Chemicals and reagents
All organic solvents were obtained from Fisher (Fair Lawn, NJ UK). Sodium chloride (≥ 99%), anhydrous sodium sulfate (≥ 99%), dipotassium hydrogen phosphate (≥ 98%), sodium bicarbonate in powder form and ACS grade (≥ 98%) were obtained from Aldrich (Milwaukee, WI. US). Purified water was prepared from Milli-Q water purification system (Millipore, Beford, MA). The purity of pesticide standards and internal standard solutions were ≥ 95% and were obtained from Restek (Thames Restek CO., UK).
Phosphate buffer (1 mol/L) was prepared by dissolving 105 g of dipotassium hydrogen phosphate and 61 g of potassium dihydrogen phosphate in 1 L of water.
Stock standard solutions of 47 organochlorine pesticides (OC), organophosphorus pesticides (OP) and other pesticides were prepared in ethyl acetate at concentrations ranging from 50 to 200 ng/μL. Stock pesticide solutions were kept at 4 °C in a dark place until analysis. Spiking solutions contain pesticides at concentrations of 10–50 ng/μL. Aldrin solution at 50 ng/μL in acetone was used as internal standard.
Sampling
A total of 85 samples including 10 watermelons (
Citrullus lunatus), 10 muskmelons (
Cucumis melo L.), 10 melons (
Cucumis melo), 20 cucumbers (
Cucumis sativum), 5 carrots (
Daucus carota), 5 lettuces (
L.S. Var. Longifolia), 20 tomatos (
Solanum lycopersicum), and 5 cabbages (
Brassica oleraceae var. capitatef. alba) were obtained from wholesale fruit and vegetable center in Tehran. Sampling and transporting of fresh fruit and vegetables were performed in accordance with the general principles and methods of the Codex Alimentarius 2000 (
25). 47 pesticides were subjected to analysis, including organochlorine, organophosphorus, organonitrogen, dicarboximides, strobilurin, triazine, pyrethroids, and other chemical groups. Two kilograms of each sample were collected and prepared for residue analysis according to the European Commission (EC) directive 91/414 method (
26,
27). All samples were packaged in plastic bag and cooled by storing at -18 °C before analysis.
Sample preparation
For analysis, 30 g of chopped sample was blended and dissolved in 60 mL of ethyl acetate then mixed with 5 g of sodium bicarbonate and 40 g of anhydrous sodium sulfate (
28). The subsample was mixed for 2 min using an Ultra Turrax homogenizer (IKA T10 Cole-Parmer, US) and followed by centrifugation at 3000 rpm for 5 min at -10 °C (Becton, Dickinson, & Co., Parsippany, NJ, US). The extract was evaporated to 0.5 mL under nitrogen gas. Co-extractives are performed by passing sample through SPE and GPC columns. Envirosep-ABC (Phenomenex J2, Scientific, Colombia, USA) 60×21.2 mm i.d. columns were used in series. A 2-mL aliquot of crude-extract (i.e., equivalent of l g of original matrix) was loaded onto the HPGPC.
The flow-rate of the mobile phase (cyclohexan:ethyl acetate, 1:1, v v-1) was 5 mL min-1. The dump time was 16 min; then, the collected sample (14 mL) was evaporated to dryness using a rotary evaporator. The remaining solvent was dried using a gentle stream of nitrogen and the remainder was dissolved in l mL of toluene prior to GC analysis. The pesticide fraction obtained by HPGPC cleaning of the crude extract was evaporated to dryness using a mild nitrogen stream.
The residue was then dissolved in 10 mL ethyl acetate. To prepare “spiked sample”, a 5-mL aliquot, was carefully evaporated and the reminder was dissolved in 0.5 mL of toluene. The SPE cartridge used in the study was 500 mg, and it also consisted of 6 mL of ENVI-Carb graphitized carbon black (GCB) sorbent (Supelco, Bellefonte, PA, USA).
The cleanup procedure was conducted in 5 replicates and there was a control in which no cleanup was performed; these were set aside for later analysis. Ethyl acetate extract (10 mL) was passed through a preconditioned cartridge and the extract was collected. After filtration to remove the precipitated proteins and solids, the pH of the sample was adjusted.
The collected eluate was evaporated to <1 mL with the aid of a nitrogen stream at 40 °C. Then, acetone (10 mL) was added, and the eluate was again evaporated under nitrogen to <1.0 mL. Add 50 μL internal standard (1.0 ng/μL, final concentration of aldrin), and dilute extract to 2.5 mL with acetone and 20 μL spiking solution (10 ng/μL for pesticide mix solution) to 500 μL control extract (5 g sample per 2 mL ethyl acetate) to produce a final concentration of 1.0 ng/μL for Aldrin and 0.4 ng/μL for most compounds.
Finally for analysis 0.5 µL of the sample was injected into GC-MS. Limits of detection (LODs) of the method were reported by spiking samples at 0.1−1.0 mg/kg.
Preparation of Standard Solutions
Each stock solution of the pesticide standards was prepared at four concentration levels in ethyl acetate. Calibration standards were prepared by appropriate dilution of the stock solution in ethyl acetate at different levels (0.01- 0.5 mg/kg) in accordance with the European Commission guidelines (
29).
The 47 mix pesticide standards were prepared in aceton (10 ppm). 8 μg/mL spiking solution was created by diluting 5 μL of the stock solution in 200 μL of acetonitrile (0.1% acetic acid). 200 μL of pesticide spiking solution was gently pipetted over each produce sample to create a final pesticide concentration of 100 ppb and the tubes were mixed for one minute to disperse the pesticides. The spiked samples were kept at refrigerator.
GC-MS Ion-trap
GC-MS analysis of the fruit and vegetable samples for pesticide residues was carried out by using a Varian Model 3800 gas chromatograph (GC) fitted with an ion trap mass spectrometric ITMS Varian 2200 (Varian Instrument, CA, USA).
A DB-5 (Folsom, CA, USA) capillary column (25 m×0.25 mm I.D., 0.25-mm film thickness) was employed. The injections were performed with the column oven set at 100 °C in the splitless mode. That temperature was maintained for 1 min and then programmed at 10 °C min-1 to 200 °C and at 4 °C min-1 to 300 °C, where it was held for 3 min. The carried gas in the system was helium, which was delivered at flow rate of 1.0 mL/min.
The gas pressure program was set to 70 KPa (l min hold), 110 KPa at 8 KPa min-1, 150 KPa at 1 KPa min-1, and 174 KPa at 4 KPa min-1 (5 min hold). The Varian Saturn 2200 GC-MS and CP-3800 gas chromatograph were used. For ITMS, the temperature of the injector was 250 °C. The quantification matched calibrants were run with each batch at 5 levels.
Statistical analysis
Statistical analyses were performed using SPSS V.16 software for Windows (SPSS, Chicago, USA). One-way analysis of variance (ANOVA) procedure was used for statistical analysis. The results were reported significant at the 5% level (P < 0.05).