Abstract
Patients and Methods: In this study, the vertical distances between the metal markers were measured by panoramic radiography in seven sites of two skulls in various head positions. Then the radiographic measurements were compared with the actual values.
Results: Eighty three percent of the measurements were underestimated, 8.5% were overestimated on panoramic radiography and 8.5% of the measurements had no difference with the real measurements. Overestimation was not greater than 1 mm. The difference between actual and radiographic measurements was less in the posterior areas and in the mandible . In all head positions, the most difference between actual and radiographic measurements occurred in the anterior area.
Conclusion: Based on the results of this study, linear measurements on panoramic radiography are more reliable in the posterior areas and may be used in early clinical measurements.
Keywords
Panoramic Radiography Linear Measurement Mandibular Canal Mental Foramen
Fulltext
Full text is available in PDF.
References
-
1.
The references are available in PDF file.