The USP grade halothane (2-bromo-2-chloro-1,1,1-trifluoroethane) was purchased from Nicholas Piramal (Mumbai, India). The USP grade Isoflurane (2-chloro-2-(difluoromethoxy)-1,1,1-trifluoro-ethane) was purchased from Nicholas Piramal (Morpeth, United Kingdom) and USP grade sevoflurane (1,1,1,3,3,3-hexafluoro-2-(fluoromethoxy)propane) was from Abbot Laboratories. The USP grade 2-Chlorophenol, Cholorobenzene, Chloroform, Carbon Tetrachloride, 1-Butanol (99.5%) and carbon disulphide (99%) were purchased from Merck (Darmstadt, Germany). All new fibers were conditioned in GC injector according to manufacturer’s recommendations. Seventy-five micrometers of carboxen/polydimethylsiloxane (CAR/PDMS) SPME fibers and manual holders were supplied from Supelco (Supelco, Bellefonte, PA).
In this cross-sectional study, the authors designed a dynamic atmosphere standard chamber (
15) and installed it in the laboratory (
Figure 1). The temperature in the system was controlled by a heating coil that was actuated by a thermocouple with a temperature sensor (Samwon Eng, Model SU-105, South Korea). The system has three areas of different dimensions that allow velocity to be changed in each area. The velocity in a chamber was measured with a calibrated thermo -anemometer. Air of different humidity was generated using an impinge system at a range of bubbling flows and temperatures. Humidity was continuously monitored by a Testo 601 hygrometer (Model Testo 601, Testoterm GmbH & Co, Germany).
A, air inlet; B, bypass valve; C, valve to humidity system; D, humidity generation system; E, electrical coil; F, syringe pump; G, sampling chamber; H, temperature sensor; I, humidity sensor; J, sampling location; K, hygrometer; L, thermocouple; N, dry gas meter; O, outlet (to hood).
This system was suitable for sampling at various air velocities, temperatures, and humidity (
Figure 1). Analytes were injected into the premixing chamber with a calibrated syringe pump (SEP-10S Plus, Aitecs, Lithuania). The diluents’ gas flow rate in the system was supplied from an oil free compressor that was equipped with traps for hydrocarbons and volatile compounds and was checked continuously by a calibrated dry gas meter (Elster-Handel, Germany). In the upper surface of chamber, there are some septums for location of sampling by (
Figure 1J) SPME and duration of sampling was 3.5 - 4 hours. The concentration was calculated by the following equation (
Equation 1).

Equation 1.
That ρ is the density of analyte (g/mL), IR is injection rate (mL/h), MW is the molecular weight of analyte, C is the desirable concentration in the chamber in ppmv, Vm is the molar volume (24.45 L in the case of air at 25°C and 1 atm.) and Q is diluents’ gas flow rate (liters per minute).
A Varian 3800 GC-Saturn 2200 mass spectrometer was used to analysis SPME samples. The capillary VOCOL column 60 m × 0.25 mm ID and 1.5 μm film thickness (Supelco, Bellefonte, PA) used in gas chromatography–mass spectrometry (GC-MS) and carrier gas was helium in flow rate of 1 mL/min. The MS transfer line temperature was set at 220°C. The temperature program for analytes was set at 60°C for 1 minute and then increased at 4°C/min to 170°C and held for 1 minute. Calibration standards for analysis of samples were prepared in the range 0.5 - 1000 mg/mL in 1-butanol because of its low vapor volume.
For obtaining a cablibration curve based on the peak area, 0.1 mL of each liquid standard was injected to GC in triplicate
A total size of 54 samples obtained to assess the effects of temperature, humidity and velocity on sample rate and also the effect of storage time on samples recovery.
Concentration in the SPME samples could be calculated by the following equation according to Fick’s first law (
13) (
Equation 2).

Equation 2.
In this equation, n is mass of analyte absorbed on the SPME fiber (in nanograms) and determined by external liquid calibration. Dg is binary gas phase diffusion coefficient (in square centimeters per second), A is the diffusion cross-sectional area (in square centimeters), t is the sampling time in minutes, and C is the concentration of analyte in the sampling area (in nanograms per cubic centimeter).
The term DgA/L in
Equation 1 is called the sampling rate (SR). Dg could be theoretically calculated according to the method developed by Fuller et al. from following equation (
16).

Equation 3.
In this equation, T is temperature (in kelvin) and MA and MB are molecular mass of air and analyte respectively, P the air pressure (in atmosphere) VA and VB are molar volume of air and analyte respectively (in cubic metres per mole).
2.1. Statistical Methods
Statistical analyses were carried out by statistical package for social science (SPSS, version 17, Inc., Chicago). Charts and monograms were also created by Microsoft Excel®2007. The concentration difference among data was performed by analysis of variance (ANOVA) test and Tukey’s test. A P < 0.05 was considered statistically significant.