Biventricular ICD

authors:

avatar Mohammad Ali Akbarzadeh ORCID 1 , 2 , * , avatar Mohammad Hasan Namazi 1 , 2 , avatar Morteza Safi 1 , 2 , avatar Mehdi Sheibani 3

MD, Cardiovascular Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran
MD, Cardiology Department, Shahid Modarress Hospital, Tehran, Iran
Loghman-Hakim Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran

how to cite: Akbarzadeh M A, Namazi M H, Safi M, Sheibani M. Strategy to manage T-Wave Oversensing in a Biventricular ICD. Int J Cardiovasc Pract. 2017;2(2):e130065. https://doi.org/10.21859/ijcp-020205.

Abstract

T-wave oversensing is a relatively prevalent cause of intracardiac signals oversensing in patients with Implantable Cardioverter Defibrillator (ICD). Some of these oversensings are typically corrected with device reprogramming. If reprogramming fails to resolve the issue, invasive options such as repositioning the implanted lead may be necessary. We present a patient with dilated cardiomyopathy and intermittent T wave oversensing by a cardiac resynchronization therapy (CRT) that was managed by altering V-V timing.