Abstract
Methods: A cross-sectional study was conducted January2008 - May 2009 in Tehran. 1000 newborns (526 boys and 474 girls) were assessed. First, all of neonates were evaluated by TEOAE 24h after birth. If responses of OAE were failing, they were retested 10 to 15 days after birth by TEOAE. Also, All Neonates were assessed by ABR in the age of 3 months. Descriptive Statistics was used to analyze data.
Findings: Eighteen out of 1000 neonates failed double–checked TEOAE tests, of which 6 were confirmed by ABR test (12 false positive results). Nine out of 1000 neonates had impaired ABR tests, from these patients, 6 had failed OAE as well, but 3 had normal OAE (3 false negative results). From these 9 patients 2 had profound hearing loss and received cochlear implantation. We found that OAE has 66.7% sensitivity and 98.8% specificity in diagnosis of neonatal hearing impairment. Its positive and negative predictive value was 33.3% and 99.7% respectively. Also we did not find statistically significant relationship between hearing loss and risk factors.
Conclusion: TEOAE as a simple, non-invasive, short duration and cost effective method, is a suitable test for neonatal hearing screening. Even though only two thirds of patients were detected by this method, 99.7% negative predictive value makes it a good screening test. We recommend OAE as a suitable primary neonatal hearing screening all over the country.
Keywords
Hearing Loss Sensitivity Specificity Auditory Brain Stem Response Otoacustic Emission
Fulltext
References
-
1.
References are available on the PDF.