Effect of Voxel Size on Detection of External Root Resorption Defects Using Cone Beam Computed Tomography

authors:

avatar Sima Nikneshan 1 , avatar Solmaz Valizadeh 1 , avatar Anahita Javanmard 2 , avatar Leila Alibakhshi 1 , *

Department of Radiology, School of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran
Dentist, Tehran, Iran

how to cite: Nikneshan S, Valizadeh S, Javanmard A, Alibakhshi L. Effect of Voxel Size on Detection of External Root Resorption Defects Using Cone Beam Computed Tomography. I J Radiol. 2016;13(3):e34985. https://doi.org/10.5812/iranjradiol.34985.

Abstract

Background:

Selecting a voxel size that yields minimal radiation dose with no significant compromise of the diagnostic accuracy of cone beam computed tomography (CBCT) is particularly important.

Objectives:

This study aimed to assess the effect of voxel size on detection accuracy of simulated external root resorption defects using NewTom CBCT system.

Materials and Methods:

In this diagnostic study, the roots of 90 extracted human central incisors were hypothetically divided into cervical, middle, and apical thirds. Variable-size defects were prepared in the buccal and lingual surfaces of the roots and CBCT scans were obtained with four different voxel sizes (150, 200, 250, and 300). Presence or absence of defects on CBCT scans was determined by three radiologists and the results were compared with the gold standard (actual size and depth of defects). Sensitivity and specificity values were calculated and reported for different groups.

Results:

In the lingual surfaces, the highest specificity and sensitivity belonged to 300µm voxel size (92.9%), and 200 and 250µm voxel sizes (both 97.4%) in the cervical third, respectively. In the middle third, the highest specificity and sensitivity belonged to 250 and 200µm voxel sizes (84.6% and 100%, respectively). In the apical third, the highest specificity and sensitivity belonged to 300 µm voxel size (100% and 97.9%, respectively). In the buccal surfaces, the highest specificity and sensitivity belonged to 150 (75.0%), 300 and 250 µm voxel sizes (100%) and the minimum values belonged to 200 µm voxel size (60% and 97.3%) in the cervical third. In the apical third, the highest specificity and sensitivity were noted in 300 µm voxel size (100% and 97.6%, respectively) and the minimum values were seen in 200 and 150 µm voxel sizes (93.8%, 90.5%, respectively). In the middle third, 300, 250 and 200 µm voxel sizes yielded the highest specificity (88.9%), while 150 µm voxel size yielded the highest sensitivity (98.8%).

Conclusion:

Considering the similar diagnostic efficacy of all voxel sizes, 300 µm voxel size can be used with adequate efficacy for detection of external root resorption defects with minimal patient radiation dose and the shortest scanning time.

1. Background

External root resorption is an irreversible, multi-factorial process that may lead to tooth loss (1). External root resorption in non-traumatized primary teeth is part of a physiological process to replace the primary teeth with their permanent successors. However, root resorption may also occur as a pathological process. It may also occur following orthodontic tooth movement, dental trauma, pulp infections, bleaching or periodontal disease. Moreover, impacted teeth, cysts, tumors and pressure of erupting canines applied to lateral incisors may cause external root resorption (2, 3). If diagnosed early and the causative agent is eliminated, the resorbed area may be restored followed by cementum deposition. Thus, early detection of external root resorption is necessary for a prompt treatment. To date, no general protocol has been defined for CBCT for specific diagnostic tasks in dentistry (4).

External root resorption is diagnosed by clinical and radiographic examinations (5). However, use of conventional radiography for this purpose results in false negative results in 51.9% and false positive results in 15.3% of cases (6). The CBCT technology aids the diagnosis of endodontic pathosis and analysis of resorption defects (7). Defects smaller than 0.6mm in diameter and less than 0.3mm in depth are not detectable on conventional periapical radiographs; this results in under-diagnosis and progression of root resorption (8, 9). Radiographic detection of external root resorption on buccal and lingual root surfaces is a real challenge (10). In 2007, Da Silveira et al. conducted a study on the elimination of superimposition of structures to improve the detection accuracy of resorption defects. This study showed that axial multislice computed tomography (CT) had a high diagnostic sensitivity and excellent specificity for detection of simulated defects in the buccal surfaces of teeth (except for small defects located in the apical third) (11).

On the other hand, diagnostic efficacy and correct estimation of the position and size of root resorption defects are critical factors for selection of an appropriate treatment plan and achieving a successful outcome (9). Thus, three-dimensional (3D) images may serve as important diagnostic tools in dental treatments. Use of CBCT scans can greatly help in this regard due to advantages such as the use of collimator that limits the radiation from the X-ray tube to a specific area, use of convergent voxels that enable image reconstruction of dental structures with an original quality, high speed (10 to 70 seconds) and low patient radiation dose (approximately 1/60 of that of multislice CT) (12-14). Accuracy of measurements using CBCT was comparable with those made on dry mandible (15). In CBCT, data acquisition is done via the rotation of X-ray beam around an object that gradually moves forward. On the other hand, voxel size in this technique has a direct correlation with the quality of the image and the exposure dose. In a study performed by Liedke et al. in 2009, CBCT was reported to be valid for detection of simulated external root resorption defects and 0.3 mm voxel size yielded the highest diagnostic accuracy and the least patient radiation dose (16). In their study, no correlation was noted between voxel size (0.2, 0.3 and 0.4 mm) and tomographic plane (axial, frontal, sagittal), size of defect (small, medium, large) or position of defects (apical, middle or cervical third of the root). Moreover, sensitivity and specificity of CBCT were equal for all voxel sizes. In another study conducted by Patel et al. in 2009, CBCT was found to be a valid and reliable technique for estimation of presence or absence of root resorption compared to intraoral digital radiography (17). Thus, considering the advantages of CBCT, it may enable early detection of internal and external root resorption and improve the prognosis of treatment.

Researchers are searching for dental imaging techniques with high diagnostic accuracy and low patient radiation dose to enhance diagnosis and decrease possible side effects of radiation.

2. Objectives

Considering the importance of early detection of external root resorption, this study sought to assess the effect of voxel size on detecting external root resorption using NewTom CBCT system.

3. Materials and Methods

This diagnostic study was conducted on 90 extracted sound human mandibular central incisors. The teeth had been extracted due to periodontal disease (hopeless prognosis) or upon patient’s demand for a complete denture. Sample size was calculated to be 90 considering 80% power of study and type 1 error of 0.05 using the sample size calculation formula and also based on previous studies (16, 18).

Sampling was non-randomized but variable-size defects were created randomly in the cervical, middle, and apical thirds of the roots. The teeth were visually inspected to ensure they were sound. After collection, the teeth were immersed in a solution of sodium hypochlorite in water (1/5 ratio) and were then immersed in 70% alcohol. The teeth were randomly coded from 1 to 90. The roots were hypothetically (without sectioning) divided into cervical, middle, and apical thirds and a total of 450 regions were obtained as such. In each region, a defect was created or the region remained intact as a control site. For each region, four possibilities existed with regard to cavity preparation:

1. Small defect (0.3 mm in depth and 0.6 mm in diameter)

2. Medium defect (0.6 mm in depth and 1.2 mm in diameter)

3. Large defect (0.9 mm in depth and 1.8 mm in diameter)

4. No defect (control)

Therefore, 90 teeth were selected to simulate all possibilities of occurrence of external root resorption defects in the clinical setting. To simulate external root resorption defects, cavities were prepared by a round bur. Defects were created in the buccal and lingual root surfaces by 0.6, 1.2, and 1.8 mm round diamond burs. When preparing the defects, the bur shank was positioned tangent to the root to ensure maximum precision in preparing the defects. To ensure the accurate size of the defects, dimensions were measured by a digital caliper. Defects were randomly prepared in the cervical, middle, and apical thirds. To simulate the periodontal ligament (PDL), roots were coated with a 0.2 mm-thick layer of wax using the dipping method to the level of the cementoenamel junction (Figure 1) (16). Six wax molds were formed in the form of the mandibular arch, then filled with a mixture of plaster and sawdust, and 15 teeth were mounted in each mold (Figure 2). Next, CBCT scans were obtained of each mold with sagittal, frontal and axial slices using four different voxel sizes of 150, 200, 250, and 300μm (NewTom VGi, QR SRL Company, Verona, Italy). The reason for selecting these voxel sizes was because root resorption defects are very small, and the smaller the voxel size, the higher the odds of visualization of smaller defects. Also, selection of these voxel sizes was based on previous studies (16, 18). To obtain radiographs, molds were placed in a container filled with water to simulate soft tissue. Qualitative and quantitative indices for presence or absence of defects and size of defects were evaluated by three radiologists who were blinded to the group allocation of specimens using the NewTom software (Figure 3). To assess intra-observer agreement, the same observers evaluated the samples again after 15 days. The observers were allowed to use all features of NewTom software. Data were recorded in data sheets and compared with the gold standard. The gold standard values were recorded earlier when the defects were being preparing.

The teeth after wax coating
The teeth after wax coating
The teeth were mounted in molds filled with plaster and sawdust and immersed in water to simulate soft tissue.
The teeth were mounted in molds filled with plaster and sawdust and immersed in water to simulate soft tissue.
Sagittal, frontal, and axial sections of CBCT scans taken with four different voxel sizes of 150, 200, 250 and 300 μm
Sagittal, frontal, and axial sections of CBCT scans taken with four different voxel sizes of 150, 200, 250 and 300 μm

The agreement of diagnoses of simulated external root resorption defects with the gold standard was evaluated based on voxel size, defect size, presence of defect in the buccal or lingual root surface, and cervical, middle or apical third of the root and reported using descriptive statistics (number and percentage). The actual size and depth of defects created on the root surfaces (namely no cavity, small cavity, medium cavity, and large cavity) were considered as the gold standard. Also, sensitivity and specificity of diagnoses based on each of the above-mentioned parameters were calculated and reported. The data were analyzed using SPSS version 22 software (Microsoft, IL, USA). The Kappa coefficient for the three observers was calculated as 1. The proportion test was used to compare sensitivity and specificity values among different voxel sizes.

4. Results

Assessment of the agreement of diagnoses with the gold standard (Tables 1 and 2) revealed that in the cervical third of the lingual surfaces of the roots, the highest specificity belonged to 300 μm voxel size (92.9%), and the lowest to 150μm voxel size (64.3%) (P = 0.198). In this group, the highest sensitivity belonged to 200 and 350 μm voxel sizes (97.4%) and the lowest to 300 μm voxel size (94.7%) (P = 0.920) (Table 3).

Table 1.

Agreement of Diagnoses with the Gold Standard Based on the Size of Defects in the Buccal Root Surfaces Using Different Voxel Sizes (16, 18)

Anatomical PartDiagnosed Defect SizeVoxel Sizes
300 μm250 μm200 μm150 μm
Gold Standard Defect Size
No DefectSmall DefectMedium DefectLarge DefectTotalNo DefectSmall DefectMedium DefectLarge DefectTotalNo DefectSmall DefectMedium DefectLarge DefectTotalNo DefectSmall DefectMedium DefectLarge DefectTotal
Cervical ThirdNo Defect11000111100011910111901111
Small Defect5184027321302731941271260027
Medium Defect1515021061502117130211812021
Large Defect0003131201283120029311212731
Total1723193190162719289015271731901236142890
Middle ThirdNo Defect81009810098100972009
Small Defect2211024219212422110241211124
Medium Defect0514019051311902152190711119
Large Defect0013738014333803332380223438
Total102716379010261935901027193490832143690
Apical ThirdNo Defect4800048470014845102484510248
Small Defect1160017214101711510172150017
Medium Defect041401808100180711018099018
Large Defect00077200572005720057
Total49201479051221169048231279049259790
Table 2.

Agreement of Diagnoses with the Gold Standard Based on the Size of Defects in the Lingual Root Surfaces Using Different Voxel Sizes

Anatomical PartDiagnosedDefect SizeVoxel Size
300 μm250 μm200 μm150 μm
Gold Standard Defect Size
No DefectSmall DefectMedium DefectLarge DefectTotalNo DefectSmall DefectMedium DefectLarge DefectTotalNo DefectSmall DefectMedium DefectLarge DefectTotalNo DefectSmall DefectMedium DefectLarge DefectTotal
Cervical ThirdNo defect131001410202141200214921214
Small defect1141016015101601510161131116
Medium defect27210302919030252303021117030
Large defect1012830001293001128300102930
Total1722232890122621319014212530901227193290
Middle ThirdNo defect1000313112001311110131030013
Small defect0281130027213002811301262130
Medium defect1321025051912509151250915125
Large defect0012122010212200121220012122
Total1131232590113521239011381823901138192290
Apical ThirdNo defect4200042391214339111423911142
Small defect1162019118001921610191171019
Medium defect0414018051301805130180711018
Large defect0011011110810102811112711
Total4320171090412515990422217990412615890
Table 3.

Sensitivity, Specificity, Standard Error and 95% Confidence Interval Values for Different Voxel Sizes for Detection of Defects in the Lingual Surfaces of the Roots

Voxel Sizes, µmSpecificity, %Sensitivity, %SE95% CI
Cervical Third350-97.40.016770.974 ± 0.03355
30092.9-0.02710.920 ± 0.0541
300-94.70.023620.947 ± 0.04723
250----
200-97.40.016770.94 ± 0 04723
15064.3-0.05050.643 ± 0.1010
Middle Third30076.9-0.04440.769 ± 0.0889
300-98.70.011940.987 ± 0.02388
25084.6-0.03800.846 ± 0.0761
250-100--
20084.6-0.03800.846 ± 0.0761
200-100--
15076.9-0.04440.769 ± 0.0889
150-98.70.011940.987 ± 0.02388
Apical Third300100---
300-97.90.015110.979 ± 0.03023
25092.2-0.02710.920 ± 0.0541
20092.2-0.02710.920 ± 0.0541
200-93.80.02540.938 ± 0.05084
20092.2-0.02710.920 ± 0.0541

In the middle third of the lingual surfaces of the roots, the highest specificity belonged to 200 and 250 μm voxel sizes (84.6%) and the lowest to 150 and 300 μm voxel sizes (76.9%) (P = 0.518). In this group, the highest sensitivity belonged to 200 and 250 μm voxel sizes (100%) and the lowest to 150 and 300 μm voxel sizes (98.7%) (P = 0.953) (Table 3).

In the apical third of the lingual surfaces of the roots, the highest specificity belonged to 300 μm voxel size (100%) and other voxel sizes showed equal specificity (92.9%) (P = 0.883). In this group, the highest sensitivity belonged to 300 μm voxel size (97.9%) and the lowest to 200 μm voxel size (93.8%) (P = 0897) (Table 3).

In the cervical third of the buccal surfaces of the roots, the highest specificity belonged to 150 μm voxel size (75.0%) and the lowest to 200 μm voxel size (60%) (P = 0.411). In this group, the highest sensitivity belonged to 250 and 300 μm voxel sizes (100%) and the lowest to 200 μm voxel size (97.3%) (P = 0.916) (Table 4).

Table 4.

Sensitivity, Specificity, Standard Error and 95% Confidence Interval Values for Different Voxel Sizes for Detection of Defects in the Buccal Surfaces of the Roots

Voxel Sizes, µmSpecificity, %Sensitivity, %SE95% CI
Cervical Third300-100--
250-100--
20060-0.05160.60 ± 0.1033
200-97.30.017090.973 ± 0.03417
15075-0.04560.75 ± 0.0913
Middle Third300100---
300-97.60.016130.976 ± 0.03227
250-90.50.030910.905 ± 0.06182
20093.8-0.02540.938 ± 0.0508
150-90.50.030910.905 ± 0.06182
Apical Third30088.9-0.03310.889 ± 0.0662
300-97.50.016460.975 ± 0.03291
25088.9-0.03310.889 ± 0.0662
250-97.50.016460.975 ± 0.03291
20088.9-0.03310.889 ± 0.0662
200-97.50.016460.975 ± 0.03291
15077.8-0.04380.778 ± 0.0876
15098.8-0.01140.988 ± 0.0229

In the apical third of the buccal surfaces of the roots, the highest specificity belonged to 300 μm voxel size (100%) and the lowest to 150 and 200 μm voxel sizes (93.8%) (P = 0782). In this group, the highest sensitivity belonged to 300 μm voxel size (97.6%) and the lowest to 150 and 250μm voxel sizes (90.5%) (P=0.775) (Table 4).

In the middle third of the buccal surfaces of the roots, the specificity of 200, 250 and 300 μm voxel sizes was equal (88.9%) and the specificity of 150 μm voxel size was 77.8% (P = 0.591). In this group, the highest sensitivity belonged to 150 μm voxel size (98.8%) and the sensitivity of other voxel sizes was equal (97.5%) (P = 0.976) (Table 4).

No significant differences were noted in terms of sensitivity and specificity of different voxel sizes for detection of variable-size root resorption defects.

Agreement of diagnoses with the gold standard in 150, 200, 250, and 300 μm voxel sizes for variable-size defects in the cervical, middle and apical thirds of buccal and lingual surfaces are shown in Tables 1 and 2.

5. Discussion

Halting the progression of defects by early treatment plays a fundamental role in the prognosis of teeth with invasive cervical (19) or inflammatory (20) root resorption defects. Therefore, early and accurate detection of root resorption can significantly increase the treatment success. On the other hand, conventional radiographs do not provide sufficient diagnostic value and therefore, digital diagnostic systems were introduced due to their inherent advantages (21). Assessment of root resorption using CBCT is limited to patients in whom defects have been previously detected by radiographic examination and they have taken 3D scans for therapeutic purposes (12). Considering the importance of early detection of root resorption defects in increasing the treatment success rate, this study aimed to assess the effect of voxel size on the diagnostic accuracy of root resorption by CBCT.

Previous studies have shown that conventional radiography does not provide sufficient accuracy for detection of small external root resorption defects in buccal or lingual surfaces (6, 8-10). Thus, researchers have been in search for more efficient techniques for detection of external root resorption defects. In 2007, Da Silveira et al. evaluated the advantages of multislice CT to detect external root resorption defects and reported that this modality had a high sensitivity and specificity in detecting external root resorption defects on the buccal root surfaces (11). However, defects in the apical third of the roots were significantly more difficult to detect. On the other hand, in 2009, Liedke et al. showed that CBCT had a very high sensitivity and specificity for detection of external root resorption defects and they reported no significant difference in this regard based on the size of the defect, its position or the plane of section (16).

The current study showed that CBCT with different voxel sizes had relatively equal diagnostic efficacy in terms of sensitivity and specificity for detection of variable-size defects in different areas and surfaces of the roots and this technique had sufficient diagnostic value for detection of external root resorption. Although small differences in sensitivity and specificity were found in 150, 200, 250, and 300 μm voxel sizes, they were not significant.

The results of the current study are in line with those of Liedke et al. in 2009. They also reported equal specificity and sensitivity values for different CBCT voxel sizes. However, they only evaluated defects on buccal root surfaces and used iCAT CBCT system (16). In 2010, Kamburoglu et al. evaluated two CBCT systems with different voxel resolutions for detection of external and internal root resorption defects and showed that high resolutions of both systems had similar efficacy for detection of internal root resorption defects and their efficacy was superior to that of low resolutions in one of the systems (18). In addition, the results showed that the diagnostic accuracy of different voxel sizes of CBCT increased as the size of defects increased, but no significant difference was found between small and medium-size defects. In 1998, Goldberg et al. evaluated the detection accuracy of simulated external root resorption defects in the maxillary incisors and showed that detection of small defects was more difficult than that of medium and large size defects. This result was in accord with our findings (9). In 2007, Da Silveira et al. evaluated the diagnostic accuracy of CBCT for detection of external root resorption defects and showed accurate detection of apical small resorption defects in 28.6%, medium-size defects in 86.66%, and large defects in 100% of cases. The diagnostic accuracy of CT was higher for larger defects (11). Moreover, in 2009, Hahn et al. used flat panel volumetric CT (fpVCT) and showed that non-cavitary defects were accurately diagnosed in 53% of cases. Small resorption defects were accurately diagnosed in 69% of cases. These values were 96% for medium size and 89% for severe, large defects (22). In 2012, Neves et al. evaluated external root resorption using different voxel sizes of CBCT and showed that by an increase in size of defects, the accuracy, sensitivity, positive predictive value and negative predictive value of diagnoses increased as well (23, 24).

In order to three dimensionally reconstruct images using data retrieved from the axial scans, each initial voxel must be dimensionally converted to several cubic voxels. This process, called interpolation, creates same-size cubic voxels that occupy the same volume. The CT number of these cubes is equal to the mean initial voxel CT number. Creation of these new cubic voxels allows image reconstruction at each surface with no reduction in resolution.

Use of CBCT, depending on the device model and the protocol applied, significantly decreases the patient radiation dose (25). Similarly, radiation dose has a direct correlation with the number of slices. Thus, it has been suggested that minimum number of slices should be used for diagnostic purposes (14). Based on the results of the current study, all three voxel sizes had almost equal diagnostic efficacy and accuracy for detection of external root resorption defects. Thus, 300 μm voxel size, with the least patient radiation dose and the shortest scanning time compared to other voxel sizes may provide sufficient diagnostic accuracy for external root resorption defects in the buccal and lingual root surfaces.

Advent of CBCT revolutionized dental imaging because with only a slight increase in the patient radiation dose, high level of diagnostic information with adequate quality and quantity is obtained compared to conventional radiography and in most cases, this slight increase in dose is justifiable by taking into account the value of diagnostic information obtained. However, if this imaging modality cannot significantly increase the detection accuracy, this increase in patient radiation dose is not accepted. The patient radiation dose in CBCT scans is 3-7 times higher than that of conventional radiography. Moreover, CBCT is time consuming. But, the patient radiation dose in dental CBCT is much lower than that of medical CT.

On the other hand, the diagnostic advantages of each imaging technique must be considered by taking into account the risk of exposure to excess radiation. This is particularly important in younger patients since excess radiation adversely affects the development of organs (25). According to Farman in 2005, the “as low as reasonably achievable” (ALARA) principle is a fundamental rule for diagnostic radiographies that also applies to CBCT and some new principles must also be added for CBCT (26). At the same time, clinicians should try to minimize the patient radiation dose without negatively affecting the quality of images. Future studies are required to compare the diagnostic accuracy of different CBCT systems for detection of external root resorption as well as other defects.

The main limitation of this study was its in vitro design; thus, generalization of results to the clinical setting must be done with caution.

Acknowledgements

References

  • 1.

    Bakland LK. Root resorption. Dent Clin North Am. 1992;36(2):491-507. [PubMed ID: 1572510].

  • 2.

    Ericson S, Bjerklin K, Falahat B. Does the canine dental follicle cause resorption of permanent incisor roots? A computed tomographic study of erupting maxillary canines. Angle Orthod. 2002;72(2):95-104. [PubMed ID: 11999943]. https://doi.org/10.1043/0003-3219(2002)072<0095:DTCDFC>2.0.CO;2.

  • 3.

    Fuss Z, Tsesis I, Lin S. Root resorption--diagnosis, classification and treatment choices based on stimulation factors. Dent Traumatol. 2003;19(4):175-82. [PubMed ID: 12848710].

  • 4.

    Spin-Neto R, Gotfredsen E, Wenzel A. Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review. J Digit Imaging. 2013;26(4):813-20. [PubMed ID: 23254628]. https://doi.org/10.1007/s10278-012-9562-7.

  • 5.

    Kim E, Kim KD, Roh BD, Cho YS, Lee SJ. Computed tomography as a diagnostic aid for extracanal invasive resorption. J Endod. 2003;29(7):463-5. [PubMed ID: 12877264]. https://doi.org/10.1097/00004770-200307000-00009.

  • 6.

    Nance RS, Tyndall D, Levin LG, Trope M. Diagnosis of external root resorption using TACT (tuned-aperture computed tomography). Endod Dent Traumatol. 2000;16(1):24-8. [PubMed ID: 11202852].

  • 7.

    Abella F, Morales K, Garrido I, Pascual J, Duran-Sindreu F, Roig M. Endodontic applications of cone beam computed tomography: Case series and literature review. Giornale Italiano Endodonzia. 2015;29(2):38-50. https://doi.org/10.1016/j.gien.2015.08.002.

  • 8.

    Andreasen FM, Sewerin I, Mandel U, Andreasen JO. Radiographic assessment of simulated root resorption cavities. Endod Dent Traumatol. 1987;3(1):21-7. [PubMed ID: 3471513].

  • 9.

    Goldberg F, De Silvio A, Dreyer C. Radiographic assessment of simulated external root resorption cavities in maxillary incisors. Endod Dent Traumatol. 1998;14(3):133-6. [PubMed ID: 9863423].

  • 10.

    Borg E, Kallqvist A, Grondahl K, Grondahl HG. Film and digital radiography for detection of simulated root resorption cavities. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998;86(1):110-4. [PubMed ID: 9690255].

  • 11.

    da Silveira HL, Silveira HE, Liedke GS, Lermen CA, Dos Santos RB, de Figueiredo JA. Diagnostic ability of computed tomography to evaluate external root resorption in vitro. Dentomaxillofac Radiol. 2007;36(7):393-6. [PubMed ID: 17881597]. https://doi.org/10.1259/dmfr/13347073.

  • 12.

    Cohenca N, Simon JH, Mathur A, Malfaz JM. Clinical indications for digital imaging in dento-alveolar trauma. Part 2: root resorption. Dent Traumatol. 2007;23(2):105-13. [PubMed ID: 17367458]. https://doi.org/10.1111/j.1600-9657.2006.00546.x.

  • 13.

    Swennen GR, Schutyser F, Barth EL, De Groeve P, De Mey A. A new method of 3-D cephalometry Part I: the anatomic Cartesian 3-D reference system. J Craniofac Surg. 2006;17(2):314-25. [PubMed ID: 16633181].

  • 14.

    Scarfe WC, Farman AG, Sukovic P. Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc. 2006;72(1):75-80. [PubMed ID: 16480609].

  • 15.

    Torres MG, Campos PS, Segundo NP, Navarro M, Crusoe-Rebello I. Accuracy of linear measurements in cone beam computed tomography with different voxel sizes. Implant Dent. 2012;21(2):150-5. [PubMed ID: 22382754]. https://doi.org/10.1097/ID.0b013e31824bf93c.

  • 16.

    Liedke GS, da Silveira HE, da Silveira HL, Dutra V, de Figueiredo JA. Influence of voxel size in the diagnostic ability of cone beam tomography to evaluate simulated external root resorption. J Endod. 2009;35(2):233-5. [PubMed ID: 19166780]. https://doi.org/10.1016/j.joen.2008.11.005.

  • 17.

    Patel S, Dawood A, Wilson R, Horner K, Mannocci F. The detection and management of root resorption lesions using intraoral radiography and cone beam computed tomography - an in vivo investigation. Int Endod J. 2009;42(9):831-8. [PubMed ID: 19627378]. https://doi.org/10.1111/j.1365-2591.2009.01592.x.

  • 18.

    Kamburoglu K, Kursun S. A comparison of the diagnostic accuracy of CBCT images of different voxel resolutions used to detect simulated small internal resorption cavities. Int Endod J. 2010;43(9):798-807. [PubMed ID: 20609023]. https://doi.org/10.1111/j.1365-2591.2010.01749.x.

  • 19.

    Hiremath H, Yakub SS, Metgud S, Bhagwat SV, Kulkarni S. Invasive cervical resorption: a case report. J Endod. 2007;33(8):999-1003. [PubMed ID: 17878092]. https://doi.org/10.1016/j.joen.2007.02.014.

  • 20.

    Hommez GM, Browaeys HA, De Moor RJ. Surgical root restoration after external inflammatory root resorption: A case report. J Endod. 2006;32(8):798-801. [PubMed ID: 16861085]. https://doi.org/10.1016/j.joen.2005.10.038.

  • 21.

    Nair MK, Nair UP. Digital and advanced imaging in endodontics: a review. J Endod. 2007;33(1):1-6. [PubMed ID: 17185117]. https://doi.org/10.1016/j.joen.2006.08.013.

  • 22.

    Hahn W, Fricke-Zech S, Fricke J, Gruber RM, Dullin C, Zapf A, et al. Detection and size differentiation of simulated tooth root defects using flat-panel volume computerized tomography (fpVCT). Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;107(2):272-8. [PubMed ID: 18602316]. https://doi.org/10.1016/j.tripleo.2008.04.030.

  • 23.

    Neves FS, Vasconcelos TV, Vaz SL, Freitas DQ, Haiter-Neto F. Evaluation of reconstructed images with different voxel sizes of acquisition in the diagnosis of simulated external root resorption using cone beam computed tomography. Int Endod J. 2012;45(3):234-9. [PubMed ID: 22003884]. https://doi.org/10.1111/j.1365-2591.2011.01966.x.

  • 24.

    Neves FS, de Freitas DQ, Campos PS, de Almeida SM, Haiter-Neto F. In vitro comparison of cone beam computed tomography with different voxel sizes for detection of simulated external root resorption. J Oral Sci. 2012;54(3):219-25. [PubMed ID: 23047032].

  • 25.

    Ludlow JB, Davies-Ludlow LE, Brooks SL. Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol. 2003;32(4):229-34. [PubMed ID: 13679353]. https://doi.org/10.1259/dmfr/26310390.

  • 26.

    Farman AG. ALARA still applies. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;100(4):395-7. [PubMed ID: 16182157]. https://doi.org/10.1016/j.tripleo.2005.05.055.