EktaspeedThe effect of alterations in processing solution and exposure times on Dentus M2 and Ektaspeed intra-oral films
Abstract
Background: The processing procedure influences the degree of density and contrast of the resulting radiograph. Objective: To determine the effect of changes in processing solution and exposure times on Dentus M2 and Ektaspeed intra-oral films. Methods: This was an analytical study in which 44 samples of each type of the films were used. The distance between the source and the film surface was 28 inches. The exposure was performed using 11 different times, vertically, with maximum voltage of 70 and 8 mA. Processing was carried out at days 1, 5, 10 and 15 using an automatic machine. Later, the density of each film was measured by Digital Density Meter. Data were analyzed by T-Test and paired T-Test. Finding: Longer exposure time caused an increase in density of both types of the films. The mean density of either type decreased through days 1 to 15. The mean density was 0.92±0.09 for Dentus M2 and 0.75±0.09 for Ektaspeed in all processing time. However, a significant difference was found between two groups (p<0.05). Conclusion: The Dentus M2 is recommended asboth result in a intra-oral film showed to have higher density with lower variability than the Ektaspeed film
Copyright
© 2024, Journal of Inflammatory Diseases. This open-access article is available under the Creative Commons Attribution-NonCommercial 4.0 (CC BY-NC 4.0) International License (https://creativecommons.org/licenses/by-nc/4.0/), which allows for the copying and redistribution of the material only for noncommercial purposes, provided that the original work is properly cited.
Similar Articles
Compairing accuracy of Kodak Ektaspeed and Agfa Dentus M2 dental X-ray films in detection of caries
Marami A, Taghavi N, Zareyosefi A. Compairing accuracy of Kodak Ektaspeed and Agfa Dentus M2 dental X-ray films in detection of caries. J Inflamm Dis. 2024;10(3):e155245. doi:
Accuracy of linear tomography in comparison with computed tomography to assess bone quantity for dental implant treatment
Tofangchiha M, Mohammadpour M, Shojaei fard B, Marami A, Rahrotaban S. Accuracy of linear tomography in comparison with computed tomography to assess bone quantity for dental implant treatment. J Inflamm Dis. 2024;17(6):e155805. doi:
Diagnostic Accuracy of Digital and Conventional Radiography in the Detection of Non-Cavitated Approximal Dental Caries
Abesi F, Mirshekar A, Moudi E, Seyedmajidi M, Haghanifar S, et al. Diagnostic Accuracy of Digital and Conventional Radiography in the Detection of Non-Cavitated Approximal Dental Caries. I J Radiol. 2012;9(1):. doi: https://doi.org/10.5812/iranjradiol.6747
Dosimetry of organ at risks in Orthopantomography
Rafati M, Eftekhari-Moghadam A, Mehri-Kakavand G, Hasanzadeh H, Maftoon A. Dosimetry of organ at risks in Orthopantomography. koomesh. 2015;17(1):e150800. doi:
Evaluating the Affect of Tube Current’s on Different Dental Material Artifacts in Cone Beam Computed Tomography Using Signal Difference to Noise Ratio Index
Safi Y, Moshfeghi M, Vasegh Z, Aghdasi MM, Haghani J, et al. Evaluating the Affect of Tube Current’s on Different Dental Material Artifacts in Cone Beam Computed Tomography Using Signal Difference to Noise Ratio Index. I J Radiol. 2017;14(2):e38620. doi: https://doi.org/10.5812/iranjradiol.38620
- Scopus by DOI: 0
Last Update: 0 months ago
- Scopus by Title: 0
Last Update: 0 months ago
- Scopus by Title (Ref): 0
Last Update: 0 months ago
- CrossRef: 0
Last Update: 4 days ago
Ordering Reprints
Articles are published under the Creative Commons license stated on each article. No permission or royalty fee is required for uses permitted by that license. CCC handles optional bulk and customized reprint orders. Any quotation covers production and delivery services only, not copyright permission. > Request Reprints from CCC