The electrochemical behaviors of the modified electrode
Figure 1A displays the cyclic voltammograms of 0.1 mM of 4,4′-biphenolat different pH values. As can be seen, 4,4′-biphenolat cyclic voltammogram (B) depicts an anodic (A
1) and its corresponding cathodic peak (C1) in the positive- and negative-going scans at each pH value, respectively.
This can be attributed to 4,4′-biphenolat transformation into 4, 4′-diphenoquinone (Q) and vice versa under a quasi-reversible electron-transfer condition (
27-
28).
Upon increasing the pH, the activation overpotential was seen to reduce and a shift towards negative oxidation peak potentials was resulted due to the participation of protons in the B-to-Q oxidation reaction as follows:
B Q + ne- + mH+ (1)
where m represents the number of protons taking part in the reaction. The half-wave potential is calculated as follows:
E1/2 = Eº1/2 – (2.303 mRT/2F) pH (2)
where Eº
1/2 stands for the half-wave potential at pH 0.0 and m, R, T, F are associated with their typical values. The cyclic voltammogram also demonstrated the average of anodic and cathodic peak potentials as calculated below (
29):
E1/2 = (Epa + Epc)/2 (3)
This plotting E1/2 vs. pH can be illustrated through the following Equation 4:
E1/2 = -0.057 pH + 0.705 R² = 0.993 (4)
Being close to the theoretical value of 59 mV, a slope of 57 mV pH
-1 within the pH range of 2.0-11.0 suggested the 2-electron and 2-proton processes of the electrode surface reaction as exhibited by
Figure 1B,
Scheme 1, and Equation 4.
As shown in
Figure 2A, the different scan rates of B cyclic voltammograms occur in the presence of CySH in a buffer solution at pH 10.0. According to
Figure 2B, a linear relationship exists between the plot of IpA
vs. v
1/2 in the range of 5-200 mVsec
-1.
Moreover, by increasing the potential scan rate, the peak current ratio (I
pC1/I
pA1) is enhanced, which is due to the insufficient reaction time for CySH and B (
Figure 2C) (
27).
Scheme 1 portrays how the process of electron transfer leads to a catalytic reaction.
The current research aimed at examining the cyclic voltammograms of 2 mM of CySH on GCE at a scan rate of 10 mVs
-1 and a pH value of 10 (
Figure 3, Curve a).
Figure 3 (Curve c) demonstrates the cyclic voltammogram of 0.1 mM of B at the presence of 2 mM of CySH. After comparing the cyclic voltammograms of B in the presence and absence of CySH (Curve b), a significantly greater anodic peak current (I
pA1) was clearly observed in the latter case.
The electrochemical behaviors of 0.1 mM of B were investigated in the presence of 0.5 mM of CySH at the scan rate of 10 mVsec
-1 at different pH values associated with various buffers (
2-
11) through LSV (
Figure 4A). The gradual growth of the oxidation peak current (Ip
A1) induced by pH enhancement up to a value of 10.0 and its subsequent reduction are displayed in
Figure 4B. It was expected that the raised pH level of the solution provoke the anodic peak potential (E
pA1) shift towards negative potentials (
Figure 4C) (
29).
Analytical measurements
Figure 5A shows the cyclic voltammograms of B at various concentrations of CySH oxidation on GCE for the investigation of the electrocatalytic activity of B.
Figure 5B shows the relevant calibration curve. As mentioned above, the anodic peak current (I
pA1) was increased by CySH concentration enhancement, which further showed a linear calibration range of 10-1000 μM with R² = 0.993. This indicated the complete fitness of the regression line to the experimental data. Therefore, the regression equation (I/µА = 0.005[CySH] + 0.484) could be used for determining the unknown samples. Detrmination of CySH Limit of Detection (LOD) was based on LOD definition of 3 Sb/m, in which S
b is the blank Standard Deviation (SD) (n = 7) and m is the calibration graph slope. Consequently, CYSH LOD was found to be 0.99 μM. Furthermore, its reproducibility was measured to be 100 μM based on the cyclic voltammograms. The Relative Standard Deviation (RSD) was calculated to be 2.5% for 7 replicates.
The voltamvmetric determination of CySH is done by comparing this approach with the previously reported methods as shown in
Table 1. Our analytical parameters of CySH determination clearly demonstrated comparable or better results than those of the previous reports.
Chronoamperometric studies
In addition to the above methods, the chronoamperometry method was utilized for studying the electrochemical behaviors of CySH in 4,4′-biphenol (B) solution at the GCE.
Figure 6A shows double-step chronoamperograms both in the absence (a) and presence of CySH, representing the mediator (4,4′-biphenol (B) solution) at the GCE. The potential steps applied in this study included the values of 300 and -10 mV
vs. Saturated Calomel Electrode (SCE). These selected potentials were more positive and negative than the anodic and cathodic peak potentials when completing the oxidation and reduction processes at the electrode surface, respectively, while both the anodic and cathodic reactions could be controlled in the diffusion process. The presence of CySH leads to a reduced cathodic current corresponding to the decrease in 4,4′-biphenol (B). Furthermore, a linear dependency was observed when plotting the net current
vs. the negative square root of time.
Therefore, the main process of the diffusion control was established. The diffusion coefficient value (D) was obtained by using the slope of the following line based on Cottrell′s equation (
Figure 6B) (
29):
I = nFAD1/2Cπ−1/2t−1/2 (5)
where D and C denote the diffusion coefficient (cm
2 sec
-1) and bulk concentration (mol cm
−3), respectively. The diffusion coefficient of CySH was calculated to be 8.79 × 10
-6 cm
2 sec
–1. In addition, the catalytic rate constant (k) was measured for the chemical reaction of CySH with 4,4′-biphenol through the chronoamperometric analysis (
27-
28):
IC/IL = π1/2 γ1/2 = π1/2 (kC0t) 1/2 (6)
where I
C and I
L stand for 4,4′-biphenol (B) currents in the presence and absence of CySH, respectively; k represents the catalytic rate constant (mol
–1 L sec
–1); t shows the elapsed time (s); and C
0 demonstrates bulk cysteine concentration (mol L
–1). A linear dependency was shown by plotting the current ratio (I
C/I
L) in relation to the square root of time. k
cat value was measured to be 6.17 × 10
3 M
-1 sec
-1 based on the slope of this line (
Figure 6C).
Measurements of CySH concentrations in human plasma and commercial tablets were done to determine the applicability of CySH as an electrochemical sensor. CySH can be simply and rapidly assessed via the standard addition method.
The practical usefulness of the prepared CySH sensor was evaluated by analyzing it in various human plasma samples and commercial tablets. A specified amount of CySH was used in the spiking drugs to measure its recovery in the samples and tablets. As displayed in
Table 2, the analysis results of CySH contents in the real samples revealed that 4,4′-biphenol (B) could serve as a relatively good redox mediator for the voltammetric determination of CySH in the different matrices of the real samples, indicating the reasonable rates of recovery and reproducibility.
Interference studies
The major challenge in interference studies is that the presence of ascorbic acid (AA) and uric acid (UA) makes it difficult to determine CySH. The selectivity of the prepared sensor was evaluated in the presence of 4,4′-biphenol considering its response to AA, UA, and the materials listed in
Table 3, in which the tolerance limit was defoned as a range of relative errors in determining the maximum concentration of troublesome materials, which fall around 5 percent. The results indicated that 40 μM CySH was just the response detected to the prepared sensor, without any trace of the other species.