1. Background
2. Objectives
3. Methods
3.1. Design and Sample
3.2. Statistical Analysis
4. Results
| Item | I-CVI | CVR | Result |
|---|---|---|---|
| Before operation | |||
| The cable of the ESU is twisted around the electric generator. | 0.85 | 0.5 | Pass |
| The ESU is directly plugged into a power outlet or a power distributor with earth connection. | 1 | 0.9 | Pass |
| It was assured that the ESU was in good working order before the operation. | 1 | 0.8 | Pass |
| The periodic calibration of the unit has been done as prescribed by the manufacturer. | 0.65 | 1 | Pass |
| The cable of the ESU is overstretched. | 1 | 0.6 | Pass |
| The cable of the ESU disrupts the movements of the surgical team. | 0.85 | 0.6 | Pass |
| The manual of the ESU is in full view of all the members of the surgical team. | 1 | 0.8 | Pass |
| The generator of the ESU is not used as a table for other surgical tools. | 1 | 0.9 | Pass |
| The cable of the electrosurgical pencil is kinked. | 1 | 1 | Pass |
| The cable of electrosurgical pencil was attached to the sterile drape with a metal clamp. | 1 | 0.9 | Pass |
| To prevent fires and burning the patient’s skin, inflammable prep solutions were dried before draping and application of the ESU. | 1 | 1 | Pass |
| Active monopolar electrodes were used for patients with cardiac pacemakers. | 1 | 0.9 | Pass |
| The electrosurgical plate has been placed on a muscular area free of moisture, hair, scars, or tattoos. | 1 | 1 | Pass |
| The electrosurgical plate was in full contact with the patient’s body during the entire course of the surgery. | 1 | 1 | Pass |
| The patient’s metallic belongings (jewelry, hairclips, etc.) were removed before they were admitted into the operating room. | 1 | 0.9 | Pass |
| There is contact between the patient’s body and the operating table and its metallic accessories. | 1 | 0.7 | Pass |
| The patient’s skin condition was examined and recorded before applying the ESU. | 1 | 0.9 | Pass |
| The size of the electrosurgical plate has been selected according to the patient’s size and age. | 1 | 0.5 | Pass |
| In the course of operation | |||
| The tip of the electrosurgical pencil is cleaned with a special pad or moist gauze (scalpels or other metallic tools are not used). | 0.80 | 0.8 | Pass |
| The sound of the active mode of the electrosurgical pencil is audible to the entire members of the surgical team. | 1 | 0.8 | Pass |
| The tip of the electrosurgical pencil is never dipped into a liquid environment. | 1 | 0.7 | Pass |
| When not in use, the electrosurgical pencil is placed in a safe and insulated case. | 0.85 | 0.8 | Pass |
| A suction system equipped with smoke absorber filters is used to extract the smokes from the ESU. | 1 | 0.9 | Pass |
| Those members of the surgical team who are exposed to the smokes from the ESU are wearing masks with high filtration capacity (N95). | 1 | 0.6 | Pass |
| In the application of buzzing forceps, an active current of over 3 seconds is employed. | 1 | 0.9 | Pass |
| In the application of buzzing forceps, the tip of the active electrode is in contact with the lower part of the clamp. | 1 | 0.7 | Pass |
| In the application of buzzing forceps, the tip of the active electrode is activated after it contacts the body of the clamp. | 1 | 0.6 | Pass |
| The ESU is set according to the conditions (age, tissue, etc.) of the patient. | 0.80 | 0.9 | Pass |
| Liquids are not poured over any of the sections of the ESU. | 1 | 0.9 | Pass |
| After operation | |||
| Blood or any other fluids splashed over the ESU are cleaned after each surgery according to the instructions of the manufacturer. | 1 | 0.8 | Pass |
| The tip of the electrosurgical pencil is considered as part of the tools that are counted. | 1 | 0.8 | Pass |
| The patient’s skin condition is examined and recorded after vapplying the ESU. | 1 | 0.8 | Pass |
Abbreviation: ESU, electrosurgical unit.
aS-CVI/Ave = 0.97; S-CVI/UA = 0.82; according to Lawshe table, given that 20 experts are consulted, the minimum acceptable value for CVR is 0.42.
| Group | Time | Wilcoxon Statistic (z) | P Value | |
|---|---|---|---|---|
| Pretest | Posttest | |||
| Intervention | 17.2 ± 2.57 | 28.9 ± 1.52 | -2.807 | 0.005 |
| Control | 18.1 ± 2.23 | 18.1 ± 2.23 | -0.179 | 0.855 |
| The Mann-Whitney statistic (z) | -0.651 | -3.8 | ||
| P value | 0.515 | 0.001 | ||
aValues are expressed as mean ± SD.
| Number | Item | Kappa |
|---|---|---|
| 1 | The cable of the ESU is twisted around the electric generator. | 1 |
| 2 | The ESU is directly plugged into a power outlet or a power distributor with earth connection. | 0.61 |
| 3 | It is assured that the ESU is in good working order before the operation. | 0.61 |
| 4 | The periodic calibration of the unit is done as prescribed by the manufacturer. | 1 |
| 5 | The cable of the ESU is overstretched. | 1 |
| 6 | The cable of the ESU disrupts the movements of the surgical team. | 0.9 |
| 7 | The manual of the ESU is in full view of all the members of the surgical team. | 0.61 |
| 8 | The generator of the ESU is not used as a table for other surgical tools. | 0.6 |
| 9 | The cable of the electrosurgical pencil is kinked. | 1 |
| 10 | The cable of electrosurgical pencil is attached to the sterile drape with a metal clamp. | 1 |
| 11 | To prevent fires and burning the patient's skin, inflammable prep solutions are dried before draping and application of the ESU. | 1 |
| 12 | Active monopolar electrodes are used for patients with cardiac pacemakers. | 1 |
| 13 | The electrosurgical plate is placed on a muscular area free of moisture, hair, scars, or tattoos. | 0.9 |
| 14 | The electrosurgical plate is in full contact with the patient’s body during the entire course of the surgery. | 0.73 |
| 15 | The patient’s metallic belongings (jewelry, hairclips, etc.) are removed before they are admitted into the operating room. | 0.61 |
| 16 | There is contact between the patient’s body and the operating table and its metallic accessories. | 0.73 |
| 17 | The patient’s skin condition are examined and recorded before applying the ESU. | 0.73 |
| 18 | The size of the electrosurgical plate is selected according to the patient’s size and age. | 1 |
| 19 | The tip of the electrosurgical pencil is cleaned with a special pad or moist gauze (scalpels or other metallic tools are not used). | 1 |
| 20 | The sound of the active mode of the electrosurgical pencil is audible to the entire members of the surgical team. | 1 |
| 21 | The tip of the electrosurgical pencil is never dipped into a liquid environment. | 1 |
| 22 | When not in use, the electrosurgical pencil is placed in a safe and insulated case. | 1 |
| 23 | A suction system equipped with smoke absorber filters is used to extract the smokes from the ESU. | 1 |
| 24 | Those members of the surgical team exposed to the smokes from the ESU are wearing masks with high filtration capacity (N95). | 1 |
| 25 | In the application of buzzing forceps, an active current of over 3 seconds is employed. | 1 |
| 26 | In the application of buzzing forceps, the tip of the active electrode is in contact with the lower part of the clamp. | 1 |
| 27 | In the application of buzzing forceps, the tip of the active electrode is activated after it contacts the body of the clamp. | 1 |
| 28 | The ESU is set according to the conditions (age, tissue, etc.) of the patient. | 0.61 |
| 29 | Liquids are not poured over any of the sections of the ESU. | 1 |
| 30 | Blood or any other fluids splashed over the ESU are cleaned after each surgery according to the instructions of the manufacturer. | 1 |
| 31 | The tip of the electrosurgical pencil is considered as part of the tools that are counted. | 1 |
| 32 | The patient’s skin condition is examined and recorded after applying the ESU. | 0.73 |