Abstract
Patients and Methods: We proposed and developed a computerized algorithm for wall thickness measurements in 2D echocardiographic image frames. In this program, wall thickness measurement is based on intensity profile function and adaptive bilateral thresholding operation. For validation, thicknesses of septum base and mid segments were estimated in constituent image frames using proposed technique followed by comparing them with conventional manual results from same images of the cardiac cycle by statistical methods.
Results: In our sample image frames (240 corresponding segments; with different range of image quality), a bias of 0.10 mm and 0.12 mm with SD differences of ±0.81 mm and ±0.72 mm and correlation coefficients of 0.87 and 0.89 were found in base and mid segments, respectively. Interobserver variability using the computer-assisted method (CAM) and conventional manual technique (CMM) were 4.0% and 4.7% for the basal and 2.8% and 3.9% for the middle segments.
Conclusion: The method introduced in the present study permits precise thickness assessment of base and mid segments of the interventricular septum wall with high concordance with CMM.
Keywords
Echocardiography Interventricular Septum Wall Thickness Intensity Profile