Validating a Stability Indicating HPLC Method for Kinetic Study of Cetirizine Degradation in Acidic and Oxidative Conditions

authors:

avatar Effat Souri , * , avatar Ali Hatami , avatar Nazanin Shabani Ravari , avatar Farhad Alvandifar , avatar Maliheh Barazandeh Tehrani

Department of Medicinal Chemistry, Faculty of Pharmacy and Pharmaceutical Sciences Research Center, Tehran University of Medical Sciences, Tehran, Iran.

How To Cite Souri E, Hatami A, Shabani Ravari N, Alvandifar F, Barazandeh Tehrani M. Validating a Stability Indicating HPLC Method for Kinetic Study of Cetirizine Degradation in Acidic and Oxidative Conditions. Iran J Pharm Res. 2013;12(2):e125693. https://doi.org/10.22037/ijpr.2013.1312.

Abstract

A stability indicating High-Performance Liquid Chromatography (HPLC) method was validated and used to study the degradation of cetirizine dihydrochloride in acidic and oxidative conditions. The separation was carried out on a Symmetry C18 column and a mixture of 50 mM KH2PO4 and acetonitrile (60:40 v/v, pH = 3.5) was used as the mobile phase. The method was linear over the range of 1-20 μg/mL of cetirizine dihydrochloride (r2 > 0.999) and the within-day and between-day precision values were less than 1.5%. The results showed that cetirizine dihydrochloride was unstable in 2 M HCl and 0.5% H2O2. The kinetics of the acidic degradation showed a pseudo-first-order reaction in the temperature range of 70-90°C. In addition, the kinetics of hydrogen peroxide mediated degradation was pseudo-first-order in the temperature range of 50-80°C.

Introduction

Cetirizine dihydrochloride, chemically known as (RS)-2-[2-[4-(4-chlorophenyl) phenylmethyl] piperazin-1-yl] ethoxy] acetic acid dihydrochloride (Figure 1), is a non-selective antagonist of H1-receptor. Cetirizine dihydrochloride is a potent and non-sedating antihistamine belongs to the piperazine class of second generation of antihistamines. Cetirizine dihydrochloride is used for symptomatic treatment of allergic conditions including seasonal rhinitis and chronic urticaria (1).

Literature survey showed several reported High-Performance Liquid Chromatography (HPLC) (2-9), capillary zone electrophoresis (10) or spectrophotometric (2, 11) methods for determination of cetirizine dihydrochloride in pharmaceutical dosage forms either alone or in combination with other drugs. In addition, a potentiometric method is reported in BP for determination of cetirizine dihydrochloride in bulk powder (12).

According to the ICH guidelines, the stability of drug substances should be studied in different conditions. There are some reports in the literature about the stability of cetirizine dihydrochloride in different conditions. According to Makhija et al. (13), no degradation was observed for cetirizine in combination with pseudoephedrine under acidic (1 M HCl) or basic (1 M NaOH) conditions at 70°C after 2 h. On the other hand cetirizine was unstable under oxidative conditions. By using 0.1 M HCl, 0.1 M NaOH or 1% H2O2 at 80°C for 10 h, it has been concluded that cetirizine dihydrochloride was stable in basic condition but unstable in acidic or oxidative conditions (4). There is another report regarding the stability of cetirizine dihydrochloride under 0.5 M HCl or 0.5 M HCl at 80°C after 4 h (7). Using 1 M HCl or 30% H2O2 decomposition of cetirizine dihydrochloride was observed after 12 h at 80°C (8). To the best of our knowledge, these reports are at a descriptive level and there is no research published in the literature in regard to the kinetics of degradation of cetirizine dihydrochloride under acidic, basic or oxidative conditions.

The aim of this study was to validate an HPLC method according to the ICH guidelines to be performed for kinetic study of the degradation of cetirizine dihydrochloride under acidic, basic and oxidative degradation processes.

Experimental

Materials

Cetirizine dihydrochloride was from Auctus Pharma Limited (Unit-II), India (Batch No, CZH-10 01 001). All the HPLC grade solvents and analytical grade chemicals were purchased from Merck (Darmstadt, Germany). HPLC grade water was obtained by a Milli-Q purification system (Millipore, Milford, MA, USA).

Instrumentation

A Waters HPLC system consisted of an isocratic pump (Model 515), an autosampler (Model 710 plus) and a variable UV-vis detector (Model 480) was employed. The HPLC data were processed by using a multi-channel Chrom and Spec software for chromatography, version 1.5 x.

Chromatographic conditions

Chromatographic Conditions was partly based on the method reported by Zaater et al. (14). A Symmetry® C18 5 μm column (4.6 mm × 150 mm, Waters) and a mixture of KH2PO4 50 mM and acetonitrile (60:40 v/v, pH = 3.5) at a flow rate of 1 mL/min were used. The mobile phase was prepared daily and degassed by filtration through a 0.45 μm Teflon membrane filter (Millipore, Milford, MA, USA) and sonication for 10 min. The wavelength for UV-detection was 230 nm and all determinations were performed at ambient temperature.

Standard solution

Stock standard solution of cetirizine dihydrochloride was prepared by dissolving an accurately weighed portion of drug in water to reach a concentration of 5000 μg/mL. The stock standard solution was kept at 4°C protected from light. Standard solutions for calibration curve (1, 2, 5, 10, 15 and 20 μg/mL) were prepared daily by appropriate dilution of stock standard solution with mobile phase to reach the desired concentration.

System suitability

System suitability parameters of the HPLC method were assessed by six replicate injections of a solution of cetirizine dihydrochloride (10 μg/mL in mobile phase) to the HPLC system. The coefficient of variations for the peak areas and retention times were calculated.

Linearity

Six series of cetirizine dihydrochloride solutions in mobile phase at the concentrations of 1, 2, 5, 10, 15 and 20 μg/mL were prepared and injected to the HPLC system. The calibration curves were constructed by plotting the measured peak areas of each concentration against the corresponding concentration and the statistical analysis was performed.

Precision and accuracy

Cetirizine dihydrochloride solutions at three different concentration levels within the calibration range (1, 5 and 20 μg/mL of mobile phase) was injected to the HPLC system in triplicate. The concentration of each solution was measured using a calibration curve in the range of 1-20 μg/mL. The within-day precision and accuracy was calculated. The same method was performed during three separate days to find out the between-day precision and accuracy.

Stability

The stability of cetirizine dihydrochloride stock solution was checked after storing at 4°C for 1 week.

Application of the method

Twenty cetirizine dihydrochloride tablets were weighed and ground into fine powder using a glass mortar and pestle. An accurately weighed portion of powder equivalent to 10 mg cetirizine was transferred to a 100 mL volumetric flask and 50 mL of mobile phase was added. After 15 min sonication the volumetric flask was adjusted to the volume with the same solvent. The solution was injected to the HPLC system after filtration through a 0.45 μm polypropylene syringe filter (Teknokroma, Spain) and ten times dilution. The drug concentration was determined in comparison with the same concentration of a standard solution of cetirizine dihydrochloride.

Recovery

To find out the relative recovery of cetirizine from dosage forms, an accurately weighed amount of tablet powder equal to 50% of one tablet was spiked with cetirizine dihydrochloride standard solution in a 100 mL volumetric flask. The same procedure for determination of drug dosage form was performed. The peak area of the triplicate injections was compared by a standard solution of cetirizine dihydrochloride with the same concentration level and the relative recovery was calculated.

Kinetic investigation of cetirizine degradation

For acidic degradation, 1 mL of stock standard solution of cetirizine dihydrochloride (5000 μg/mL) was transferred into a 10 mL volumetric flask and 2 M HCl was added to volume. The flask was placed in a dry air oven (Melag, Germany) at four different temperatures (70, 80, 85 and 90°C). At specified time intervals, 500 μL of the solution was transferred to a 10 mL volumetric flask and after neutralization with sodium hydroxide, adjusted to the volume by mobile phase. The resulting solution was injected to the HPLC system and the peak area of cetirizine was compared with a freshly prepared standard solution. The percentage of remained cetirizine was calculated and plotted against time. Each experiment was repeated three times at each temperature.

For basic degradation, the same experiment was performed using 5 M NaOH and for neutralization, a solution of hydrochloric acid was used.

Oxidative degradation was also performed using the same procedure by using 0.5% hydrogen peroxide at four different temperatures (50, 60, 70 and 80°C).

Results and Discussion

Chromatographic conditions

As mentioned in the experimental section, the chromatographic conditions were partly based on a previously published method (14). Using a Symmetry C18 column and a mixture of KH2PO4 50 mM and acetonitrile (60:40 v/v, pH = 3.5), good peak shape without tailing and interferences from degradation products in acidic or oxidative conditions, was observed (Figure 2a). The system suitability parameters showed in Table 1 were within the acceptable criteria.

Typical chromatograms obtained from stability studies of cetirizine DIhydrochloride. A: cetirizine DIhydrochloride standard solution (25 μg/mL); B: cetirizine dihydrochloride solution in 2 M HCl at 70ºC after 72 h; C: cetirizine dihydrochloride solution in 0.5% H2O2 at 70ºC after 3 h.
Table 1

System suitability parameters for chromatographic conditions

ParametersFoundAcceptable limits
USP theoretical plates (n = 6)3500N > 1500
USP tailing factor (n = 6)1.08T < 1.5
Repeatability (tR) (n = 6)0.23RSD < 1%
Repeatability (peak area)(n = 6)0.50RSD < 1%

Linearity

The HPLC method was linear over the concentration range of 1-20 μg/mL of cetirizine dihydrochloride. According to the data shown in Table 2, acceptable linearity with r2 > 0.999 was observed. The limit of quantification (LOQ) with CV < 1.5% was found to be 1 μg/mL. The limit of detection (LOD) based on S/N ratio of 3 was found to be 0.2 μg/mL.

Table 2

Statistical data of calibration curves of cetirizine dihydrocholride (n = 6)

ParametersResults
  Linearity range1-20 µg/mL
 Regression equationy = 39.49x - 8.39
 Standard deviation of slope0.15
 Relative standard deviation of slope (%)0.39
 Standard deviation of intercept0.42
 Correlation coefficient (r2)0.9994

Precision and accuracy

The results of the precision and accuracy of the method at three different concentration levels are demonstrated in Table 3. The intermediate precision was also studied by comparison of the assay results for cetirizine tablets by two analysts using two different HPLC systems. The CV values did not exceed 2%.

Table 3

Precision and accuracy of the method for determination of cetirizine dihydrochloride (Three sets for 3 days).

Added ( µg/mL)Recovered         (mean ±SD) (µg/mL)CV(%)Error(%)
Within-day (n =              3)
 1.00                1.01 ± 0.01        1.13       1.00
 5.00                5.00 ± 0.03        0.62       0.00
 20.00               19.98 ± 0.16        0.80      - 0.10
Between-day (n =9)
 1.00                1.01 ± 0.01        1.45       1.00
 5.00                5.00 ± 0.04        0.88       0.00
 20.00               19.90 ± 0.16        0.81      - 0.45

Relative recovery

The mean recovery of cetirizine dihydrochloride calculated by standard addition method to tablet powder was about 99% and no interfering peaks from excipients were observed.

Solution stability

The stock standard solution of cetirizine dihydrochloride in water was stable over a period of one weak at 4°C with a recovery of about 99.5%.

Analysis of pharmaceutical product

The cetirizine tablets were determined by using the proposed method. The results showed good agreement with the labeled amount (10.04 ± 0.06 mg per tablet).

Degradation studies

In a preliminary study the degradation of cetirizine dihydrochloride was studied in 0.1 M HCl, 0.1 M NaOH and 1% H2O2 at 80°C according to the reported method of Jaber et al. (4). The degradation in 1% H2O2 was too fast and no degradation was observed in 0.1 M NaOH. Subsequent studies were repeated by using higher strength of NaOH and HCl and lower strength of H2O2.

Cetirizine dihydrochloride was more stable in basic condition even under higher strength of NaOH (1 M, 2 M and 5 M). In fact, a very small loss of peak area was observed after long exposure to 5 M NaOH at 90°C up to 48 h.

Kinetics of degradation in acidic condition

The degradation of cetirizine dihydrochloride was investigated in acidic condition using 2M hydrochloric acid. The drug was found to degrade in this condition at different temperatures with a same pattern (Figure 2b). A regular decrease in cetirizine dihydrochloride concentration was observed with increasing time. A linear relationship was observed by plotting the 1og of residual percent cetirizine dihydrochloride versus time (Figure 3) which indicates a pseudo-first-order kinetics according to the following equation:

log Ct = log C0 - kt / 2.303

Pseudo-first-order plots for the degradation of cetirizine dihydrochloride in 2 M HCl at various temperatures using HPLC method. Key; Ct, percent remained cetirizine dihydrochloride at time t, and C0, percent cetirizine dihydrochloride at zero time

Here, Ct is the percent remained cetirizine peak area, C0 is the initial percent of cetirizine (100%), k is the apparent first order rate constant with a negative sign and t is the time. Good correlation coefficient for all plots in different temperatures was observed (CV > 0.99). The values for the apparent first order rate constants and the half-lives are presented in Table 4. The rate of degradation was decreased with temperature reduction.

Table 4

Degradation equation, apparent rate constant (k) and half-life (t1/2) for cetirizine dihydrochloride in 2 M HCl

Temperature (°C)Equationr2-valuek (h-1)t1/2 (h)
70y = - 0.002x + 2.0030.9990.005138.6
80y = - 0.006x + 1.9980.9920.01449.5
85y = - 0.013x + 2.0000.9940.03023.1
90y = - 0.027x + 1.9170.9910.06211.2

Based on the Arrhenius relationship in the temperature range of 70-90°C (Figure 4), the following equation was used to calculate the activation energy for the acid-mediated degradation process of cetirizine which was about 121.8 KJ/mole.

K=Ae_Eact/RT

Arrhenius plot for degradation of cetirizine dihydrochloride in 2 M HCl

Here, k is the apparent first order rate constant. A is the pre-exponential factor, Eact is the activation energy and T is the temperature.

No significant degradation was observed at lower temperatures. So, the kinetics of degradation could not be monitored in these temperatures.

Kinetics of degradation in oxidative condition

The degradation of cetirizine dihydrochloride was too fast in 1% H2O2 medium. Using 0.5% H2O2 the degradation reaction could be followed at four different temperatures (50, 60, 70 and 80°C). Degradation pattern was similar in all conditions and a few small new peaks were detected in the chromatogram (Figure 2c).

The results for degradation of cetirizine dihydrochloride in oxidative condition at four different temperatures were also shown in Figure 5. A linear relationship was observed between the log residual percent of cetirizine dihydrochloride versus time which indicates a pseudo-first-order kinetics. The values for the apparent first order rate constants and half-lives are presented in Table 5. In addition, by plotting the Arrhenius equation (Figure 6), the activation energy for the hydrogen peroxide mediated degradation was found to be 36.3 KJ/mole.

Pseudo-first-order plots for the degradation of cetirizine dihydrochloride in 0.5% H2O2 at various temperatures using HPLC method. Key; Ct, percent remained cetirizine dihydrochloride at time t, and C0, percent cetirizine dihydrochloride at zero time
Arrhenius plot for degradation of cetirizine dihydrochloride in 0.5% H2O2
Table 5

Degradation equation, apparent rate constant (k) and half-life (t1/2) for cetirizine dihydrochloride in 0.5% H2O2

Temperature (°C)Equationr2-valuek (h-1)t1/2 (min)
50y = - 0.127x + 1.9970.9940.293141.9
60y = - 0.193x + 1.9660.9940.44493.6
70y = - 0.301x + 2.0370.9960.69360.0
80y = - 0.409x + 1.9650.9990.94244.1

References

  • 1.

    Brunton LL, Lazo JS, Parker KL. Goodman and Gillman's. The Pharmacological Basis of Therapeutics. 11th edition. USA: Mc Graw-Hill Medical Publishing Division; 2005.

  • 2.

    El Walily AFM, Korany MA, El Gindy A, Bedair MF. Spectrophotometric and high performance liquid chromatographic determination of cetirizine dihydrocholride in pharmaceutical tablets. J. Pharm. Biomed. Anal. 1998;17:435-442. [PubMed ID: 9656155].

  • 3.

    Paw B, Misztal G, Hopkata H, Drozd J. Development and validation of a HPLC method for the determination of cetirizine in pharmaceutical dosage forms. Pharmazie. 2002;57:313-315. [PubMed ID: 12061254].

  • 4.

    Jaber AMY, Al Sherife HA, Al Omari MM, Badwan AA. Determination of Cetirizine dihydrochloride, related impurities and preservatives in oral solution and tablet dosage forms using HPLC. J. Pharm. Biomed. Anal. 2004;36:341-350. [PubMed ID: 15496327].

  • 5.

    Arayne MS, Sultana N, Siddiqui FA. Determination and quantification of cetirizine HCl in dosage formulations by RP-HPLC. Pak. J. Pharm. Sci. 2005;18:7-11.

  • 6.

    Haghighi S, Shapouri MR, Amoli-Diva M, Pourghazi K, Afruzi H. HPTLC-Densitometric Determination for Cetirizine and Montelukast Analysis in Combined Tablet Dosage Forms. Iranian J. Pharm. Res. 2013;12.

  • 7.

    Karakus S, Kucukguzel I, Kucukguzel SG. Development and validation of a rapid RP-HPLC method for the determination of cetirizine or fexofenadine with pseudoephedrine in binary pharmaceutical dosage forms. J. Pharm. Biomed. Anal. 2008;46:295-302. [PubMed ID: 18054459].

  • 8.

    Hadad GM, Emara S, Mahmoud WMM. Development and validation of a stability- indicating RP-HPLC method for the determination of paracetomol with dantrolene or/and cetirizine and pseudoephedrine in two pharmaceutical dosage forms. Talanta. 2009;79:1360-1367. [PubMed ID: 19635371].

  • 9.

    Likar MD, Mansour HL, Harwood JW. Development and validation of a dissolution test for a once-a-day combination tablet of immediate-release cetirizine dihydrochloride and extended-release pseudoephedrine hydrochloride J. Pharm. Biomed. Anal. 2005;39:543-551.

  • 10.

    Azhagvuel S, Sekar R. Method development and validation for the simulthaneous determination of cetirizine dihydrocholride, paracetamol, and phenylpropanolamine hydrochloride in tablets by capillary zone electrophoresis J. Pharm. Biomed. Anal. 2007;43:873-878.

  • 11.

    Gowekar MM, Pande VB, Kasture AV, Tekade AR, Chandorkar JG. Spectrophotometric estimation of ambroxol and cetirizine hydrochloride from tablet dosage form. Pak. J. Pharm. Sci. 2007;20:250-251. [PubMed ID: 17545112].

  • 12.

    “The British Pharmacopeia,”. London, UK: British Pharmacopoeial Commision; 2008.

  • 13.

    Makhija SN, Vavia PR. Stability indicating HPTLC method for the simultaneous determination of pseudoephendrine and cetirizine in pharmaceutical formulations. J. Pharm. Biomed. Anal. 2001;25:663-667. [PubMed ID: 11377047].

  • 14.

    Zaater MF, Tahboub YR, Najib NM. RP-LC method for the determination of cetirizine in serum. J. Pharm. Biomed. Anal. 2000;22:739-744. [PubMed ID: 10815716].