Optimization of thermal field parameters of the High Intensity Focused Ultrasound: HIFU simulator

Authors

S Gharloghi,*, V Dehlaghi
*Corresponding Author: Email: [email protected]

Journal of Inflammatory Diseases:Vol. 19, issue 1; 32-37
Published online:Apr 18, 2015
Article type:Research Article
How to Cite:Gharloghi S, Dehlaghi V. Optimization of thermal field parameters of the High Intensity Focused Ultrasound: HIFU simulator. J Inflamm Dis. 2024;19(1):e155888. doi:

Abstract

  Background: The High intensity focused ultrasound (HIFU) is a noninvasive and effective technique for tumor ablation. Frequency and acoustic power are effective parameters for temperature distribution and the extent of tissue damage.   Objective: The aim of this study was to optimize the frequency and acoustic power and to achieve appropriate temperature and thermal fields.   Methods: This analytical study was conducted in Kermanshah, 2014. Computer simulation was performed using MATLAB software (HIFU simulator toolbox). The KZK nonlinear wave equation was used to model sound propagation. The Bio heat equation was used to calculate the transient temperature in the liver tissue. Extracorporeal transducer was assessed over a frequency range of 2, 3 and 4 MHz and a power range of 50, 100 and 150 W.   Findings: In 2 MHz frequency, the temperature in the focal point reached to 44.5°C, 55°C and 75°C for 50, 100 and 150-W power, respectively. In 3 MHz frequency and the mentioned powers, the temperature reached to 47.5°C, 67°C and 94°C, respectively. The temperature reached to 45.5°C, 55 °C and 67 °C in 4 MHz frequency, respectively. 2, 3 and 4 MHz frequencies with 150-W power and 3 MHz frequency with 100-W power induced considerable extent of thermal dose.   Conclusion: 2-MHz frequency and 150W power can lead to more thermal dose in the same exposure time . These parameters can reduce treatment period and complications in larger tumors ablation.       Citation: Gharloghi S, Dehlaghi V. Optimization of thermal field parameters of the High Intensity Focused Ultrasound : HIFU simulator. J Qazvin Univ Med Sci. 2015 19 (1): 32-37.    

Copyright

© 2024, Journal of Inflammatory Diseases. This open-access article is available under the Creative Commons Attribution-NonCommercial 4.0 (CC BY-NC 4.0) International License (https://creativecommons.org/licenses/by-nc/4.0/), which allows for the copying and redistribution of the material only for noncommercial purposes, provided that the original work is properly cited.

Similar Articles

30
Apr
2007

Ultrasound Monitoring of Temperature Change during Interstitial Laser Thermotherapy of Liver: An In Vitro Study  

T. Gorji-Ara,
Manijhe Mokhtari Dizaji,
H. Ghanaati

Gorji-Ara T, Mokhtari Dizaji M, Ghanaati H. Ultrasound Monitoring of Temperature Change during Interstitial Laser Thermotherapy of Liver: An In Vitro Study  . I J Radiol. 2007;4(2):e79110. doi:

28
Sep
2015

Effect of Therapeutic Sequence of Hot Pack and Ultrasound on Physiological Response Over Trigger Point of Upper Trapezius

Dararat Benjaboonyanupap,
Aatit Paungmali,
Ubon Pirunsan

Benjaboonyanupap D, Paungmali A, Pirunsan U. Effect of Therapeutic Sequence of Hot Pack and Ultrasound on Physiological Response Over Trigger Point of Upper Trapezius. Asian J Sports Med. 2015;6(3):e23806. doi: https://doi.org/10.5812/asjsm.23806

24
Dec
2016

Added Value of Contrast-Enhanced Ultrasonography as a Planning Modality for Radiofrequency Ablation of Hepatocellular Carcinoma: Comparison with Gray-Scale Ultrasonography

So Hee Kim,
Jae Woong Choi,
Yang Shin Park,
Jongmee Lee,
Chang Hee Lee,
Kyeong Ah Kim
,et al.

Kim SH, Choi JW, Park YS, Lee J, Lee CH, et al. Added Value of Contrast-Enhanced Ultrasonography as a Planning Modality for Radiofrequency Ablation of Hepatocellular Carcinoma: Comparison with Gray-Scale Ultrasonography. I J Radiol. 2017;14(2):e43232. doi: https://doi.org/10.5812/iranjradiol.43232

29
Sep
2024
Shear Wave Elastography in Evaluating the Efficacy of Liver Microwave Ablation: An Experimental Study

Shear Wave Elastography in Evaluating the Efficacy of Liver Microwave Ablation: An Experimental Study

Dandan Xu,
Chao Tian,
Shan Wei

Xu D, Tian C, Wei S. Shear Wave Elastography in Evaluating the Efficacy of Liver Microwave Ablation: An Experimental Study. I J Radiol. 2024;21(2):e143996. doi: https://doi.org/10.5812/iranjradiol-143996

10
Nov
2021
Focused Ultrasound Surgery Applications in Treating Vulvar Non-neoplastic Epithelial Disorders

Focused Ultrasound Surgery Applications in Treating Vulvar Non-neoplastic Epithelial Disorders

Felix Wu Shun Wong,
Chengzhi Li

Wu Shun Wong F, Li C. Focused Ultrasound Surgery Applications in Treating Vulvar Non-neoplastic Epithelial Disorders. J Skin Stem Cell. 2021;8(3):e119766. doi: https://doi.org/10.5812/jssc.119766

More by these authors

S GharloghiPubMedScholar
V DehlaghiPubMedScholar
Share
Cited by
Metrics