Comparison of Maximum Signal Intensity of Contrast Agent on T1-Weighted Images Using Spin Echo, Fast Spin Echo and Inversion Recovery Sequences

Author(s):
Mahmood NazarpoorMahmood Nazarpoor1,*, Masoud PoureisaMasoud Poureisa2, Mohammad Hossein DaghighiMohammad Hossein Daghighi2
1Department of Radiology, Faculty of Paramedicine, Tabriz University of Medical Sciences, [email protected], Iran
2Department of Radiology, Faculty of Medicine, Tabriz University of Medical Sciences, Iran


IJ Radiology:Vol. 10, issue 1; 27-32
Published online:Dec 27, 2012
Article type:Research Article
Received:May 05, 2012
Accepted:Oct 25, 2012
How to Cite:Mahmood NazarpoorMasoud PoureisaMohammad Hossein DaghighiComparison of Maximum Signal Intensity of Contrast Agent on T1-Weighted Images Using Spin Echo, Fast Spin Echo and Inversion Recovery Sequences.I J Radiol.2012;10(1):27-32.https://doi.org/10.5812/iranjradiol.5452.

Abstract

Background:

MRI is not able to directly measure the concentration of contrast agent. It is measured indirectly from the signal intensity (SI). It is very important to know how much contrast agent should be injected to receive a maximum SI in the region of interest (ROI).

Objectives:

The aim of this study was to investigate the maximum relationship between contrast concentration and signal intensity (SI) on T1-weighted images using spin echo (SE), fast spin echo (FSE) and inversion recovery (IR) sequences.

Materials and Methods:

To assess the relationship between SI and concentration, a water-filled phantom containing vials of different concentrations of gadolinium DTPA (Gd-DTPA) (0 to 19.77 mmol/L) or a constant concentration (1.2 mmol/L) of contrast agent was used. The vials of constant concentration were used to measure coil nonuniformity. The mean SI was obtained in the ROI using T1-weighted images. All studies were carried out using a 0.3 T clinical MR scanner with a standard head coil.

Results:

This study shows that maximum SI will appear at different ranges in different sequences. The maximum SI can be seen at concentrations of 5.95, 4.96 and 3.98mmol/L for SE, FSE and IR, respectively.

Conclusion:

Using standard imaging parameters, each MRI sequence reaches its maximum SI in a specific contrast concentration, which is highest in SE and least in IR in a comparison between SE, FSE and IR sequences.

Full Text

Full text is available in PDF

comments

Leave a comment here


Crossmark
Crossmark
Checking
Share on
Metrics

Purchasing Reprints

  • Copyright Clearance Center (CCC) handles bulk orders for article reprints for Brieflands. To place an order for reprints, please click here (   https://www.copyright.com/landing/reprintsinquiryform/ ). Clicking this link will bring you to a CCC request form where you can provide the details of your order. Once complete, please click the ‘Submit Request’ button and CCC’s Reprints Services team will generate a quote for your review.