Reagents and chemicals
Doxorubicin hydrochloride and daunorubicin hydrochloride were purchased from Sigma Chemical Company (St. Louis, MO, USA). Analytical grade chloroform, HPLC grade acetonitrile, methanol, ethyl acetate and methyl tert-butyl ether (MTBE) were purchased from Merck (Darmstadt, Germany). Ultrapure water was obtained using a Millipore Direct-QTM (Millipore Corporation, Bedford, MA, USA).
Instrumentation
The HPLC system consisted of a Wellchrom K-1001 HPLC pump, a Wellchrom online degasser and a Rheodyne auto injector equipped with a 50 µL loop were all from Knauer coupled with a fluorescence detector (RF-10A XL Shimadzu). The chromatographic data was acquired by Chromgate 3.1 software from Knauer.
Chromatographic conditions
The chromatographic separation was performed on a PerfectSil C18 column (4.6 × 150 mm, 5 μm particle size, MZ-Analysentechnik, Mainz, Germany). The optimum mobile phase consisting of acetonitrile and water (32: 68, v/v), pH adjusted to 2.6 with 85% orthophosphoric acid was delivered isocratically at a flow rate of 1 mL/min. The column temperature was maintained at 35 °C and excitation and emission wavlengths were set at 475 and 555 nm, respectively. The injection volume was 50 µL.
Preparation of stock solutions and standards
Stock solutions (1 mg/mL) of doxorubicin and daunorubicin (IS) were separately prepared in methanol and stored at -20 °C.
To prepare working solutions of doxorubicin, the stock solution was diluted with water to give concentrations of 1, 5, 10, 50, 100, 200, 500 µg/mL. Solutions containing 800 ng/mL IS were also prepared by serial dilutions of stock solution with methanol.
Calibration standards of doxorubicin were freshly prepared by spiking appropriate amounts of working solutions in pooled drug free plasma at concentrations of 5, 10, 20, 50, 70, 100, 200, 500, 700 and 1000 ng/mL. Quality control (QC) samples in the rat plasma were prepared at concentrations of 10, 50, 100, 500 and 1000 ng/mL. The spiked samples were then treated following the sample preparation procedure as indicated below.
Sample preparation
To 60 µL of plasma sample, 50 µL of IS solution (800 ng/mL daunorubicin hydrochloride in methanol) was added and vortex-mixed for 30 s. The extraction of drug was performed by adding 900 µL of a mixture of chloroform/methanol (4:1, v/v). After vortex mixing for 10 min and centrifugation (10 min, 10000 g), the organic phase was collected, transferred to a clean tube and evaporated to dryness under a stream of nitrogen at 40 °C. Dry residues from plasma were dissolved in 60 µL of mobile phase, and after centrifugation for 5 min (10000 g), 50 µL of the supernatant was injected into the chromatographic column.
Method validation
The developed method was validated in terms of selectivity, linearity, accuracy and precision (intra and inter-day variability), limits of detection (LOD) and quantification (LOQ), recovery and stability under different storage conditions.
Selectivity
In order to verify the selectivity of the method, blank plasma samples of six different rats were analyzed and possible interferences with doxorubicin and IS were checked by visual comparing of chromatograms.
Linearity
To evaluate the linearity of HPLC method, plasma calibration curves of at least 9 points were constructed using freshly prepared spiked samples, by plotting the peak-area ratio of doxorubicin to internal standard versus the nominal concentration of doxorubicin. Linearity of the method was established by the least-squares linear regression analysis.
Precision and accuracy
The precision of the method is expressed by the relative standard deviation (RSD %) of replicate measurements as a measure of random error. Accuracy of analytical method is reported as a relative error (RE %) and defines the difference between measured and nominal concentrations.
The intra-day and inter-day precisions were determined by analyzing five replicates of QC samples on the same day and three times on three days, respectively.
Extraction recovery
The extraction recovery of doxorubicin was determined at three concentrations of 10, 100 and 500 ng/mL by comparing the responses obtained from processed plasma samples with those obtained from aqueous standard solutions (n=5). The extraction recovery of the IS was calculated at the concentration of 800 ng/mL as well.
Limit of detection (LOD) and Limit of quantitation (LOQ)
Limit of detection (LOD), defined as the lowest detectable concentration is considered a concentration that has a signal to noise ratio of 3:1 (
38). The limit of quantification (LOQ) was taken as the lowest concentration that can be accurately (relative error < 20%) and precisely (RSD < 20%) determined (
39).
Stability
The stability of QC samples was established under following conditions: after going through three freeze-and-thaw cycles, from -20 °C to room temperature (freeze-thaw stability), short term stability at room temperature for the period of routine sample work-up (at least 3 h), stability of dry extract and long term stability in plasma and stock solutions at storage conditions (-20 °C) for 1 month.
Pharmacokinetic study
Animal treatment
Male Sprague-Dawley rats weighing 200–220 g were purchased from the Razi Institute of Iran (Tehran, Iran). Each rat was housed in a cage with a 12 h light/12 h dark cycle at ambient temperature (21-22 °C) and the relative humidity of 55 ± 5%. The rats were fasted overnight before experimentation and had ad libidum access to water. All protocols and procedures were approved by the local ethics committee for animal experiments of Shahid Beheshti University in Tehran, Iran.
A solution of doxorubicin (1 mg/mL in normal saline) at a dose of 5 mg/Kg was administered intravenously (I.V.) via the rat tail vein and blood samples (150 µL) were taken from the tail vein before (blank sample) and after the drug administration at several time points (5 min, 0.25, 0.5, 1, 2, 4, 6, 8, 10, 12, 24 and 48 h) in tubes containing EDTA as an anticoagulant. Each blood sample was gently inverted several times to ensure complete mixing with the anticoagulant. After centrifugation at 5000 g for 10 min, plasma was separated and stored at -20 °C until analysis.
Pharmacokinetic analysis
Concentrations of doxorubicin in plasma samples were analyzed by the developed HPLC method. Plasma concentration versus time data were analyzed and pharmacokinetic parameters including elimination rate constant (K), elimination half-life (t
1/2), mean residence time (MRT), volume of distribution at steady- state (Vss), systemic clearance (Cl) and the area under the plasma concentration versus time curve (AUC
0-∞) in plasma were determined by a noncompartmental analysis (
40).
The elimination rate constant (K) was estimated by the least-square regression of plasma concentration-time data points in the terminal log-linear region of the curves. The Half-life was calculated as 0.693 which was divided by K. AUC0-∞ was calculated using the trapezoidal rule with extrapolation to infinity. The clearance was calculated by dividing the dose by AUC0-∞. The volume of distribution at steady- state (Vss) and MRT was calculated using the following non-compartmental equations:
MRT = AUMC / AUC0-∞
Vss = Cl / MRT
In previous equations, AUMC (area under the first moment curve) is the area under the curve of C × t versus t from time zero to infinity.