3.1. Clinical Specimens
This prospective study was approved by the institutional review board of Seegene medical foundation, South Korea (IRB No SMF-IRB 2015004), and it was performed in accordance with the Helsinki Declaration of 1975, as revised in 1983. In total, 1489 stool specimens were collected from July 2015 to November 2015. Each specimen was immediately stored at -70°C until further simultaneous analyses.
3.2. Multiplex PCR
To evaluate the quality of the real-time one-step PCR assay, Seeplex Diarrhea-V ACE Detection (DR6411Y, Seegene, Seoul, South Korea), a commercial multiplex viral gastroenteritis assay, was used for comparison. This assay is designed for the detection of the Norovirus GI (NVGI), Norovirus GII (NVGII), Rotavirus A (ROV), Adenovirus F (ADV-F; Serotype 40/41), and Astrovirus (ASV). It consists of reverse transcription and PCR amplification steps, followed by the resolution and detection of amplified DNA products by capillary electrophoresis. Complementary DNA (cDNA) synthesis was performed with 8 µL of extracted nucleic acids using a first strand cDNA synthesis kit (Fermentas, Ontario, Canada), according to the manufacturer’s instructions. The thermal cycling conditions consisted of an initial activation of 15 minutes at 94°C, followed by 40 cycles of 30 s at 94°C, 90 s at 60°C, and 90 s at 72°C. This process was performed using an Applied Biosystems GeneAmp 9700 system (Life Technologies, Carlsbad, CA, USA). The amplification products were detected using agarose gel electrophoresis.
3.3. Nucleic Acid Extraction
Nucleic acid molecules were extracted from 200 µL of each sample using QIAamp DSP DNA Mini Kit (Qiagen, Hilden, Germany), which can be used to extract viral RNA as well. For the Allplex GI-virus Assay, bacteriophage MS2 was added to each sample as an internal control, according to the manufacturer’s instructions.
3.4. In Vitro Transcription
Rotavirus A, Astrovirus and sapovirus (SV) were amplified using the primers for the Allplex GI-virus Assay (GI9701Y, Seegene, Seoul, South Korea). The thermal cycling conditions consisted of an initial activation for 15 min at 94°C, followed by 40 cycles of 30 s at 90°C, 90 s at 60°C, and 90 s at 72°C. The reaction was performed using an Applied Biosystems GeneAmp 9700 system (Life Technologies, Carlsbad, CA, USA). Each PCR product was cloned into a TOPO TA vector (Invitrogen, Carlsbad, CA, USA) containing the T7 polymerase promoter. The complete inserts, including the T7 promoter, were amplified with vector-specific primers. The PCR products were purified and transcribed in vitro using the MegaScript T7 in vitro transcription kit (Ambion, TX, USA), and the obtained RNAs were purified using MEGAclear kit (Ambion, TX, USA). The isolated RNA samples were quantified spectrophotometrically and the copy numbers were calculated.
3.5. One-step Multiplex Real-time PCR Based on the MuDT Technique
The Allplex GI-virus Assay (Seegene, Seoul, South Korea) is a one-step multiplex real-time PCR that permits simultaneous amplification and detection of target nucleic acids of NVGI, NVGII, ROV, ADV-F, ASV, and SV (genogroups G1, 2, 4). For this assay, one-step real-time PCR was performed with 5 µL of nucleic acid extract, using a CFX96 PCR detection system (Bio-Rad, Hercules, CA, USA) under the following conditions: 20 minutes at 50°C and 15 minutes at 95°C, followed by 45 cycles of 95°C for 10 seconds, 60°C for 1 minute, and 72°C for 30 seconds. To analyze two targets in one channel, the fluorescence intensity was measured at 60°C and 72°C. The analysis was performed using the designated software (Seegene Viewer, Seegene, Seoul, South Korea).
3.6. SV Sequencing
Sequencing was performed using the BigDye Terminator v3.1 cycle sequencing kit (Applied Biosystems, Forster City, CA, USA) on an ABI PRISM 3730XL analyzer (Applied Biosystems, Forster City, CA, USA). The obtained sequences were compared with the reference sequences through NCBI BLAST, using the GenBank database (NIH, Bethesda, MA, USA).
3.7. Determination of the Detection Limit of the One-step Multiplex Real-Time PCR Assay
The limit of detection (LoD) range was first estimated and then, a preliminary LoD study was performed for each culture fluid (ZeptoMetrix Corporation, Buffalo, NY, USA) or in vitro transcribed RNA. NVGI (0810086CF), NVGII (0810087CF), and ADV-F (0810085CF) were purchased from ZeptoMetrix. NVGI and NVGII represent recombinant culture fluids.
Culture fluid viral stocks were used in a series of six 10-fold dilutions in a negative matrix, and in vitro transcribed RNAs were used in a series of five 10-fold dilutions in TE buffer. The test was performed using three Allplex GI-virus lots with four replicate tests for each lot. Afterward, the analytical sensitivity of the assay was determined. The results obtained in the LoD estimation study were used to determine the range for the LoD analysis. The last dilution with a 100% detection rate was used to prepare either five or four dilutions that were tested on three Allplex GI-virus Assay lots, in replicates of eight. The results obtained using the three lots were combined, and the detection rate was calculated for each dilution. LoD was defined as the lowest concentration of the target that was detected ≥ 95% of the time.
3.8. Cross-Reactivity with Common Enteric Bacteria, Viruses, and Protozoa
Cross-reactivity with bacteria, viruses, and protozoa was examined using nucleic acids isolated from bacteria and protozoa. The control isolates were obtained from the American Type Culture Collection (ATCC, Manassas, VA, USA), ZeptoMerix (ZMC, Buffalo, NY, USA), Korean Collection for Type Culture (KCTC, Daejeon, South Korea), Korean Culture Center of Microorganisms (KCCM, Seoul, South Korea), the Biological and Emerging Infections Resources Program (BEI Resources, Manassas, VA, USA), and National Culture Collection for Pathogens (NCCP, New York, NY, USA). The cross-reactivity of the Allplex GI-virus Assay was assessed using 38 species of different bacteria, viruses, and protozoa.
3.9. Repeatability Study
Nineteen different samples were prepared consisting of one negative sample and 18 stimulated analytes, which included moderately positive (3 × LoD, positive results obtained 100% of the time), minimally positive (1 × LoD, positive results obtained approximately 95% of the time), and highly negative samples (0.5 × LoD, negative results obtained approximately 20% to 80% of the time). The repeatability of the test was assessed using the data obtained from the LoD calculations. Repeatability was tested using the known positive samples in three different concentrations and a negative control for each target. Each target was tested twice a day for a total of 20 days by a single operator, and duplicates of each panel sample were tested per each run. Within-run, run-to-run, and day-to-day CV% (coefficient of variation) values were analyzed, and the accuracy of the test for each product was evaluated.
3.10. Statistical Analysis
Statistical analyses were performed using SPSS software, version 18 (SPSS Inc., Chicago, IL, USA) and an online tool for assessing statistical agreement (www.john-uebersax.com/stat/raw.htm). Cohen’s kappa value was calculated and used to compare the capacity of viral strain detection between the Allplex GI-virus Assay and Seeplex Diarrhea-V ACE Detection.