How Low Can We Go Dose Saving Through Low-Dose Examination for Urolithiasis Via Modern Computer Tomography with Tinfilter

Author(s):
Youssef ErfanianYoussef Erfanian1,*, Nika GuberinaNika Guberina1, Saravanabavaan SuntharalingamSaravanabavaan Suntharalingam1, Jens TheysohnJens Theysohn1
1University Hospital Essen/ Germany

IJ Radiology:Vol. 14, issue Special Issue; e48020
Published online:Apr 12, 2017
Article type:Abstract
Received:Dec 20, 2016
Accepted:Feb 07, 2017
How to Cite:Erfanian Y, Guberina N, Suntharalingam S, Theysohn J. How Low Can We Go Dose Saving Through Low-Dose Examination for Urolithiasis Via Modern Computer Tomography with Tinfilter. I J Radiol. 2017;14(Special Issue):e48020. doi: https://doi.org/10.5812/iranjradiol.48020

Abstract

Objectives:

To compare radiation dose and image quality of three different CT devices using low dose abdominal protocols in patients with suspected urolithiasis. Differences in radiation dose due to usage of additional hardware (tin filter) and software (newer generation iterative reconstruction) were evaluated.

Methods:

Low - dose examinations from two standard CT (siemens AS +, siemens flash) devices with a tin - filter CT (siemens force) were compared regarding dose-length product (DLP) and computed tomography dose index volume (CTDIvol). Image quality of each CT scan was assessed using a five point Likert scale (0 = major blurring, 4 = excellent depiction). Kruskal - Wallis analysis was performed to test for significant subgroup differences in DLP, CTDIvol and image quality. An interrater agreement concerning image quality was evaluated.

Results:

CT examinations of a total of 143 patients were assessed. DLP of Force was 56% and 55% lower in comparison to AS+ and flash, (74.3, 166.1 and 164.6 mGycm respectively; P < 0.001). CTDIvol of Force showed 60% and 58% significantly lower values as compared to AS + and Flash, respectively (1.4 vs. 3.5vs. 3.3 mGy; P < 0.001). Image quality in force proved to be slightly higher as compared to flash and AS + (3.67, 3.50 and 3.59 respectively; P > 0.05). Interrater agreement regarding image quality was substantial for all CT - devices (? = 0.75, n = 143).

Conclusions:

Modern CTs with the innovative technology of a built-in tin filter allow for significant reduction of radiation exposure in patients with suspected urolithiasis by optimizing the X - Ray spectrum with image quality as high as in standard CT devices

Full Text

Full text is available in PDF

Similar Articles

17
Jun
2018
https://www.roseradiology.com/multidetector16slicect.htm

Assessment of Radiation Dose and Image Quality of Multidetector Computed Tomography

Haytham Al Ewaidat,
Xiaoming Zheng,
Yousef Khader,
Mostafa Abdelrahman,
Mostafa Khaled Mustafa Alhasan,
Mohammad Ahmmad Rawashdeh
,et al.

Al Ewaidat H, Zheng X, Khader Y, Abdelrahman M, Mustafa Alhasan MK, et al. Assessment of Radiation Dose and Image Quality of Multidetector Computed Tomography. I J Radiol. 2018;15(3):e59554. doi: https://doi.org/10.5812/iranjradiol.59554

26
Apr
2023
Innov J Radiol

A Study of Fixed Current and Automatic Tube Current Modulation at a Fixed Low Tube Voltage in Low-dose Computed Tomography (CT) Imaging of Infant Lungs

Wushuang Chen,
Shilong Tang,
Zhen Xu,
Xin Chen,
Yinglan Zheng,
Ling He

Chen W, Tang S, Xu Z, Chen X, Zheng Y, et al. A Study of Fixed Current and Automatic Tube Current Modulation at a Fixed Low Tube Voltage in Low-dose Computed Tomography (CT) Imaging of Infant Lungs. I J Radiol. 2023;20(2):e126362. doi: https://doi.org/10.5812/iranjradiol-126362

20
Jun
2016

Dosimetry of Three Cone Beam Computerized Tomography Scanners at Different Fields of View in Terms of Various Head and Neck Organs

Sima Nikneshan,
Mahmood Reza Aghamiri,
Ehsan Moudi,
Nika Bahemmat,
Hoora Hadian

Nikneshan S, Aghamiri MR, Moudi E, Bahemmat N, Hadian H. Dosimetry of Three Cone Beam Computerized Tomography Scanners at Different Fields of View in Terms of Various Head and Neck Organs. I J Radiol. 2016;13(3):e34220. doi: https://doi.org/10.5812/iranjradiol.34220

15
Oct
2019
Iranian Journal of Radiology

Effect of Arm Position on Image Quality and Radiation Dose in Multidetector Computed Tomography

Dae Cheol Kweon,
Jiwon Choi

Kweon DC, Choi J. Effect of Arm Position on Image Quality and Radiation Dose in Multidetector Computed Tomography. I J Radiol. 2019;16(4):e86280. doi: https://doi.org/10.5812/iranjradiol.86280

7
Sep
2016

Comparison of Radiation Dose and Image Quality with Various Computed-Tomography Scout Views: The Angular Modulation Technique Based on Information Calculated from Scout Views

Myeong Seong Kim,
Jong-Woong Lee,
Sun Geun Kim,
Dae Cheol Kweon

Seong Kim M, Lee J, Geun Kim S, Cheol Kweon D. Comparison of Radiation Dose and Image Quality with Various Computed-Tomography Scout Views: The Angular Modulation Technique Based on Information Calculated from Scout Views. I J Radiol. 2017;14(1):e13477. doi: https://doi.org/10.5812/iranjradiol.35606


Crossmark
Crossmark
Checking
Share on
Cited by
Metrics

Ordering Reprints

Articles are published under the Creative Commons license stated on each article. No permission or royalty fee is required for uses permitted by that license. CCC handles optional bulk and customized reprint orders. Any quotation covers production and delivery services only, not copyright permission. > Request Reprints from CCC 

Search Relations

Author(s):

Related Articles