Diagnostic Accuracy of Sixty Four Multi-Slice CT Angiography in Assessment of Arterial Cut-Off and Run-Off in Comparison with Surgical Findings
Authors
Abstract
Background/Objective: The accurate anatomic mapping and determination of the severity of arterial disease an important health problem of the elderly is of great significance. We aimed to determine the diagnostic value of 64-multislice CT angiography (MSCTA) in arterial disease in run-off and cut-off sites, taking into consideration the results of surgical findings of our study as a standard of reference.
Patients and Methods: Throughout the study, MSCTA followed by an operative intervention was carried out on a total of 38 patients with clinical signs and symptoms suggestive of arterial disease (AD) and all had the indication for vascular surgery. The mean age of patients was 34±15.86 (range, 23 to 93) years. MSCTA was executed using a 64-slice CT scanner, during the arterial phase of injecting the nonionic, contrast medium with a power injector at the rate of 5 ml/sec into the antecubital vein and exploration and revascularization of peripheral arterial disease was performed intraoperatively.
Results: The most common cause of vascular occlusion was atherosclerosis and arterial disease was more common in the lower extremities. The most frequent site of stenosis due to MSCTA findings was in the superficial femoral artery. Spearman’s correlation coefficients showed a high degree of agreement amongst the raters. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and the accuracy of MSCTA compared to surgery were 83.8%, 96%, 96.8%, 81.3% and 89%, respectively. MSCTA findings were compared with surgery as a standard of reference which showed concordance in the majority of cases (81.6%). Cut-off sites were correctly identified by MSCTA in 97.3% of the patients and the most common sites of discordance observed in our study were in run-off sites (18.2%).
Conclusion: MSCTA angiography as a novel diagnostic modality may be a suitable alternative and a viable choice for routine clinical diagnosis.
Highlights
References
- 1.The references are available in PDF file.
Copyright
© 2011, Author(s). This open-access article is available under the Creative Commons Attribution 4.0 (CC BY 4.0) International License (https://creativecommons.org/licenses/by/4.0/), which allows for unrestricted use, distribution, and reproduction in any medium, provided that the original work is properly cited.
Similar Articles
Diagnostic Accuracy of Dual-Source Computerized Tomography Coronary Angiography in Symptomatic Patients Presenting to a Referral Cardiovascular Center During Daily Clinical Practice
Mahdavi A, Mohammadzadeh A, Joodi G, Tabatabaei MR, Sheikholeslami F, et al. Diagnostic Accuracy of Dual-Source Computerized Tomography Coronary Angiography in Symptomatic Patients Presenting to a Referral Cardiovascular Center During Daily Clinical Practice. I J Radiol. 2016;13(2):e59322. doi: https://doi.org/10.5812/iranjradiol.24350
Stent Stenosis After Percutaneous Coronary Intervention
Motevalli M, Ghanaati H, Karami P, Rahmani M, Karami A, et al. Stent Stenosis After Percutaneous Coronary Intervention. I J Radiol. 2014;11(30th Iranian Congress of Radiology):e21255. doi: https://doi.org/10.5812/iranjradiol.21255
Assessment of the Coronary Artery in High-Pitch, Dual-Source CT Aortography Without ECG Synchronization
Choi HG, Park MJ, Choi HC, Choi HY, Shin HS, et al. Assessment of the Coronary Artery in High-Pitch, Dual-Source CT Aortography Without ECG Synchronization. I J Radiol. 2017;14(3):e22024. doi: https://doi.org/10.5812/iranjradiol.22024
Preoperative Multi-Detector Computed Tomography in Kidney Donors; Quantitative Data Report from Operation Rooms
Nezami N, Naghavi-Behzad M, Piri R, Salari B, Hool S, et al. Preoperative Multi-Detector Computed Tomography in Kidney Donors; Quantitative Data Report from Operation Rooms. I J Radiol. 2018;15(4):e59025. doi: https://doi.org/10.5812/iranjradiol.59025
Prevalence of Congenital Coronary Artery Anomalies and Variants in 2697 Consecutive Patients Using 64-Detector Row Coronary CTAngiography
Arjmand Shabestari A, Akhlaghpoor S, Tayebivaljozi R, Fattahi Masrour F. Prevalence of Congenital Coronary Artery Anomalies and Variants in 2697 Consecutive Patients Using 64-Detector Row Coronary CTAngiography. I J Radiol. 2012;9(3):111-121. doi: https://doi.org/10.5812/iranjradiol.8070
- Scopus by DOI: 0
Last Update: 1 month ago
- Scopus by Title: 2
Last Update: 1 month ago
- Scopus by Title (Ref): 2
Last Update: 1 month ago
- CrossRef: 0
Last Update: 5 days ago
- OpenAlex: 0
Last Update: 1 day ago
- OpenCitations: 0
Last Update: 2 days ago