In the present study we examined 90 infants, 45 (50%) females and 45 (50%) males. The results of the ultrasound compared to radiography as the gold standard for evaluation of the PICC tip had a sensitivity of 100% and specificity of 89.5%. The ultrasound, with a PPV of 97.3% and NPV of 100%, showed good diagnostic capabilities. In line with these results, a similar study conducted by Telang et al. (
12) estimated the sensitivity (96.55%), specificity (100%), PPV (100%), and NPV (75%) of RTUS.
In 2006, Lanza et al. (
13) conducted a study in Italy on neonates that had an average age of 31.7 days. They reported that the B-mode and Doppler ultrasounds had an 84.6% sensitivity and 100% specificity. The reported NPV value was 97.9% and PPV was 100%. They reported a correlation between ultrasound and chest radiography of 98.1%, which was relatively consistent with the results of our study.
Another study evaluated the catheter pathway by ultrasound in 61 umbilical cord catheters in 60 infants and compared the results with radiography. They reported a sensitivity of 96.4% and specificity of 93.9% for ultrasound, and a sensitivity of 92.8% and specificity of 78.8% for radiography. Ultrasound had a sensitivity of 93.3% and specificity of 95.6% for determining the position of the tip of the catheter, whereas radiography had a sensitivity of 66.7% and specificity of 63% (P < 0.001). Radiographic dysfunction increased in determining the position of the tip of the umbilical cord catheter with increased birth weight (P < 0.005). Their results showed that radiography was reliable in determining the path of the umbilical cord catheter (central or otherwise), but the ultrasound method had a higher priority in determining the position of the tip of the umbilical cord catheter compared to radiography (
14). These findings were in line with their study results.
In the present study, 81.1% of the PICC placements were identified as appropriate by radiography and 78.9% were confirmed by ultrasound. On the other hand, radiography detected 17 (18.8%) cases of inappropriate placement, whereas ultrasound detected 19 (21.1%) cases of inappropriate placement. The findings indicated that ultrasound was more accurate than radiography in determining the PICC placement. These findings were similar to the results of a study by Karber et al. (
15).
Ultrasound assessments after catheter implantation have been shown to identify the depth of the catheter placement points in the heart and portal system. However, radiography has not been able to do this properly (
16,
17). Results of previous studies have indicated that ultrasound can minimize the potential side effects of improper catheter placement and reduce the need for radiography reassessment (
18,
19), which was confirmed in the present study.
The practical and beneficial role of sonography during catheter implantation as a guide in neonatal intensive care has been investigated, and it appears that ultrasound examination can be performed even on premature infants (
20,
21).
In our study, the consistency between the results of radiographic and ultrasound was independent of the underlying variables of sex, gestational age, chronological age, catheter embedding location, embedding indication, underlying disease, and need for surgery. Similar studies have also reported that the results of the two tests are independent of the underlying variables (
11,
12,
22,
23).
Ultrasound sensitivity in the present evaluation was 100%, with a specificity of 89.5%, PPV of 97.3%, and NPV of 100%. Saul et al reported both the sensitivity and specificity of ultrasound as 100% (
24).
At present, radiography is mentioned as the standard gold method in diagnosing the end of PICC catheter in books and articles, and this method is used routinely in most centers. However, the disadvantages of radiography include the following: exposure of the baby to radiation; in case of malposition and correction, need for re-graphing and re-imposition of radiation; sometimes AP and LAT imaging is needed to determine the location of the catheter; creating a static image that requires a proper position on the baby’s torso and limbs is difficult in practice; risks of cumulative radiation dose in patients, staff and other patients present at the NICU due to its bedside. It is now inclined to gradually replace chest radiographs with repeated radiographs in the diagnosis of neonatal respiratory diseases (
1). In this regard, in order to see the end of the catheter, small portable ultrasound devices have been considered for this purpose in the present study.
Advantages of the present study included the larger sample size compared to most of the similar studies. The study conducted by Tauzin et al. (
19) evaluated a series of 109 cases, which was more than that in the current study. Also, our study was performed on 1 and 2 French silicone PICC catheters, which are difficult to visualize with sonography because of their small size. Most studies examined umbilical artery catheters (UAC) and umbilical venous catheters (UVC), which are larger than the PICC (
25-
32). In addition, there were no observed side effects in the 90 studied infants. However, we conducted this study on one population over a limited period of time and in one center, which was a limitation of this study. Future studies that enroll more patients and in several centers are recommended.
5.1. Conclusions
This study showed that the ultrasound had good PPV and NPV in determining the location of the tip of the PICC catheter. Ultrasound appears to be a suitable bedside tool for identifying PICC placement and can prevent babies from exposure to additional X-rays. However, multi-center studies with larger sample sizes need to be conducted.