Using Kalman Filter to Improve the Accuracy of Diffusion Coefficients in MR Imaging: A Simulation Study

Author(s):
Sam Sharifzadeh JavidiSam Sharifzadeh Javidi1, 2,*, Hamidreza Saligheh RadHamidreza Saligheh Rad1, 2
1Physics and Medical Engineering Department, Medicine School, Tehran University of Medical Sciences, Tehran, Iran
2Quantitative Medical Imaging Systems Group, Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran

IJ Radiology:Vol. 16, issue Special Issue; e99153
Published online:Dec 10, 2019
Article type:Abstract
Received:Oct 26, 2019
Accepted:Dec 10, 2019
How to Cite:Sharifzadeh Javidi S, Saligheh Rad H. Using Kalman Filter to Improve the Accuracy of Diffusion Coefficients in MR Imaging: A Simulation Study. I J Radiol. 2019;16(Special Issue):e99153. doi: https://doi.org/10.5812/iranjradiol.99153

Abstract

Background:

The diffusion coefficient of water molecules in different tissues is a biomarker to diagnose and differentiate diseases, tumors, injuries, etc.. Using a motion-sensitizing gradient, it is possible to map the diffusion coefficient into diffusion-weighted MRI (DWI). Although DWI is a good tool for diagnosing, its accuracy in low regime SNR and even especially in high blood consumer organs is in question.

Objectives:

We aimed to improve the accuracy of the diffusion coefficient, especially in the presence of noise.

Methods:

Diffusion of water molecules at each voxel causes a signal intensity decay that can be measured using a motion-sensitizing gradient. Blood perfusion in capillary network artifacts and measurement noise can affect the real amount of D. In light of considering process noise and measurement noise, Kalman filter is used to cancel perfusion artifacts and measurement noise. Based on the diffusion model (S = S0 exp (-b D)), signal intensity was produced several times and a complex Gaussian noise was added to it. Using a Kalman filter, a noise cancelation process was designed to improve the quality of results. The Kalman filter solved a linear problem in the form of state space. Therefore, the diffusion model was rewritten as log(S/S0)/-b = D in state space. The Kalman filter predicted the amount of D and then modified it based on measurements iteratively (Figure 1). Finally, the results of the proposed method and conventional method were compared with true values (Figure 2).

Results:

Statistical tests showed that the proposed method was significantly better than the conventional method (P < 0.01). The conventional method caused a bias in the results of DWI due to eliminating micro-vessel perfusion in the capillary network. However, Kalman filter could consider the effects of microvessel perfusion as a process noise and reduce its effects on results. Kalman filter results (Table 1) were compared with true values and the t-test showed no significant alteration (P = 0.25). Conventional method results were significantly different from true values and Kalman filter results (P < 0.0001).

Conclusion:

Diffusion coefficients in the presence of noise and capillary network suffer from bias; however, the proposed method can be used in these situations to improve the quality of DWI images.

To see figures, table, and references, please refer to the PDF file.

Similar Articles

28
Feb
2016

Brain Tissue Classification Based on Diffusion Tensor Imaging: A Comparative Study Between Some Clustering Algorithms and Their Effect on Different Diffusion Tensor Imaging Scalar Indices

Ihab Elaff

Elaff I. Brain Tissue Classification Based on Diffusion Tensor Imaging: A Comparative Study Between Some Clustering Algorithms and Their Effect on Different Diffusion Tensor Imaging Scalar Indices. I J Radiol. 2016;13(2):e23726. doi: https://doi.org/10.5812/iranjradiol.23726

8
Jan
2018
Calculating eGFR Using Apparent Diffusion Coefficient (ADC) Values Obtained Through MR Imaging

Calculating eGFR Using Apparent Diffusion Coefficient (ADC) Values Obtained Through MR Imaging

Jaturat Kanpittaya,
Patrawan Jaimook,
Theerayut Thongkrau,
Chaiyapon Keeratikasikorn,
Kittisak Sawanyawisuth

Kanpittaya J, Jaimook P, Thongkrau T, Keeratikasikorn C, Sawanyawisuth K. Calculating eGFR Using Apparent Diffusion Coefficient (ADC) Values Obtained Through MR Imaging. I J Radiol. 2018;15(1):e13682. doi: https://doi.org/10.5812/iranjradiol.13682

12
Apr
2017

The Importance of Diffusion Weighted Imaging in Breast MRI

Maryam Noori,
Bahareh Siahlou,
Hamidreza Salighehrad,
Anahita Fathi,
Mohammad Fathi

Noori M, Siahlou B, Salighehrad H, Fathi A, Fathi M. The Importance of Diffusion Weighted Imaging in Breast MRI. I J Radiol. 2017;14(Special Issue):e47964. doi: https://doi.org/10.5812/iranjradiol.47964

16
Jan
2016

Additional Value of Diffusion-Weighted Imaging to Evaluate Prognostic Factors of Breast Cancer: Correlation with the Apparent Diffusion Coefficient

Eun Kyung Park,
Kyu Ran Cho,
Bo Kyoung Seo,
Ok Hee Woo,
Sung Bum Cho,
Jeoung Won Bae

Park EK, Cho KR, Seo BK, Woo OH, Cho SB, et al. Additional Value of Diffusion-Weighted Imaging to Evaluate Prognostic Factors of Breast Cancer: Correlation with the Apparent Diffusion Coefficient. I J Radiol. 2016;13(1):e33133. doi: https://doi.org/10.5812/iranjradiol.33133

30
Dec
2009

Improvement of White Matter Fiber Tracking Based on Diffusion- Tensor MR Imaging Data Using Modified Speed Functions

F. Darki,
Mohammad Oghabian

Darki F, Oghabian M. Improvement of White Matter Fiber Tracking Based on Diffusion- Tensor MR Imaging Data Using Modified Speed Functions. I J Radiol. 2009;6(4):e79457. doi:


Crossmark
Crossmark
Checking
Share on
Cited by
Metrics

Ordering Reprints

Articles are published under the Creative Commons license stated on each article. No permission or royalty fee is required for uses permitted by that license. CCC handles optional bulk and customized reprint orders. Any quotation covers production and delivery services only, not copyright permission. > Request Reprints from CCC 

Search Relations

Author(s):

Related Articles