Comparison of Maximum Signal Intensity of Contrast Agent on T1-Weighted Images Using Spin Echo, Fast Spin Echo and Inversion Recovery Sequences

Author(s):
Mahmood NazarpoorMahmood Nazarpoor1,*, Masoud PoureisaMasoud Poureisa2, Mohammad Hossein DaghighiMohammad Hossein Daghighi2
1Department of Radiology, Faculty of Paramedicine, Tabriz University of Medical Sciences, [email protected], Iran
2Department of Radiology, Faculty of Medicine, Tabriz University of Medical Sciences, Iran

IJ Radiology:Vol. 10, issue 1; 27-32
Published online:Dec 27, 2012
Article type:Research Article
Received:May 05, 2012
Accepted:Oct 25, 2012
How to Cite:Nazarpoor M, Poureisa M, Daghighi MH. Comparison of Maximum Signal Intensity of Contrast Agent on T1-Weighted Images Using Spin Echo, Fast Spin Echo and Inversion Recovery Sequences. I J Radiol. 2012;10(1):. doi: https://doi.org/10.5812/iranjradiol.5452

Abstract

Background:

MRI is not able to directly measure the concentration of contrast agent. It is measured indirectly from the signal intensity (SI). It is very important to know how much contrast agent should be injected to receive a maximum SI in the region of interest (ROI).

Objectives:

The aim of this study was to investigate the maximum relationship between contrast concentration and signal intensity (SI) on T1-weighted images using spin echo (SE), fast spin echo (FSE) and inversion recovery (IR) sequences.

Materials and Methods:

To assess the relationship between SI and concentration, a water-filled phantom containing vials of different concentrations of gadolinium DTPA (Gd-DTPA) (0 to 19.77 mmol/L) or a constant concentration (1.2 mmol/L) of contrast agent was used. The vials of constant concentration were used to measure coil nonuniformity. The mean SI was obtained in the ROI using T1-weighted images. All studies were carried out using a 0.3 T clinical MR scanner with a standard head coil.

Results:

This study shows that maximum SI will appear at different ranges in different sequences. The maximum SI can be seen at concentrations of 5.95, 4.96 and 3.98mmol/L for SE, FSE and IR, respectively.

Conclusion:

Using standard imaging parameters, each MRI sequence reaches its maximum SI in a specific contrast concentration, which is highest in SE and least in IR in a comparison between SE, FSE and IR sequences.

Full Text

Full text is available in PDF

Similar Articles

30
Dec
2009

The Effect of Repetition Time on the Maximum Linear Relationship Between Contrast Agent Concentration and Signal Intensity on T1-Weighted Image Using Inversion Recovery (IR) Sequence

Mahmood Nazarpoor

Nazarpoor M. The Effect of Repetition Time on the Maximum Linear Relationship Between Contrast Agent Concentration and Signal Intensity on T1-Weighted Image Using Inversion Recovery (IR) Sequence. I J Radiol. 2009;6(4):e78908. doi:

15
Jun
2014

The Effect of Inversion Time on the Relationship Between Iron Oxide Nanoparticles Concentration and Signal Intensity in T1-Weighted MR Images

Hodaiseh Saharkhiz,
Nahideh Gharehaghaji,
Mahmood Nazarpoor,
Asghar Mesbahi,
Masoud Pourissa

Saharkhiz H, Gharehaghaji N, Nazarpoor M, Mesbahi A, Pourissa M. The Effect of Inversion Time on the Relationship Between Iron Oxide Nanoparticles Concentration and Signal Intensity in T1-Weighted MR Images. I J Radiol. 2014;11(2):e12667. doi: https://doi.org/10.5812/iranjradiol.12667

22
Apr
2015

Effect of Flip Angle on the Correlation Between Signal Intensity and Different Concentrations of Iron Oxide Nanoparticles Using T1-Weighted Turbo-FLASH Inversion Recovery Sequence

Nahideh Gharehaghaji,
Mahmood Nazarpoor,
Hodaiseh Saharkhiz

Gharehaghaji N, Nazarpoor M, Saharkhiz H. Effect of Flip Angle on the Correlation Between Signal Intensity and Different Concentrations of Iron Oxide Nanoparticles Using T1-Weighted Turbo-FLASH Inversion Recovery Sequence. I J Radiol. 2015;12(2):e22887. doi: https://doi.org/10.5812/iranjradiol.22887

9
Oct
2021

Evaluation of the Effect of Multiple Linear Gadolinium-Based Contrast Agent Exposures on the Signal Intensity of the Dentate Nucleus in Multiple Sclerosis Patients

Maryam Mohammadzadeh,
Zahra Sheibani,
Madjid Shakiba,
Amir Reza Azimi,
Arezou Hashemzadeh,
Mojtaba Barzegar
,et al.

Mohammadzadeh M, Sheibani Z, Shakiba M, Azimi AR, Hashemzadeh A, et al. Evaluation of the Effect of Multiple Linear Gadolinium-Based Contrast Agent Exposures on the Signal Intensity of the Dentate Nucleus in Multiple Sclerosis Patients. I J Radiol. 2021;18(4):e114397. doi: https://doi.org/10.5812/iranjradiol.114397

20
Jan
2018
Time-Dependent Changes in the Magnetic Resonance Detection of Brain Metastases Using Single-Dose Gadobutrol

Time-Dependent Changes in the Magnetic Resonance Detection of Brain Metastases Using Single-Dose Gadobutrol

Bing Fan,
Mei-jiao Li,
Xiao-ying Wang,
Yu-feng Xu,
Fei-yu Li,
Lai-yun Zhang
,et al.

Fan B, Li M, Wang X, Xu Y, Li F, et al. Time-Dependent Changes in the Magnetic Resonance Detection of Brain Metastases Using Single-Dose Gadobutrol. I J Radiol. 2018;15(1):e66054. doi: https://doi.org/10.5812/iranjradiol.66054


Crossmark
Crossmark
Checking
Share on
Cited by
Metrics

Ordering Reprints

Articles are published under the Creative Commons license stated on each article. No permission or royalty fee is required for uses permitted by that license. CCC handles optional bulk and customized reprint orders. Any quotation covers production and delivery services only, not copyright permission. > Request Reprints from CCC 

Search Relations

Author(s):

Related Articles