Cover image of the article: Determination of a medical linac wedge factor dependency on the field size, depth and separation using Monte Carlo method to introduce an algorithm for treatment planning

Image Credit:Koomesh

Determination of a medical linac wedge factor dependency on the field size, depth and separation using Monte Carlo method to introduce an algorithm for treatment planning

Authors

peyman Hejai, Bizhan Hashemi,*, Majid Shahriari, MohammadTaghi Eivazi, Anoshiravan KazemNezhad
*Corresponding Author: Email: bhashemi@modaresd.ac.ir

Koomesh:Vol. 10, issue 1; 13-20
Published online:Oct 24, 2008
Article type:Research Article
How to Cite:Hejai P, Hashemi B, Shahriari M, Eivazi M, KazemNezhad A. Determination of a medical linac wedge factor dependency on the field size, depth and separation using Monte Carlo method to introduce an algorithm for treatment planning. koomesh. 2008;10(1):e152214. doi:

Abstract

Introduction: The study was performed to investigate the dependency of the wedge factors (WF) of a linac internal wedge with the depth and field size for 6MV photon beams. In addition, the current study was performed in order to accurate determination of the WF to introduce an algorithm to be incorporated it in treatment planning to decrease the level of systematic error due to this factor in clinical applications. Materials An Elekta linac head structure with and without the internal wedge was precisely simulated based on the data provided by the manufacturer using MCNP 4C Monte Carlo code. The simulated linac was then benchmarked against the experimental measurement. Thereafter, a mini - phantom was simulated using the same Monte Carlo code. Then, the relevant linac wedge factors due to the internal wedge were determined for different symmetric field sizes of 4, 5, 10, 15, and 20 cm2 at various depths ranged from 0.5 to 34cm, while the mini phantom was present in the fields. The wedge factors were determined based on the values, which are measured at 100cm SAD and 10 cm depth for each field size. Linear trend lines were fitted on the set of depth dependent wedge factors for each of the field sizes. Then the trend lines’ constants were analyzed for investigating their field size dependency. Results: Results indicated that the wedge factor is dependent on the depth for each field size. In addition, it was noted that the level of the wedge factor dependency on the depth varies for different field sizes. For example the wedge factor calculated at a depth of 10 cm had a variation of %7.4 when the field size varied from 4 to 20 cm2 symmetric square fields. Conclusion: ‎: In this study, a more accurate algorithm was determined that it can be used instead of the ‎usual wedge factor look-up tables used for MU calculation in radiotherapy planning. The variation of the ‎photon beam due to these circumstances can be determined accurately using Monte Carlo method and the ‎proposed algorithm. This algorithm enables us to reduce the level of possible systematic errors ‎encountered in clinical practices

Highlights

Copyright

© 2008, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.

Similar Articles

30
Jun
2013
Isodose 2D determining in water phantom regular and irregular fileds in radiation therapy for caculation modulated dose

Isodose 2D determining in water phantom regular and irregular fileds in radiation therapy for caculation modulated dose

Mohamad Javad Tahmasebi Biragani,
Nahid Chegeni,
Mansur Zabihzade,
Shole Arvandi,
Sasan Razmjoo Ghalaee,
Fatemeh Seif

Tahmasebi Biragani MJ, Chegeni N, Zabihzade M, Arvandi S, Razmjoo Ghalaee S, et al. Isodose 2D determining in water phantom regular and irregular fileds in radiation therapy for caculation modulated dose. Jentashapir J Cell Mol Biol. 2013;4(3):e94075. doi:

15
Feb
2007
Monte Carlo stimulation of the angular, radial and energy distribution of a medical linear accelerator using MCNP4C code

Monte Carlo stimulation of the angular, radial and energy distribution of a medical linear accelerator using MCNP4C code

Peyman Hejazi,
Bijan Hashemi,
Majid Shahriari,
Anoshiravan KazemNezhad

Hejazi P, Hashemi B, Shahriari M, KazemNezhad A. Monte Carlo stimulation of the angular, radial and energy distribution of a medical linear accelerator using MCNP4C code. koomesh. 2007;8(2):e152139. doi:

25
Mar
2011
An assessment of the factors involved in effective attenuation coefficient of the compensator material for the treatment with 6MV photons using intensity modulated radiation therapy method

An assessment of the factors involved in effective attenuation coefficient of the compensator material for the treatment with 6MV photons using intensity modulated radiation therapy method

Abbas Haghparast,
Bijan Hashemi,
Mohammad Taghi Eyvazi

Haghparast A, Hashemi B, Eyvazi MT. An assessment of the factors involved in effective attenuation coefficient of the compensator material for the treatment with 6MV photons using intensity modulated radiation therapy method. koomesh. 2011;12(3):e152458. doi:

26
Feb
2019
Comparison of Dosimetric Parameters Between Field in Field and Conformal Radiation Therapy Techniques in Early Stage of Left Breast Cancer Patients

Comparison of Dosimetric Parameters Between Field in Field and Conformal Radiation Therapy Techniques in Early Stage of Left Breast Cancer Patients

Sephora Nokhasteh,
Hamideh Nazemi,
Payman Hejazi,
Mahdieh Dayyani

Nokhasteh S, Nazemi H, Hejazi P, Dayyani M. Comparison of Dosimetric Parameters Between Field in Field and Conformal Radiation Therapy Techniques in Early Stage of Left Breast Cancer Patients. Int J Cancer Manag. 2019;12(2):e84123. doi: https://doi.org/10.5812/ijcm.84123

10
May
2016
Determining the Effective Source-Surface Distance for Therapeutic Electron Beams

Determining the Effective Source-Surface Distance for Therapeutic Electron Beams

Mohammad Javad Tahmasebi Birgani,
Mansour Zabihzadeh,
Sholeh Arvandi,
Elahe Gharibreza

Tahmasebi Birgani MJ, Zabihzadeh M, Arvandi S, Gharibreza E. Determining the Effective Source-Surface Distance for Therapeutic Electron Beams. Jentashapir J Cell Mol Biol. 2016;7(3):e30975. doi: https://doi.org/10.17795/jjhr-30975

More by these authors

peyman HejaiPubMedScholar
Bizhan HashemiPubMedScholar
Majid ShahriariPubMedScholar
MohammadTaghi EivaziPubMedScholar
Anoshiravan KazemNezhadPubMedScholar
Share
Cited by
Metrics