Simple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma

authors:

avatar Mohsen Amini 1 , avatar Mahnaz Khanavi 1 , avatar Abbas Shafiee 1 , *

Department of Medicinal Chemistry, Faculty of Pharmacy, Medical Sciences University of Tehran, Tehran, Iran

How To Cite Amini M, Khanavi M, Shafiee A. Simple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma. Iran J Pharm Res. 2004;3(2):e128189. https://doi.org/10.22037/ijpr.2010.583.

Abstract

A rapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82: 18 v/v). Extraction of drug was performed from plasma by liquid-liquid extraction and the average recovery was 78.2% The assay is precise, with inter-assay coefficient of variation of 6.70 % at 0.25-8 µg/ml (n=3). Using UV detection at 277 nm the detection limit for ciprofloxacin was 20 ng/ml of plasma and the mean extraction recovery was 78.2 %. Short elution time, using UV detector and usage of ofloxacin as internal standard are advantages of this method.