In this pilot study, we acquired body navigation-loaded ultrasound images, which included information regarding the inspection site and transducer location. The most important advantage of navigation-loaded ultrasonography is that it increases the operator’s convenience and is expected to allow the operator to focus solely on the ultrasound examination. In some cases, the operator’s increased effort may be required for adding appropriate body marks or text available on the ultrasound equipment to the ultrasound images. This process is both cumbersome and time-consuming, as radiologists or clinicians need to make a diagnosis by acquiring ultrasound images and add a body mark to the image using various control buttons. However, our navigation-loaded images can make this process unnecessary, because the exact information regarding the transducer location and inspection site is automatically integrated in the ultrasound image in real time. Therefore, the navigation-loaded ultrasound image is expected to assist in interpreting ultrasound images when distinction between the right and left sides is important, or when it is necessary to mark specific locations (e.g., ultrasound-guided procedures). The disadvantage of our semi-automated technology compared to the existing ultrasound body mark system is that it takes a few minutes to fix the camera and set the software right before the ultrasound examination.
In the analysis of ultrasound images, the mean interpretation scores of all raters increased significantly by using the body navigation-loaded ultrasound images compared to conventional ultrasound images. It was also confirmed that the inter-rater agreement improved in the second analysis using the navigation-loaded ultrasound image set. Therefore, body navigation-loaded ultrasound imaging allows physicians to interpret ultrasound images more accurately and objectively. Nevertheless, based on the present results, the presence or absence of navigation-loaded images did not significantly influence the target organ identification of easily recognizable organs, such as the left lateral section of the liver, portal vein bifurcation, spleen, or urinary bladder; these organs were identified in nearly 100% of cases in all primary and secondary analyses.
In the present study, there was a significant difference in the interpretation scores of the reviewers (
Table 2). Reviewer A (a senior radiologist) had a higher interpretation score than others. Apart from the low scores of the less experienced resident, the observed difference in the scores of the two experienced radiologists, interpreting the navigation-loaded ultrasound images in the second analysis, was related to the transducer location interpretation.
Considering the evaluation of the boundaries of nine abdominal regions, the senior radiologist’s interpretation of the transducer location was more consistent with the operator's intention. Since it was possible to have a subjective interpretation of the boundary area, the inter-rater agreement for the transducer location was lower than that of other evaluation categories (
Table 3). By using the four abdominal quadrants with a clear anatomical landmark of the umbilicus as the standard for interpreting the transducer location, all raters were expected to have a higher interpretation score and a high level of agreement.
There are several limitations to this pilot study on this experimental technology. First, the economic or diagnostic efficacy of this technology compared to the conventional ultrasound body mark system was not evaluated. Some physicians may assume that this technology has no significant advantage over the current graphic body mark systems. Although this pilot study was only conducted on abdominal ultrasounds, where the standard view protocol is commonly applied, it can be also useful for examining other organs, such as the musculoskeletal joints, peripheral vascular system, and breasts, where information on the inspection site and transducer is more important than the abdomen. Second, there was a trade-off between protection of patient privacy and increasing the accuracy of the inspection site and transducer information. In this study, navigation-loaded images were acquired with a low-resolution thumbnail, focusing on the protection of patient privacy. Besides, this technology uses artificial intelligence to automatically crop sensitive areas, remove the patients’ faces (
8-
10), or completely replace them with a simple body mark according to the inspection site and maintain the appropriate field of view. Nevertheless, considering the patient privacy issues, it is expected to be of limited use in gynecological transvaginal or translabial ultrasounds. To resolve this issue, the camera function in the software should be disabled, followed by a conventional ultrasound examination. This method can be easily applied even if the patient refuses a new body marking system. Third, because camera installation and special software settings are required before starting the ultrasound, this process can be quite cumbersome. To overcome this shortcoming, we plan to develop an embedded system that employs the aforementioned artificial intelligence technology. Fourth, our system is expected to be inaccurate for cine-loop images, as it only provides information on the static ultrasound transducer. However, in the future, we will be able to determine the probe movements via software improvements if necessary. Fifth, although the inter-rater agreement increased using the body navigation-loaded images, the Fleiss’ kappa value did not increase above 0.80 (
Table 3). Therefore, it cannot completely replace the existing annotation methods, such as text or arrows.
In conclusion, according to the results of this pilot study, the body navigation-loaded ultrasound technology is expected to assist with a simple and objective interpretation of ultrasound images of body organs; it is also expected to improve the operator’s convenience by reducing the use of body marks or text mark systems. Moreover, the application of this technology is anticipated for various body organs through further technological advances.