Fluorescence spectra
Fluorescence excitation and emission spectra of (1) Tb
3+-Phen, (2) Tb
3+-Phen-MTX, (3) Tb
3+-MTX and (4) Tb
3+ systems are shown in
Figure 2. The characteristic peak of Tb
3+-Phen was observed with two emission peaks at 545 nm and 490 nm. The fluorescence spectrum of Tb
3+-Phen-MTX system was similar to that of Tb
3+-Phen; however, the fluorescence intensity was decreased by MTX. These results indicated that there was an interaction between MTX and Tb3+-Phen. Therefore, the Tb
3+-Phen-MTX system was utilized in the assays of MTX. Since the quenched fluorescence intensity at 545 nm was higher, this wavelength was chosen to detect the fluorescence intensities throughout all the experiments.
Terbium-sensitized fluorescence excitation (λem= 545 nm) (A) and emission (λex= 300 nm) (B) spectra: (1) Tb3+-phen; (2) Tb3+-phen -MTX; (3) Tb-MTX; (4) Tb3. Conditions: [Tb3+] = 2×10-4 mol/L, [MTX] = 2 μg/mL, [phen] = 10-4 M, (Tris-HCl = 0.01 M, pH = 7.0).
Factors affecting the fluorescence intensity of the system
Effect of pH and concentration of buffer
The effect of pH on the quenched fluorescence intensity (ΔI
f(%)) of the Tb
3+-Phen-MTX system in the pH range of 4.5-10.0 (
Figure 3) was investigated. As it is obvious in
Figure 3, the decreased fluorescence intensity (ΔI
f(%)) of Tb
3+–Phen complex is strongly dependent on pH and is reached to a maximum value between pH 6.8 and 7.2. Thus, pH of 7.0 was selected as the optimum pH.
Effect of pH on the quenched fluorescence intensity Tb3+-phen -MTX system. Conditions: [phen], 2 ×10-4 mol/L; [Tb3+], 10-4 mol/L; [MTX], 2 μg/mL.ΔIf(%) = (I0–If)/I0 × 100 in which If and I0 were the intensities of the systems with and without MTX, respectively
This pH was adjusted by Tris buffer solution. Experiments indicated that the chemical nature of the buffer has also considerable effects on the quenched fluorescence intensity. The results showed that 1 mL of Tris-HCl (0.05 mol/L, pH = 7.0) in a final 5 mL was the most suitable concentration. Tris buffer also increases the obtained signal, probably due to its penetrating the coordination sphere of the chelate, giving rise to a synergistic effect.
Effect of time and temperature
Effect of time and different temperatures on the fluorescence intensity (ΔI
f(%)) of Tb
3+-Phen and Tb
3+-Phen-MTX systems was studied within 80 min (
Figure 4). As shown in
Figure 4, in all temperatures, the fluorescence intensity of Tb3+-Phen and Tb3+-Phen-MTX systems is approximately constant. Therefore, the quenched fluorescence intensity of the systems is very stable and determination can be carried out immediately after the addition of all regents. The other results also showed that the quenched fluorescence intensity of the systems remain stable for more than 12 h.
Effect of time on the fluorescence intensity and stability of Tb-phen-MTX system in different temperature. Conditions: [phen], 2 ×10-4 mol/L; [Tb3+], 10-4 mol/L; [MTX], 2 μg/mL; (Tris-HCl = 0.01 M, pH = 7.0).
Effect of concentration of terbium
The effect of the concentration of Tb
3+ on the quenched fluorescence intensity (ΔI
f(%)) of Tb
3+-Phen-MTX system in constant concentration of Phen was studied and results are shown in
Figure 5. It can be seen that ΔI
f(%) has been the highest when the concentration of Tb
3+ in the mixture was 2.0 × 10
-4 mol/L. Therefore, the concentration of Tb
3+ in the mixture was chosen at 2.0 × 10
-4 mol/L for further research.
Effect of Tb3+ concentration on the quenched Fluorescence intensity. Conditions: [phen], 2 ×10-4 mol/L; [MTX], 2 μg/mL, (Tris-HCl, 0.01 mol/L, pH = 7.0).
Effect of concentration of phen
The effect of Phen concentration on the fluorescence intensity (ΔI
f(%)) of Tb
3+-Phen-MTX system was studied and as the results show in
Figure 6, it was found that ΔI
f(%) of Tb
3+-Phen-MTX system reached to a maximum value when the concentration of Phen was 0.5 ×10
-4 mol/L. Therefore, this value was used for further study.
Effect of phen concentration on the quenched fluorescence intensity. Conditions: [Tb3+], 2 ×10-4 mol/L; [MTX], 2 μg/mL; (Tris-HCl, 0.01 mol/L, pH = 7.0)
Effect of addition order
Finally, the effect of addition order was tested. For this purpose, a series of solutions with different addition orders of reagents and their corresponding blank solutions were measured at λex/λem= 300 nm/545 nm. Based on the results and considering the system stability along with the quenched fluorescence intensity enhancement, the order of Tb3+, Tris-HCl, Phen and MTX was the best and followed in further investigations.
Concentration ranges and calibration graphs
By using the optimized conditions described above, a spectrofluorimetric method was developed for the determination of MTX in the biological samples. The measurements are based on the Stern-Volmer equations (
31,
32) which describes dynamic or collisional quenching requiring contact between the excited luminophore and the quencher Q, as follows:
I0/If = 1 + Kqτ0 [Q]
In this equation, I0 and If are the fluorescence intensities in the absence and the presence of a quencher, respectively; Kqis the quenching rate constant (Lmol-1s-1) and τ0 is the fluorescence lifetime(s) in the absence of the quencher. The equation shows that a plot of I0/If versus the quencher concentration should give a straight line usable for the determination of the quencher. A high value of Kq leads to a low limit of detection for the quencher. The data showed that there is linearity between the MTX concentration and the quenched fluorescence intensity of the Tb3+-Phen system.
The calibration graph (the plot of I0/If versus the MTX concentration) (n = 10) was found to be linear in the range of 0.02 to 10 μg/mL for MTX and its equation was as follows:
I0/If = 1.0969 C + 1.0176 (r = 0.999)
Where I0/If is the quenched fluorescence intensity of Tb3+-Phen by MTX and C is the concentration of MTX expressed in μg/mL. Standard deviations (SD) for the slope and intercept of the calibration graph were ± 0.0037 and ± 0.0025, respectively.
The limit of detection (LOD) and the limit of quantification (LQD)
In accordance to ICH definition (
33), the limit of detection was calculated as 3S
b/m (where S
b is standard deviation of the blank and m is slope of the calibration graph). Using this formula, LOD was found to be 0.015 μg/mL. The limit of quantification was defined as 10S
b/m and found as 0.052 μg/mL.
Precision and accuracy
In order to investigate the precision of the proposed method (repeatability), series of six solutions of 0.05, 1.0 and 4.0 μg/mL of MTX were measured on the same day. By applying the ICH definition (36), the relative standard deviation (RSD) for six analyses was 0.5, 1.9 and 0.4%, respectively. To assess the day-to-day precision (intermediate precision), repeated analyses of 1.0 and 4.0 μg/mL of MTX (six analyses) were performed over one month and interday RSD were 2.9 and 0.8%, respectively.
Specificity and interference study
For the possible analytical application of the proposed method, a systematic study of the effects of coexisting substances that was likely in biological samples on the reduced fluorescence intensity (ΔI
f%) was carried out under the optimal conditions. The effects of these species were evaluated by the addition of increasing concentration of various species to a fixed amount of MTX (2 μg/mL), under the same experimental conditions, until variation greater than 10% in analytical intensity was achieved. The tolerance levels of various interferents (
e.g. ions, amino acids, proteins and saccharides) are summarized in
Table 1. Careful examination of
Table 1 revealed that the most substances in low amounts were found to show small effects on the determination of MTX under permission of ± 10% relative error. In practice, through the dilution of urine samples up to 1000-fold and deproteinization of serum samples and with applying standard addition method, interference-free determination of MTX is possible. Hence, the selectivity achieved by the proposed method is good and it is possible to determine the MTX. Folic acid or folinic acid were co-administered with MTX and their interferences were also investigated and the tolerance ratios were 0.3 and 0.2, respectively.
| Coexisting substance | Ratio of Coexisting substance to MTX | ΔIF% |
|---|
| K+ (Cl-) | 1:75 | 0.8 |
| Na+(Cl-) | 1:150 | 2.4 |
| Ca2+(Cl-) | 1:75 | 4.4 |
| Ba2+(Cl-) | 1:75 | -1.1 |
| Al3+(Cl-) | 1:0.25 | -7.8 |
| Cr3+(Cl-) | 1:0.02 | -3.5 |
| Au3+(Cl-) | 1:0.20 | 4.5 |
| Zn2+(Cl-) | 1:0.4 | 2.6 |
| Mn2+(Cl-) | 1:0.4 | 3.9 |
| Ni2+(Cl-) | 1:0.2 | -1.3 |
| Cu2+(Cl-) | 1:0.15 | 4.1 |
| Cd2+(Cl-) | 1:0.25 | 2.3 |
| Co2+(Cl-) | 1:0.2 | 0.9 |
| L-Alanine | 1:75 | 3.6 |
| L-Cysteine | 1:25 | -5.0 |
| Tryptophane | 1:70 | 3.5 |
| Glycine | 1:25 | 2.7 |
| L-Leucine | 1:125 | -5.4 |
| Tyrosine | 1:25 | 0.8 |
| Uric acid | 1:0.025 | -1.9 |
| Sacarose | 1:12.5 | 2.6 |
| Glucose | 1:12.5 | -2.9 |
| Bovine serum albumin | 1:0.20 | -3.7 |
| Folic acid | 1:0.25 | -3.8 |
| Folinic acid | 0.2:1 | -4.5 |
Analytical applications
Determination of MTX in an injection solution
The developed method was applied to the determination of MTX in an injection solution and the results were shown in
Table 2. One injection (containing 50 mg/5 mL) was directly diluted to 100 mL and then analyzed using the standard calibration method. The concentration of MTX in the injection solution was found as 51.5 mg/5 mL.
| Added (μg/mL) | Found* (μg/mL) | Recovery (%) | RSD (%) |
|---|
| 25.0 | 26.0 ± 0.4 | 104.8 ± 1.7 | 1.5 |
| 40.0 | 40.5 ± 0.4 | 101.2 ± 1.2 | 1.0 |
| 100.0 | 99.5 ± 4.1 | 98.4 ± 1.9 | 4.2 |
| 200.0 | 209.5 ± 0.7 | 104.7 ± 0.4 | 0.3 |
| 300.0 | 314.5 ± 8.0 | 104.7 ± 2.7 | 2.5 |
Determination of MTX in urine samples
To demonstrate the usefulness of the proposed method, several aliquots of MTX were added to urine. The standard addition method was used for MTX determinations due to the possible matrix effect from urine. For the assays of MTX in samples of urine, the fresh samples must be diluted appropriately within the linear ranges of determination (1000 fold). A portion of these sample solutions was analyzed through the developed method, using the standard calibration method. In these conditions, there is no interference in determination of MTX and this method can be utilized for direct determination of MTX after dilution.
The precision and accuracy of the method were investigated by adding different concentrations of MTX. Recoveries obtained for different concentrations of MTX spiked to urine samples were between 98.4% and 104.8% (
Table 2) with the RSDs less than 3%. Inter day RSD values for concentrations of 200 and 40 μg/mL were 0.7% and 1.5%, respectively. Therefore, the proposed method was easy to perform for the direct determination of MTX without requiring any pretreatment step.
Determination of MTX in patient serum samples
The developed method was applied to the determination of MTX in human serum samples after a prior protein precipitation using acetonitrile with serum (acetonitrile ratio of 1 : 2). It should be noted that the dilution (1000 fold) of serum samples following the protein precipitation step is necessary. The dilution diminishes the high background fluorescence measured for blank serum sample.
The precision and accuracy of the method were measured by the added-found procedure. Spiked serum samples were prepared with addition of different concentrations of MTX, around therapeutic concentrations in human serum (50-300 μg/mL) and analyzed on different days for measuring inter-day precision (using the calibration graph on each day). Intra-day precision was obtained by replicate analysesof the serum samples on the same day. The data of
Table 3 demonstrate an acceptable precision and accuracy over the investigated concentration range. Recoveries obtained for different concentrations of MTX in spiked serum samples were between 97.5% and 110.5% (
Table 3) with the RSDs less than 5%. Therefore, the method is appropriate for determination of MTX in serum and meets the requirement of rapidity for routine use in high dose MTX therapy. Also, the advantage of a low-cost and easily handling instrument should be taken into consideration.
| Added (μg/mL) | Found * (μg/mL) | Recovery (%) | RSD (%) |
|---|
| Intra-day | 50.0 | 55.2 ± 1.8 | 110.5 ± 3.3 | 3.3 |
| 100.0 | 106.3 ± 2.6 | 106.7 ± 2.6 | 2.4 |
| 250.0 | 254.5 ± 0.7 | 98.2 ± 0.30 | 0.30 |
| 300.0 | 314.5 ± 4.0 | 104.9 ± 1.3 | 1.3 |
| Inter-day | 50.0 | 53.2 ± 2.4 | 106.4 ± 4.8 | 4.5 |
| 100.0 | 106.4 ± 1.4 | 106.4 ± 1.4 | 1.3 |
| 250.0 | 243.7 ± 3.7 | 97.5 ± 1.5 | 1.5 |
| 300.0 | 323.0 ± 7.1 | 107.7 ± 2.4 | 2.2 |
The regression equation (n = 21) was I0/If = 0.7918 C + 0.9953 (r = 0.999) and the standard deviations (SD) for the slope and intercept of the calibration graph were ± 0.0012 and ± 0.0078, respectively. Results of analysis of a real serum sample showed that MTX level was 2244 mg/L (sample was diluted up to 4000 times to be in the linear range of the calibration graph).