Spectroscopic study
The absorption spectrum of pC was determined using an ultra-violent technique, where results showed that the absorption peak is around 280 nm (
Figure 1). To approve the absorption (excitation) and emission wavelengths, the excitation and fluorescence emissions were investigated by the fluorimetry technique. As
Figure 1 shown, pC excites at 280 nm and emits fluorescence at 310 nm. The fluorescence emission is linear with pC concentrations. The same excitation and emission were observed in both plasma and standard media, approving the independence of the fluorescence emission in complex plasma matrices.
Factors affecting the fluorescence intensity of the system
To regulate the best conditions for increasing the sensitivity of the method toward a specific analyte, the optimization step of the parameters is of great importance. Firstly, the proper slit sizes of emission and excitation were selected. It is important to select the proper slit because of its effect on the fluorescence of analyte and coexisting agents. In this study, 5 nm was selected for both excitation and emission slits. Due to the fact that pH has the main effect on the hydrolysis of pCS to pC, the acidity and basicity value of the hydrolysis reaction were also investigated. Accordingly, the same concentrations of pC at three pH values (acidic, neutral and basic) and their blanks were prepared and then fluorescence emissions were measured. As
Figure 2 shows, the acidic medium showed the highest fluorescence intensity, where the conversion of pCS to pC was strongly dependent on the pH of medium and reached a maximum value at a pH value of about 1.5. The pHs of solutions were adjusted by HCl, NaOH and phosphate buffer solution (PBS).
Various agents were checked for the effective protein precipitation of plasma samples. ACN was selected as the best protein precipitant and then the ACN/plasma ratio was optimized as 1:9. Although the higher sample volume means the higher extraction of the analyte, fluorescence emissions of coexisting agents overlap with pC fluorescence spectrum. The effect of reaction time was also tested. Obtained results revealed that 2 min is the optimal reaction time to convert all pCSs to pC. In lower and higher time periods, the fluorescence intensities dropped because of reaction incompletion and possible secondary reactions of pC, respectively. Also, the effect of shaking time was optimized and 15 min selected for completion of extraction of pC from plasma samples.
The salting-out ability of different salts was tested by introducing various salts into the ACN-extracted solution. Obtained results proposed NaCl as the most efficient agent for convincing salting-out force to the pC molecules for migration from the aqueous phase to the organic phase. Further experiments showed that saturated NaCl had the maximum effect on the extraction of pC.
After optimizing the parameters, the effect of time of analysis was investigated on the fluorescence emission of pC. The results obviously showed that fluorescence emission was constant for the first 5 min. All measurements were carried out immediately after the extraction step. Room temperature was selected as the optimal temperature for recording the fluorescence spectra since temperature control systems are not available in all fluorescence sets.
Analytical figures of merit
A simple fluorescence method was developed for the measurement of pC in plasma samples based on the intrinsic fluorescence emission of pC at 310 nm. After optimization of the influence of parameters, the calibration curve was plotted in the plasma matrix to discover the dynamic range and other figures of merit. To plot the calibration curve, various concentrations of pC were prepared and then linear range, LLOQ, precision and accuracy of the method were calculated. In the plasma matrix, various concentrations of pC were added to the 100 µL sample and then extracted as reported in subsection 2.3. The fluorescence emission was enhanced by increasing pCs ranging from 0.5 to 30 µg/mL. The fluorescence difference between blank and analyte was regarded as a signal to plot the calibration curve. The regression equation for plasma was ∆F = 8.9645 (C (µg/mL)) + 1.89 with R
2 of 0.9992 and the relative standard deviation (RSD%) of 1.2%. LOQ, LLOQ and LOD of the developed approach were 0.6, 0.5 µg/mL and 0.2 µg/mL, respectively. The high Y-intercept of the developed method is mainly caused by the fact that serum samples of healthy subjects possess some pC (or pCG and pCS) in the range from 3.65 to 56.06 µM (0.34 to 5.3 µg/mL) (
3).
Figure 3 illustrated the calibration curve and corresponded fluorescence spectra.
A summary of figures of merit of the developed method and comparison with the previously reported methods for the determination of pC are collected in
Table 1, respectively (
10,
17-
26). To date, the fluorescence technique has not been utilized for the quantification of pC in biological samples. Compared with the previously reported approaches, the developed method has a relatively simple procedure with comparable sensitivity. However, better sensitivities were obtained with more expensive and complex techniques such as liquid chromatography-mass spectroscopy/mass spectroscopy (LC-MS/MS) and GC-MS/MS. Although reported methods can provide better sensitivities, they often need complicated sample preparation steps with consuming hazardous solvents and high-cost instruments. Typically, highly skilled staff are required to work with GC-MS/MS and LC-MS/MS and interpret the results.
Specificity, repeatability and accuracy of the method
The specificity of a method is a crucial topic in the detection of an analyte in biological media. Specificity is defined as the response of a developed method towards a specific analyte in the presence of possible coexisting agents. In other words, it refers to the unequivocal determination of a proper amount of an analyte in the presence of various coexisting agents. Specific methods detect and differentiate the target analyte from any other non-target agents in the presence of various interfering macro-molecules and ions.
To quantitatively test the specificity of the developed method for the determination of pC, pC was determined in the presence of several agents. However, in the case of the analysis of the patient’s fluids, the effect of consumed pharmaceuticals on the signal of pC should be checked.
Table 2 lists the effects of prescribed drugs on the fluorescence intensity of the probe. Results showed that the developed method was highly specific towards pC in plasma samples. Some interfering agents have changed the original fluorescence intensity of pC, which mainly corresponded to the higher concentrations of the agents, approving the specificity of the developed method.
According to the FDA guidelines, repeatability of a method is the ability of the developed method to produce similar results for a given concentration of analyte, which is mainly associated to the precision of the method and stated in RSD%. The repeatability of the developed method was investigated by three concentrations of pC from three quality control (QC) levels of low (LQC), medium (MQC) and high (HQC). After processing extraction steps, fluorescence emissions were recorded on the same day (or different days) and the repeatability was expressed in RSD%. Repeatability is expressed in two forms of interday and intraday. Results revealed that the developed method behavior is repeatable for both intraday and interday assays and all RSD% values were reported lower than 15% as allowed by FDA. The repeatability of the developed method is reported in
Table 3.
The accuracy of a method for a specific analyte is the maximum allowed differences between the actual and the calculated values. Based on, the fluorescence emission of the three concentrations of pC from three points of the calibration curve was measured and then the recovered amounts of pC were calculated by calibration curve to estimate the accuracy of the method for pC determination in plasma samples. The recoveries of the developed method for 1, 5 and 20 µg/mL concentrations of pC were 112.8, 107.0 and 99.2%, respectively, approving the acceptable accuracy of the developed method for the determination of pC in plasma samples (
Table 4).
Stability
The stability of an analyte in a biological matrix is an important factor that affects the obtained results. The room temperature and freeze-thaw stability are two important stability factors and determined as follows. To check room temperature stability, three concentrations of pC were prepared and left for 24 h at room temperature and then fluorescence intensities were measured with the developed approach (
Table 5). For freeze-thaw stability, three concentrations of pC were prepared and then freeze-thawed for three cycles. After the third cycle, samples were immediately analyzed with the developed method (
Table 5). Results showed that both stabilities are acceptable and their standard deviations are lower than 20% as allowed by FDA guidelines (
15).
Also, stock solution stability of pC was measured at room temperature and in a refrigerator. To do this, 5 µg/mL of pC was prepared in two sets and one set was left at room temperature and another stored in a refrigerator for 24 h. Then, the stability of pC was tested and compared with a freshly prepared solution of pC. Results approved that pC stock solution is stable at both conditions (
Table 6).
Application of the developed method for pC detection in patients’ real samples analysis
The application of the developed method for the determination of pC concentration in patients’ plasma samples was evaluated under optimal conditions. Prior to analysis with the method, blood samples were taken from CKD patients from Imam Reza Hospital and then centrifuged to separate red blood cells. Then, they were stored at -20 °C till analysis with the developed method. Under optimal conditions, samples were thawed and fluorescence emissions were recorded and finally, concentrations were calculated using calibration curve equation. Obtained results are provided in
Table 7.
Capability to be as a routine analytical procedure and a POCT device
Daily routine analysis of biological agents is important to understand the patients’ condition and prescribe medicines with proper dosages. Reliable analytical methods with high specificity, precision, and accuracy can be routine analysis methods in clinical laboratories. To qualify an analytical method as a “routine analytical method”, it is important to check its performance in large numbers of biological samples.
It is emphasized that the developed method could be applied as a “routine analytical method” for the determination of pC in plasma samples because of its favorable analytical features. The reported method is very simple, specific, sensitive, cheap and fast, which are the perquisites for a “routine analysis method” and also it showed high repeatability (interday and intraday) and stability.
Absorption (dashed line), excitation (dotted line) and emission (solid line) spectra of pC (5 µg/mL) in absolute ACN (room temperature).
(A) Effect of pH on the hydrolysis of pCS and PCG to pC, (B) effect of type of protein precipitant and (C) salting-out agent fluorescence emission. (conditions: (A) HCl, NaOH, and PBS concentrations are 0.01 M, and temperature, 80 C (B) ZnSO4 in saturated form, ACN, MeOH in absolute form and HCl 36%, room temperature, and (C) all salts are saturated at room temperature)
(A) Fluorescence spectra and (B) corresponded calibration curve of pC at various concentrations from 0.5 to 30 µg/mL.
| Method | Sample | Approach | LOD* (µg/mL) | Linear range (µg/mL) | Reference |
|---|
| LC-FL | Serum | Extraction | 0.14 | 0–50 | (17) |
| GC-FID | Whole blood | Extraction | 0.6 | 2–20 | (18) |
| LC-FL | Plasma | Extraction | 0.324 | 0.5–8.0 | (10) |
| LC-FL | Urine | Extraction derivatization | 0.018 | 0.01–540 | (19) |
| UHPLC-MS/MS | Serum | Extraction | 0.25 | 0.25–80 | (20) |
| GC-MS | Serum | Extraction | LOD: 0.15LOQ: 0.30 | - | (21) |
| LC-MS/MS | Plasma | Extraction | 0.85 ng/mL | 0.85–200 | (22) |
| HPLC | Serum | Extraction | 0.25 | 0.5–50.0 | (23) |
| UPLC-MS/MS | Serum | Extraction | 0.05 | 0.05–5 | (24) |
| UPLC-FL | Serum | Extraction | 0.005 | 0–4.52 | (25) |
| Electrochemical | Serum | - | 1.1 | 108–2700 | (26) |
| FL | Plasma | Extraction | 0.5 (LLOQ)0.2 (LOD)0.6 (LOQ) | 0.5–30 | This work |
| Drug | Amounts of co-prescribed drug and tested concentrations (µg/mL) | Intensity* | Signal Change (%) |
|---|
| Prozosin | 0.05 | 56 | 102.3 |
| Amlodipine | 0.02 | 57 | 104.2 |
| Simvastatin | 0.03 | 57 | 103.4 |
| Folic acid | 0.015 | 49 | 89.3 |
| Propranolol | 0.3 | 60 | 109.9 |
| Metoprolol | 0.5 | 63 | 113.7 |
| Furosemide | 6.0 | 59 | 106.4 |
| Bisoprolol | 0.1 | 58 | 106.1 |
| Captopril | 0.5 | 58 | 104.6 |
| Valsartan | 6.0 | 56 | 102.3 |
| Nitroglycerin | 0.015 | 51 | 92.0 |
| Lisinopril | 0.14 | 51 | 92.7 |
| Gemfibrozil | 25.0 | 54 | 98.5 |
| Enalapril | 0.1 | 54 | 98.5 |
| Vazonidin | 0.02 | 56 | 101.1 |
| Rouuvatatin | 0.16 | 52 | 93.9 |
| Lovastatin | 1.0 | 56 | 102.3 |
| Nominal concentration (µg/mL) | Intraday precision (RSD%) | Interday precision (RSD%) | Intraday accuracy (RE%) | Interday accuracy (RE%) |
|---|
| 1.00 | 2.50 | 3.79 | -12.83 | -16.56 |
| 5.00 | 6.75 | 5.76 | -7.02 | -8.13 |
| 20.00 | 6.28 | 4.15 | 0.79 | -5.34 |
| Nominal concentration (µg/mL) | Found concentration (µg/mL) | Recovery (%) | Nominal concentration (µg/mL) |
|---|
| 1.0 | 1.13 | 112.8 | 1.00 |
| 5.0 | 5.35 | 107.0 | 5.00 |
| 20.0 | 19.84 | 99.2 | 20.00 |
| Nominal concentration(µg/mL) | Room temperature | Freeze-thaw |
|---|
| Mean found (µg/mL) | Accuracy(RE%) | Mean found (µg/mL) | Accuracy(RE%) |
|---|
| 1.0 | 1.1 | -9.1 | 1.1 | -9.1 |
| 5.0 | 5.5 | -10.4 | 5.5 | -9.6 |
| 20.0 | 20.2 | -0.8 | 20.0 | 0.1 |
| Concentration (µg/mL) | Immediately analyzed (µg/mL) | 24 h (RT*) (µg/mL) | 24 h (4-8 C) (µg/mL) |
|---|
| 5 | 4.98 | 4.56 | 4.71 |
| Patient (#) | Gender | Age | Measured concentration (µg/mL) |
|---|
| 1 | Female | 84 | 1.71 |
| 2 | Female | 57 | 3.43 |
| 3 | Female | 50 | 3.21 |
| 4 | Male | 78 | 2.08 |
| 5 | Male | 65 | 3.10 |