Background Ionization Radiation in Radiography Centers in Ahvaz, Iran

authors:

avatar Behzad Fouladi Dehaghi ORCID 1 , 2 , avatar Leila Ibrahimi Ghavam Abadi ORCID 3 , *

Department of Occupational Health, Faculty of Health, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
Environmental Technologies Research Center, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
Department of Environmental Management-HSE, Ahvaz Branch, Islamic Azad University, Ahvaz, Iran

How To Cite Fouladi Dehaghi B, Ibrahimi Ghavam Abadi L. Background Ionization Radiation in Radiography Centers in Ahvaz, Iran. Jundishapur J Health Sci. 2020;12(1):e96456. https://doi.org/10.5812/jjhs.96456.

Abstract

Background:

Exposure to ionizing radiation has harmful health effects. Research shows that people spend more their time indoors than outdoors. Therefore, the indoor background ionizing radiation can pose a noticeable health risk.

Objectives:

Since it is well established that ionizing radiation can lead to serious health problems, the present study aimed to evaluate the level of ionizing radiation in the background environment in radiological centers in Ahvaz, Iran.

Methods:

The evaluation of indoor and outdoor background ionizing radiation levels was carried out by using the calibrated digital Geiger-Muller counter (S.E. International Inc., USA) in five radiography centers. The background radiation was measured both indoor and outdoor of the selected radiology centers in four locations. The measured locations included behind the door of the X-ray room, outdoor, waiting room for the people, and the reception section in each center. The measurements were done with a device held one meter above the floor.

Results:

The indoor radiation levels were 0.13 ± 0.004, 0.11 ± 0.004, 0.13 ± 0.004, 0.16 ± 0.007, and 0.16 ± 0.006 µSv/h for centers a, b, c, d, and e, respectively, and the outdoor radiation levels were 0.12 ± 0.02, 0.11 ± 0.01, 0.10 ± 0.00, 0.12 ± 0.01, and 0.13 ± 0.00 µSv/h, respectively.

Conclusions:

The mean equivalent dose in this study was lower than the standard level (1 mSv/y); therefore, the radiology centers were safe.

1. Background

Generally, everyone can come into exposure to ionizing radiation found in the background, regardless of where he lives or works. Ionizing background radiation is present in the environment emitted from various sources (1). Generally, there are two main sources of ionizing radiation, natural sources and anthropogenic sources (1). As stated by the United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR), there are four main sources of natural radiation, including cosmic radiation and terrestrial radiation besides ingestion and inhalation of naturally occurring radionuclides. Radiation found in the background environment is the most important source of exposure for the general population. Human activities have gradually led to the increase of background ionizing radiation (2). Indoor background ionizing radiation is more important than outdoor background ionizing radiation because:

1. Some materials used in the construction contain radioactive elements.

2. Radon, a harmful radioactive gas, can be found in indoor air.

3. Generally, the time spent indoors is much more than the time spent outdoors.

The results of several studies by the World Health Organization (WHO) and the International Commission on Radiological Protection (ICRP) revealed that in a general time schedule basis, 80% of everybody’s time is spent indoors namely at offices, schools, homes, and inside different buildings (3) whereas the other 20% is spent outdoors.

Ionization radiation is used in medicine, research, and industry. It is used as a diagnostic and therapeutic tool to help people suffering from serious illnesses like cancer. Therefore, radiation is the main source of exposure for people in medicine and some related occupations. This has become of great concern regarding the well-established harmful impacts of high dosage of ionizing radiation occurring in hospitals and medical research centers. Patients can come into exposure to ionizing radiation through radiographic examination, radioisotope procedures, and therapies (4). Several studies have found evidence of injuries and symptoms caused by ionizing radiation (5). Some studies also demonstrated that chronic exposure to low doses of ionizing radiations is capable of inducing cytogenetic damage in the exposed person (6). Wall et al. showed that a linear growth of risk can be seen with an increase in exposure to ionizing radiation higher than the background level (7). Garaj-Vrhovac and Kopjar reported significantly higher DNA damage in exposed workers to ionizing radiation than in the control group (8). Therefore, with all this body of evidence, personal monitoring seems to be a vital part of any workers’ health protection program (9). Ionizing radiation has the power to destroy biomolecules and thus produce free radicals (10). Routine monitoring and assessment of the level of exposure to ionizing radiation are essential to keep the workers’ exposure as low as reasonably achievable (ALARA). A study evaluated background ionizing radiation in hospitals and radiology centers and measured ionizing radiation as 0.23 µSv/h and 0.28 µSv/h for outdoor and indoor, respectively (11). A study by Mettler et al. showed the mean background ionizing radiation level as 0.34 µSv/h (12). The mean value for the global natural dose of background ionizing radiation is reported as 0.27 µSv/h (13). Another study reported the background ionizing radiation level in a hospital at an average of 0.146 ± 0.02 μSv/h and 0.1413 ± 0.02 μSv/h for the indoor X-ray department and 0.136 ± 0.02 μSv/h inside the hospital building (14).

2. Objectives

In this study, we assessed the background ionizing radiation levels in radiology centers in Ahvaz to determine the risk of exposure to ionizing radiation.

3. Methods

A calibrated digital Geiger-Muller counter (S.E. International Inc., USA) was used for data collection. The range of doses measured by the detector was 0.01 μSv/h to 1000 μSv/h. The background radiation was measured both indoor and outdoor of the selected radiology centers in four locations. The measured locations included behind the door of the X-ray room, outdoor, waiting room for people, and the reception section in each center. The data were obtained by the gamma spectrometer oriented vertically upward. The data were obtained outdoors by the gamma spectrometer oriented vertically upward. When measuring, the device was held one meter above the floor. In total, 20 readings were taken to cover the areas adequately. The annual indoor and outdoor equivalent doses (mSv/y) were calculated by an equation developed by the UNSCEAR in 2000 (15). The data were analyzed with descriptive statistics and the one-sample t-test by SPSS V. 19. A P value of less than 0.05 was considered significant.

4. Results

The average indoor and outdoor equivalent doses are presented in Tables 1 and 2. Also, the annual equivalent dose was calculated for both indoor and outdoor in five radiology centers (Tables 1 and 2). The maximum permissible limit for non-radionuclide industrial employees and the public is 1 mSv/y as recommended by the ICRP (16). The average indoor equivalent doses in five centers are compared with the standard level in Figure 1. Figure 2 represents the comparison of the mean equivalent doses in various outdoor locations with the standard level. The results showed that the measured levels in indoor and outdoor of all centers were below the maximum permissible limit for the public.

Table 1.

Mean Indoor Dose and Annual Equivalent Dose in Radiography Centers

Radiography CenterDistance from X-Ray Room (m)Mean Dose (µSv/h)Minimum (µSv/h)Maximum (µSv/h)Annual Equivalent Dose (mSv/y)
a20.13 ± 0.0040.100.150.49 ± 0.07
b20.11 ± 0.0040.080.140.42 ± 0.06
c20.13 ± 0.0040.110.170.49 ± 0.07
d2.50.16 ± 0.0070.110.210.58 ± 0.11
e30.16 ± 0.0060.120.210.58 ± 0.11
Table 2.

Mean Outdoor Dose and Annual Equivalent Dose in Radiography Centers

Radiography CenterDistance from X-Ray Room (m)Mean Dose (µSv/h)Minimum (µSv/h)Maximum (µSv/h)Annual Equivalent Dose (mSv/y)
a40.12 ± 0.020.060.550.11 ± 0.09
b40.11 ± 0.010.010.200.10 ± 0.04
c40.10 ± 0.010.060.140.10 ± 0.02
d4.50.11 ± 0.010.090.190.11 ± 0.02
e50.13 ± 0.010.120.160.12 ± 0.01
Comparison of the standard level with the measured indoor levels in all radiology centers
Comparison of the standard level with the measured indoor levels in all radiology centers
Comparison of the standard level with the measured outdoor levels at all radiology centers
Comparison of the standard level with the measured outdoor levels at all radiology centers

5. Discussion

In this study, we measured the indoor and outdoor mean equivalent doses and the annual equivalent doses of background ionizing radiation in five centers in Ahvaz, Iran. The results showed that the mean indoor equivalent doses of background ionizing radiation were maximum in the d and e centers with 0.16 ± 0.007 values while for outdoor, the e center had the maximum mean equivalent dose with 0.13 ± 0.00. The mean doses for both indoor and outdoor measurements were lower than the standard background radiation of 0.25 μSv/h. Also, the highest values for indoor and outdoor measurements were reported in the e center. The minimum mean equivalent doses obtained in indoor and outdoor areas belonged to the b and c centers. The mean equivalent doses were 0.11 ± 0.004 μSv/h and 0.10 ± 0.00 μSv/h for the b and c centers, respectively. These values were slightly higher than the values reported by Foulady et al. (17). It can be because radiology centers do not observe all protection measures correctly. Also, Cardis et al. reported that the medical doses showed a varying trend, which was dependent on technical factors that were ignored in the study (18). The results demonstrated that even the slightest defect in the structural design of the imaging room can lead to higher exposure levels. A study carried out by Adhikari et al. reported the importance of defects in the observed leakage (19). The minimum mean annual equivalent dose obtained in outdoor areas was 0.10 mSv/y and the maximum mean annual dose equivalent was 0.12 mSv/y; the corresponding results were 0.42 mSv/y and 0.58 mSv/y for indoor areas, respectively. In this study, the average annual equivalent doses (mSv/y) were lower than the threshold limit of 1 mSv/y in radiology centers that were similar to the results reported in other studies (14, 19, 20). It should be mentioned that there were some limitations in this study, including the weak cooperation of radiography centers during the study (only 5 out of 25 centers participated voluntarily in the present study) and the lack of knowledge of fundamental principles of X-ray machine layout and assembly in the department building.

5.1. Conclusions

In this study, background ionizing radiation levels were measured both indoor and outdoor in five centers of Ahvaz. The measurements were done in four locations in each center. The study showed that the mean equivalent dose and the annual equivalent dose were within the standard permissible limits set by the ICRP. Therefore, the radiology centers were safe in this regard.

Acknowledgements

References

  • 1.

    Okeyode I, Oluseye A. Studies of the terrestrial outdoor gamma dose rate levels in Ogun-Osun river basins development authority headquarters, Abeokuta, Nigeria. Physics Int. 2010;1(1):1-8. https://doi.org/10.3844/pisp.2010.1.8.

  • 2.

    Hall EJ, Giaccia AJ. Radiobiology for the radiologist. 7th ed. Philadelphia, PA: Wolters Kluwer Health/Lippincott Williams & Wilkins; 2012.

  • 3.

    Thome C, Tharmalingam S, Pirkkanen J, Zarnke A, Laframboise T, Boreham DR. The REPAIR project: Examining the biological impacts of sub-background radiation exposure within SNOLAB, a deep underground laboratory. Radiat Res. 2017;188(4.2):470-4. [PubMed ID: 28723273]. https://doi.org/10.1667/RR14654.1.

  • 4.

    Taqi AH, Al-Ani LAA, Ali AM. Assessment of the natural radioactivity levels in Kirkuk oil field. J Radiat Res Appl Sci. 2019;9(3):337-44. https://doi.org/10.1016/j.jrras.2016.02.007.

  • 5.

    Norman A, Kagan AR. Radiation doses in radiation therapy are not safe. Med Phys. 1997;24(11):1710-3. [PubMed ID: 9394278]. https://doi.org/10.1118/1.597958.

  • 6.

    Fouladi Dehaghi B, Ibrahimi Ghavamabadi L, Boazar M, Ahmadi Angali K, Noori M. Radiation protection in Radiology Departments of Ahvaz University of Medical Sciences Teaching hospitals, Ahvaz, Iran, 2015. Jundishapur J Health Sci. 2015;7(3). e28134. https://doi.org/10.17795/jjhs-28134.

  • 7.

    Wall BF, Kendall GM, Edwards AA, Bouffler S, Muirhead CR, Meara JR. What are the risks from medical X-rays and other low dose radiation? Br J Radiol. 2006;79(940):285-94. [PubMed ID: 16585719]. https://doi.org/10.1259/bjr/55733882.

  • 8.

    Garaj-Vrhovac V, Kopjar N. The alkaline Comet assay as biomarker in assessment of DNA damage in medical personnel occupationally exposed to ionizing radiation. Mutagenesis. 2003;18(3):265-71. [PubMed ID: 12714692]. https://doi.org/10.1093/mutage/18.3.265.

  • 9.

    Lehrer S, Rosenzweig KE. Lung cancer hormesis in high impact states where nuclear testing occurred. Clin Lung Cancer. 2015;16(2):152-5. [PubMed ID: 25458560]. [PubMed Central ID: PMC6587186]. https://doi.org/10.1016/j.cllc.2014.09.010.

  • 10.

    Olisekodiaka M, Bello T, Onuegbu A, Olowookere C, Lebi O, Omotayo L, et al. Evaluation of the serum total antioxidant status level in health workers exposed to low radiation doses in a large Nigerian hospital. Int J Res Rev Appl Sci. 2009;1(2):152-6.

  • 11.

    James IU, Moses IF, Vandi JN, Ikoh UE. Measurement of indoor and outdoor background ionising radiation levels of Kwali General Hospital, Abuja. J Appl Sci Environ Manag. 2015;19(1):89. https://doi.org/10.4314/jasem.v19i1.12.

  • 12.

    Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: A catalog. Radiology. 2008;248(1):254-63. [PubMed ID: 18566177]. https://doi.org/10.1148/radiol.2481071451.

  • 13.

    United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation. UNSCEAR; 2008.

  • 14.

    Okoye PC, Avwiri GO. Evaluation of background ionising radiation levels of braithwaite memorial specialist hospital Port Harcourt, Rivers State. Am J Sci Ind Res. 2013;4(4):359-65. https://doi.org/10.5251/ajsir.2013.4.4.359.365.

  • 15.

    United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation. United Nations Publications; 2000.

  • 16.

    United Nations Scientific Committee on the Effects of Atomic Radiation. Sources, effects and risks of ionizing radiation. UNSCEAR; 2013.

  • 17.

    Foulady B, Ibrahimi GL, Bozar M, Mohamadi A, Ahmadi K. Evaluation of X-ray radiation levels in Radiology Departments of Two Educational Hospitals in Ahvaz, Iran. Iran J Med Physics. 2017;14(2):87-91.