Spectral analysis
Luminescence spectra
Luminescence emission and excitation spectra of Tb
3+–DFP are shown in
Figure 2.
Luminescence excitation (A) and emission (B) spectra of: Tb3+-DFP complex. Experimental conditions: [Tb3+] = 105- M; [DFP] = 7.2 × 10 -6 M; pH = 7.5; λex /λem=295 nm/545 nm
Solutions containing only Tb3+ or DFP did not show any measurable luminescence with excitation at 295 nm. Under the same conditions, the characteristic luminescence spectrum of Tb3+-DFP was observed, with two emission peaks at 545 nm and 490 nm. These peaks are the characteristic luminescence peaks of Tb3+ and correspond to 5D4→7F6 and 5D4→7F5 transitions, respectively, of which the emission at 545 nm is much stronger. The luminescence intensity was proportional to the concentration of DFP.
Absorption spectra
Absorption spectra of Tb
3+ (spectrum 5, [Tb
3+]= 10
-4 M), DFP (spectra 1 and 2, [DFP]=2.85 × 10
-5 M and [DFP]=2.85 × 10
-5 M) and Tb
3+-DFP (spectra 3 and 4, [Tb
3+]= 10
-4 M + [DFP]=2.85 × 10
-5 M and [Tb
3+]= 10
-4 M + [DFP]=1.44 × 10
-5 M) are shown in
Figure 3. It can be seen that after addition of Tb
3+ into the DFP solution, a small red shift occurred in the maximal absorption peak, which is due to the formation of Tb
3+-DFP complex.
Absorption spectra of DFP in different systems (background correction was done by using a reference solution): DFP (1, 2);.)Tb3+-DFP (3, 4) and Tb3+(5). Conditions: [Tb3+] = 1.0 × 10 -4 M, [DFP] =2.87 × 10 -5 M (1, 3), 1.44 × 10 -5 M (2, 4).
Factors affecting the luminescence intensity of the system
Effect of pH
A series of Tris buffer solutions with different pH values but the same concentrations of other reagents together with corresponding blank solutions were prepared and their luminescence signals were measured at λ
ex/λ
em= 295 nm/545 nm. As shown in
Figure 4, the luminescence intensity of Tb
3+–DFP complex strongly depended on pH and reached a maximum value at 7.5. Thus, pH 7.5 was selected for the following experiments.
Effect of pH, conditions: [Tb3+] =10-5 M, [DFP] = 7.2 × 10 -6 M, λex /λem=295 nm/545 nm
Effect of buffer concentration
The effect of different concentrations of buffer on the luminescence intensity of the system is shown in
Figure 5. At lower concentrations of Tris, the OH groups of water molecules surrounded the terbium ions and actrd as effective luminescence quenchers due to OH oscillation, thus leading to a decrease in the luminescence intensity. As the concentration of buffer is increased, Tris ligands might prevent Tb
3+ ion from coordinating water around and so the luminescence intensity is increased. The results indicated that 0.5 mL of 0.1 M Tris–HCl buffer solution in 10 mL mixture was the optimum buffer volume.
Effect of buffer concentration, conditions: [Tb3+] =10-5 M, [DFP] = 7.2 × 10 -6 M, pH=7.5.
Effect of terbium (III) concentration
The effect of Tb
3+ concentration on the luminescence intensity of Tb
3+–DFP system was studied when the pH and concentration of buffer got fixed at 7.5 and 0.005 M . The results are shown in
Figure 6.Effect of terbium (III) concentration, conditions: [DFP] = 7.2 × 10 -6 M, pH = 7.5
It can be seen that luminescence intensity was the highest when the concentration of Tb
3+ in the mixture was 3.0 × 10
-4 M. Therefore, the concentration of Tb
3+ in the mixture was chosen at 3.0 × 10
-4 M for further investigations. The stoichiometry of the complex was studied using Job’s method, i.e. equal concentration of Tb
3+ and DFP were used. The ratio of molar fraction Tb
3+: DFP was varied and luminescence intensity of the complex was recorded. The maximum intensity was obtained at a mole fraction of 0.25 (
Figure 7), thus a stoichiometry of Tb
3+: DFP achieved as 1:3. The complex formation constant was calculated using a method described elsewhere (
20) and was 1.6 × 10
16.
Determination of the stoichiometry of the terbium (III), conditions: DFP complex using Job’s method; [DFP] and [Tb3+] = 3.0 × 10-4 M, pH=7.5
Effect of temperature
Temperature had great influence on the luminescence intensities of this system. The luminescence intensity sharply decreased with temperature from 0°C to 60°C. Therefore, 0°C was selected for further study (
Figure 8).
Effect of temperature, conditions: [Tb3+] =3 × 10-4 M, [DFP] = 7.2 × 10 -6 M, pH = 7.5
Effect of time
Under the optimum conditions, the effect of time on the luminescence intensity was studied at 0°C. The results showed that the luminescence intensity is stable at the first 25 min after addition of all reagents. In this study, 3 min was set as the optimum value for all luminescence intensity measurements (
Figure 9).
Effect of time, conditions: [Tb3+] = 3 × 10-4M, [DFP] = 7.2 × 10 -6 M, pH = 7.5
Effect of the order of addition
Finally, the effect of the order of addition was tested. For this purpose, series of solutions with different addition orders of reagents were measured at λex/λem = 295 nm/545 nm. Based on the results, we selected Tb3+, Tris_HCl, and DFP as the best order for this assay.
Analytical figures of merit
Under the optimal conditions, calibration graph for the determination of DFP was constructed. The linearity parameters for DFP are shown in
Table 1. Results of linearity evaluation indicated that, this method was linear in the range of 7.2 × 10
-9 to 1.4 × 10
-5 M for determination of DFP in aqueous solutions with a correlation coefficient of 0.999. The detection and quantification limits calculated from calibration graphs, are also given in
Table 1.
| Data pointa | Slope | Y-intercept | rb | Range | LODc | LOQc |
|---|
| 13 | 2.56 × 107 | 2.413 | 0.999 | 7.2 × 10-9 to 1.4 × 10-5 | 6.3 × 10-9 | 2.1 × 10-8 |
Details of the analytical performances of the previously reported methods and the proposed method for the determination of DFP are summarized in
Table 2. Compared with the previous methods, the proposed method is simple and has relatively lower detection limit and lower linearity range.
| Method | Linear range | Detection limit | Reference |
|---|
| HPLC | 3.5 × 10-6-1.4 × 10-5 | 3.5 × 10-7 | 4 |
| Voltammetry | 9.9 × 10-5-5.3 × 10-4 | 1.9 × 10-5 | 19 |
| Cyclic voltammetry | 3.0 × 10-5-1.0 × 10-3 | 1.4 × 10-5 | 3 |
| Cyclic voltammetry | 5.0 × 10-5-1.0 × 10-3 | 5.3 × 10-7 | 18 |
| Terbium sensitized luminescence | 7.1 10-9-1.4 × 10-5 | 6.3 × 10-9 | This work |
The results for precision, accuracy and recovery of the method respectively were given in
Tables 3 and
4; these results illustrated that the proposed method is accurate and precise for determination of DFP.
| Concentration (×10-6) |
|---|
| 0.072 | 7.20 | 13.0 |
|---|
| Intra-day | Inter-day | Intra-day | Inter-day | Intra-day | Inter-day |
|---|
| 2.9 | 4.8 | 2.1 | 1.3 | 1.2 | 1.3 |
| Accuracy | Recovery |
|---|
| Concentration (×10 -6 M) |
|---|
| 0.072 | 7.20 | 13.0 | 0.072 | 7.20 | 13.0 |
| 4.80 | 1.25 | 1.31 | 100.3 | 100.1 | 102.3 |
The stability data for the proposed method was shown in
Table 5 and the results indicated that, it is stable in short term room temperature and in stock solutions.
| Short term room temperature stability | Stock solution stability |
|---|
| Concentration (×10 -6 M) | 0.072 | 7.20 | 13.0 | 0.072 | 7.20 | 13.0 |
|---|
| Recovery % | 98.9 | 100.6 | 103.1 | 93.0 | 98.6 | 101.0 |
| RSD % | 3.5 | 2.3 | 1.2 | 4.4 | 1.7 | 1.2 |
Analytical application
Determination of DFP in tablets
The developed method was applied to the determination of DFP in tablets prepared according to the sample preparation procedure. For the assay of DFP, the samples must be diluted properly within the linear range of the determination of DFP and the sample solution was analyzed , using the standard calibration method. Label claim of each DFP tablet is 500 mg, and the measured amount of DFP in tablet was 503.3 ± 2.7 mg. The relative standard deviation for determination of DFP in tablets was 2.2% and the obtained recovery was 97.3-102.7%.