We observed minimum ESD in center no. 4 (a maternity hospital), due to the high kVp (61-62) and low mAs (0.5) used. In center no. 5 (a general hospital), a higher ESD was obtained, radiation field was 1000-1200 cm2, lack of proper collimation was probably the main reason for the higher patient dose. Short FFD (58cm) caused higher mean ESD in center no. 8 (a general hospital) for both types of examinations. In center no. 2 (an educational hospital), patients who underwent abdominal radiographies received relatively higher ESD; this may be associated with the selection of high mAs equal to 4. During this study, center no. 3 used CR (Computed Radiography) technologies while the other used conventional screen film (speed 400). Data given in
Figure 1 and
2 were optimized for screen film systems and not for computed radiography systems used nowadays. The inappropriate field size is the most important mistake in the pediatric radiographic technique. A field which is too large will not only impairs the image contrast and resolution by increasing the amount of scattered radiation, but more importantly results in unnecessary irradiation of the body. However, correct beam limitation requires proper knowledge of the external anatomical landmarks by the technician. This illustrates the need for both theoretical and practical teaching of the technicians. Deviation between the radiographic parameters for each NICU indicates that examinations were taken by a large number of radiographers. It means that technicians who practice in NICUs had no special training for the job. although former studies recommended that every X-ray department should assign a group specializing in pediatric imaging. The differences found in radiographic parameters may cause variations in the doses to which neonates are exposed. It should be noted that the radiographic voltage may depend on the type and age of the X-ray tube.
Table 2 provides patients' data, mean ESD and DRL comparison with former similar investigations. Faghihi et al. determined the values of ESD, dose area products and energy imparted by three methods including direct method [using thermoluminescence dosimetry (TLD)], indirect method (using tube output) and Monte Carlo (MC) method. Their results indicate that the mean ESD per radiograph estimated by direct, indirect and MC methods are 56.6 + 4.1, 50.1 + 3.1 and 54.5 m + 3.3 µGy, respectively. They have not illustrated that these doses were related to which exam, but they have indicated dose per radiograph (
10). In our previous study, the mean ESD was determined for neonates under stationary and mobile units. The mean ESD was 191.85 µGy for the chest and 197.30 µGy for the abdomen. The main reason was usage of grid for the infant in the hospitals (
9). In addition, Smans et al. categorized the newborn infants into three birth weight groups (< 1000 gr, 1000-2500 gr and > 2500gr) and found the mean ESD is increased with the weight at birth and reported the mean ESDs of 28, 33 and 52 µGy for the groups, respectively. So for comparison, we chose their third group based on the infants’ weight at birth (
6). Olgar et al. also used TLD for measuring ESDs, their results showed that neonates received acceptable doses from common radiological examinations in Turkey (
8). Brindhaban et al. reported the birth weight range was between 750 and 2000gr for infants (
12). However, in this study, the mean ESD for chest and abdomen examination were 76.3 µGy and 61.5µGy, respectively; also DRLs for neonates in our province were 88 µGy for chest and 98µGy for abdomen examinations that were slightly higher than other studies; European national diagnostic reference levels of 80 µGy for mobile chest radiographs and the NRPB reference dose of 50 µGy for chest examination. The main reason was related to using a high mAs and a low kVp in most departments and also a low FFD. Probably lack of collimation affected some exams in NICUs. Increased kVp (reduced mAs) causes greater penetration and less absorption then reduced patient dose for a constant film density. We should mention that the range of applied potential values was 41-61 kVp for chest exam, while the CEC recommendations are (60-65 kVp for neonatal, 70-80 kVp for children up to 5 years and 100-120 kVp for older children) (
13,
14). The recommended kVp causes less contrast, but better assessment of the lung parenchyma. Lower kVp is necessary if looking for bone details (
15). It is necessary to encourage radiographers to use suitable exposure factors and good collimation. Variations in size are marked not only for adults, but also for pediatric patients and the use of a single reference size is impractical to determine exposures factors. The large variation in ESD values indicate that patient dose can be diminished by paying more attention to exposure factors (kV, mAs), without loss of image quality, pointing to the importance of quality control programs. Staff should try to nurture using shielding for neonates such as additional collimation to achieve non-regular field shapes by placing lead sheets on the top of the incubator. Therefore, for dose reduction, several actions at national as well as international levels should be taken. Those actions should prevent the increasing risk of long-term effects to neonates, a unique group which undergoes multiple diagnostic examinations. All in all, the examination technique in pediatric radiology should be optimized. Establishing local diagnostic reference levels (LDRL) in each department or maybe each X-ray room can be an effective way for paying attention to patient dose and optimization and it is much more effective if national reference dose levels (NDRL) can be established for pediatrics.
For decreasing dose, several actions should be pursued: remarking the ALARA (As Low as Reasonable Achievable) concept, defining national guidelines for good neonate radiography and, retraining radiographers to be specialized for neonatal imaging and neonatal ionization radiation hazards. The medical physicist is the best suited individual to monitor patient doses and to reduce them (if possible) without substantially compromising the efficacy of diagnostic procedures. Medical physicists are also in charge of patient safety including radiation, mechanical and electrical safety.